background image

 

 

Document Feedback 

TSL2521 ALS/Flicker 

 

 

TSL2521 ALS and Flicker Settings 

 

 

Application Note 

• PUBLIC 

 

AN001042 

• v1-01 • 2022-Jan-27 

23

 

│ 11 

 

2.5 

ALS/Flicker Gain and AGC 

Another fundamental parameter for ALS measurement is the ALS gain setting. For ALS legacy 
devices like TCS3707 there was a single AGAIN value. For TSL2521, the gain is set per modulator 
and even for every sequence individually. The gain for both modulator 0 and modulator 1 for step 0 
can be set in register 

MEAS_SEQR_STEP0_MOD_GAINX_L

. The modulator gain can range from 

0.5x to 4096x and is limited in register 

CFG8

 by the field measurement_sequencer_max_mod_gain for 

all modulators and sequencer steps. Please note that in case of enabled Automatic Gain Control 
(AGC) the modulator gain registers are changed according to the last measurement. 

The gain used for the current result set in the 

ALS_DATA

 registers can be read from the 

ALS_STATUS2

 register using the bit fields als_data0_gain_status and als_data1_gain_status. This 

gain information is important for the lux equation, since the AGC 

– if enabled – might have changed 

the originally set modulator gain.  

There are two Automatic Gain Control (AGC) methods implemented. The legacy one is the saturation 
AGC that reduces the gain in case of modulator saturation and repeats the measurement until there is 
no saturated result. This method is easy and 

– as long as the dynamic range allows it – always leads 

to a useful result but can take longer time and the time delay caused by this procedure is not 
deterministic.  

The saturation AGC can be enabled in field measurement_sequencer_agc_asat_pattern in register 

MEAS_SEQR_STEP1_MOD_PHDX_SMUX_H

 individually for each sequencer step. This setting is 

valid for both modulators.  

The second one is the predict AGC, that defines the gain for the next measurement by doing a 
measurement with reduced gain. The gain reduction can be set in register 

CFG8

 in the field 

measurement_mod_gain_reduction; by default it reduces the gain by 4 steps. The advantage of the 
predict AGC is a shorter and deterministic measurement time, but in case of abrupt changes it can 
lead to a saturated result that needs to be dropped after checking the saturation signals. 

The predict AGC can be enabled in the field measurement_sequencer_agc_predict_pattern in register 

MEAS_SEQR_STEP2_MOD_PHDX_SMUX_H

 individually for each sequencer step. This setting is 

valid for both modulators. 

The AGC measurement is done by adding an additional ALS sequence round before the actual 
measurement sequence. Both AGC methods do not need to use the same ALS integration time since 
AGC has its own dedicated agc_nr_samples fields in registers 

AGC_NR_SAMPLE[7:0] 

and 

AGC_NR_SAMPLES[10:8],

 sample_time is the same as for ALS/Flicker. That can help to reduce the 

whole ALS measurement time by using a shorter integration time for the AGC measurement. Both 
AGC methods can be used at the same time, in this case the predict AGC firstly sets the gain, but if 
still analog saturation happens the saturation AGC repeats the measurement with reduced gain. 
Please note that due to the saturation AGC method the ALS measurement time is not deterministic in 
this case too. 

How often AGC is performed is defined with mod_calib_nth_iteration in register 

MOD_CALIB_CFG0

Additionally, the ACG needs to be linked to mod_calib_nth_iteration by setting the bit 

Summary of Contents for ams TSL2521

Page 1: ...Product Document Published by ams OSRAM Group...

Page 2: ...Application Note AN001042 TSL2521 ALS Flicker Settings and Comparison to Legacy ALS Devices v1 01 2022 Jan 27...

Page 3: ...C 11 2 6 Wait Time 12 3 ALS Result Data Format 13 3 1 Internal Data Format 13 3 2 16 Bit Data Full Count Results in ALS_DATA Registers 14 3 3 20 Bit Data Result in ALS_DATA Registers Including 4 Bit R...

Page 4: ...e data compression features and measurements can be planned in sequences with individual settings Additionally a new residual measurement feature offers higher resolution at lower gain These new featu...

Page 5: ...odes like Clear C and Infrared IR but after the two available ALS modulators modulator 0 and modulator 1 Figure 1 shows the number and filter type of the six available photodiodes on TSL2521 The photo...

Page 6: ...ement_sequencer_als_pattern in register MEAS_SEQR_ALS_FD_1 for both modulators by default ALS with both modulators is enabled for step 0 only Flicker measurement can be enabled for each modulator indi...

Page 7: ...ution results in an effective gain of 4096x in this case but at the gain non linearity of 256x which is rather low In order to use the second gain table mod_gain_select has to be set to 3 before start...

Page 8: ...red or not the way to get back to the same number space comparable to legacy devices is to divide the results by 16 in this case the decimal places show the residual amount The GUI provided by ams OSR...

Page 9: ...al measurement time This reduces the ALS integration time accordingly but only by a very small amount and in case of no flicker measurement during the same step most times it can be neglected as shown...

Page 10: ...n as set in fd_nr_samples in registers FD_NR_SAMPLES0 and FD_NR_SAMPLES1 and increases at the same time the ALS result as often as set in als_nr_samples in registers ALS_NR_SAMPLES0 and ALS_NR_SAMPLES...

Page 11: ...ation and Proposed Settings Item Value dec Customer Example Specification Update interval 0 5 s ALS integraton time 50 ms Max flicker frequency 2 kHz FFT bin size 10 Hz Proposal Sample time 125 s Samp...

Page 12: ...nt_sequencer_agc_asat_pattern in register MEAS_SEQR_STEP1_MOD_PHDX_SMUX_H individually for each sequencer step This setting is valid for both modulators The second one is the predict AGC that defines...

Page 13: ...sabled by clearing bit mod_calib_nth_iteration_az_enable 2 6 Wait Time Related to ALS measurement on legacy ALS devices like TCS3707 was the Wait time that defined the sample rate of the ALS measureme...

Page 14: ...As shown in Figure 10 the maximum result width of 26 bit including 4 bit Residual counts requires als_msb_position to be set to 6 In most applications the second shown format of 24 bit is a good setti...

Page 15: ...taining Residuals later since the number space is a different one without the 4 bits on LSB side 3 3 20 Bit Data Result in ALS_DATA Registers Including 4 Bit Residuals It is possible to transfer the 1...

Page 16: ...full counts result in the ALS_DATAx is shifted by 4 bits to the left or simply multiplied by 16 on host side to keep the data consistent over the whole range and to stay in the same 20 bit number spa...

Page 17: ...urement in different steps of the sequencer If this is not possible the bit do_als_final_processing in register CFG1 has to be set This delays the writing of ALS data until all flicker data is written...

Page 18: ...0xFFFE in case of 24 bit FIFO format to 0xFFFFFE Digital saturation sets the modulator interrupt flag mint in register STATUS as well Like for analog saturation this can be solved by reducing the gain...

Page 19: ...Document Feedback TSL2521 ALS Flicker ALS Result Data Format Application Note PUBLIC AN001042 v1 01 2022 Jan 27 23 18 Equation 8 _ _ _ _ _ _ 1 _ 1 22 16...

Page 20: ...to reduce the number of transferred flicker data bits in order to reduce I2 C traffic For high flicker samples rates the dynamic range is lower so the absolute number of bits of a transferred sample...

Page 21: ...t reduces the I2 C traffic significantly the drivers provided by ams OSRAM show and document the use of these features 4 3 Flicker Data Transfer Via FIFO Flicker data is always written to the FIFO if...

Page 22: ...1 Bit and Filled Up FIFO Byte The flicker data can be read out from register FIFO_DATA at any time The right procedure is to get firstly fifo_lvl by reading FIFO_LEVEL 0xFD and FIFO_STATUS0 0xFE as bl...

Page 23: ...formation Changes from previous version to current revision v1 01 Page Initial version Corrected number of preferred modulator for flicker measurements 4 Page and figure numbers for the previous versi...

Page 24: ...otice RoHS Compliant ams Green Statement RoHS Compliant The term RoHS compliant means that ams OSRAM AG products fully comply with current RoHS directives Our semiconductor products do not contain any...

Reviews: