background image

www.ossur.com

QRC 0042 EN Rev. 1

©Copyright Össur 2013

Training checklist

1)  Train user on level ground.
2)  Train user on ramps/stairs.

Make sure the user knows how to:

1)  Charge the unit and recognize low battery level warning.
2)  Turn power ON/OFF and unlock the keypad.
3)  Adjust the heel height when changing shoes.

Charging and power

Insert the charger cable into the charging port on the user interface panel on the knee module.

Charging

:  The yellow light on the ankle unit blinks when  

 

charging and becomes a solid light when it is fully  

 charged. 
 

The knee module does not indicate charging status. 

Power

:  

Fully charged battery is sufficient for 16-24 hours of  

 

continuous use, depending on activity.

Warnings

: Auditory warning signals indicate impending power  

 loss.

Relax Mode OFF
Relax Mode ON

(with chair exit mode)

Chair Exit Mode only

1 BEEP

2 BEEP

3 BEEP

x5

Mode Sequence

Hold Press Sound

RELAx MOdE

2x BEEP
3x BEEP

Approx. 5 hrs left

Warning

Indicates

4x BEEP
5x BEEP

Approx. 4 hrs left
Approx. 3 hrs left
Approx. 2 hrs left
Ankle will return to neutral 
(last set heel height) 
Yellow LED is solid ON

1 x long high pitch 

beep

Ankle will return to neutral position and 

SYMBIONIC LEG will turn OFF

Reviews: