11
II.
If the fuses located in the fuse drawer at the back of the unit need changing, the
unit must be placed in a safe mode by switching the unit off and disconnecting
the power cord from the power socket.
4.
Unpacking
4.1
Packing List
The standard items included with the UV Ozone Cleaner are:
•
The UV Ozone Cleaner unit
•
Power supply cord
4.2
Damage Inspection
Examine the components for evidence of shipping damage. If damage has occurred, please
contact Ossila directly for further action. The UV lamp is a mercury vapour lamp and the user
may be exposed to mercury only if the glass of the lamp is cracked or broken. If this happens,
please contact Ossila after abiding by these rules to minimise exposure:
1.
Ensure the power is disconnected to avoid the risk of electrocution.
2.
Store the container and UV Ozone Cleaner in a fume hood and leave the
room under ventilation for at least 15 minutes.
3.
Using cut-resistant gloves, collect any stray pieces of glass in a sealed
container. Dispose of the glass at a suitable collection point.
5.
Specifications
The UV Ozone Cleaner specifications are shown in
Table 5.1