background image

Series3 Auto User Guide

11

NOTE: 

Please follow instructions below for 

detailed operating procedure of each type.
Step 1  Turn on the power. A beep sound will 

begin the operation.

Step 2  The pump will enter into static mode 

when the power is initially turned on.

 

First it will inflate the mattress to 

25mmHg and then start weight 

calculation automatically. During 

calculation, the 8 LED indicators will 

flash until the calculation is finished, then 

mattress will enter alternate mode.

Step 3  The alternating cycle time is set at 10 min. 

Select from the touch panel to adjust the 

cycle time.

 

NOTE: Press the static mode button from 

the touch panel to provide a firm surface 

that makes it easier for the patient to 

transfer or reposition. The static mode 

can support the patient by keeping the 

mattress from bottoming out when the 

patient is in a sitting position.

Step 4  During normal operation, the unit will 

monitor pressure. If the mattress pressure 

is lower than the set pressure, the pump 

will automatically inflate the mattress to 

the set level and then stop. The alarm will 

beep and its LED will come on to alert a 

low pressure condition.

           Press the Alarm Mute button to mute 

the alarm. The alarm LED will continue 

flashing.

           Press the Alarm Mute button again to 

re-enable the audible alarm.

Step 5  Press the Max Firm button to 

automatically inflate the mattress to the 

maximum level (55mmHg) for about 20 

minutes. The pressure will return to a 

previously set level after 20 minutes.

           In case there is a bad connection of 

tubing or any air leakage from the 

mattress, this will result in a low pressure 

condition, and it will activate the alarm 

within 15 mins.

 

NOTE:

 For suitable pressure, please refer 

to page 8 for the hand check procedure.

CPR function 

When there is an emergency requirement to 

perform CPR on the patient, press the CPR button 

at the head section of the mattress and disconnect 

the air hose from the pump to release the air 

quickly from the mattress. 

Pressure set up

The pump unit can automatically set the pressure 

according to patient’s weight. Other functions will 

not work during weight calculation, except the 

Max Firm. Users can also adjust the pressure level 

of the air mattress to a desired firmness manually 

(press the button for 2 seconds).  
NOTE: It is recommended to press Max Firm on 

the panel when the mattress is first inflated. The 

auto weight calculation will start after Max Firm is 

finished

.   

Low pressure warning 

When abnormal pressure occurs, the Low 

Pressure indicator will come on. For some types 

of pumps, the alarm will be activated to alert of a 

low pressure condition. Check if the connections 

are secure and correctly installed according to the 

relevant instructions.   

NOTE: If the pressure is consistently low, open 

the zipper and confirm that all of the hoses are 

properly connected. Then check for any noticeable 

leakage in any of the tubes. If necessary, contact 

your local dealer to replace any damaged tubes 

or hoses.

Power failure

If the power is cut, only the power failure alarm 

works. The low pressure alarm can only work 

under power connected. The power failure alarm 

could last for 2 hours.

If the pump remains connected, the air will leak 

gradually and after 4 hours, the upper air chamber 

(dynamic) will be totally deflated. 

If the pump is disconnected and the transport cap 

is in place the air leaks gradually and after 8 hours, 

the upper air chamber (dynamic) will be totally 

deflated.
The lower air chamber (static) could last for at 

least 24 hours under CPR valve being well installed 

(not be unplugged).

Summary of Contents for 3 Auto Series

Page 1: ...Version 3 17 11 2022 User Guide Series3 Auto ...

Page 2: ...ure warning 12 Cleaning 13 Environment requirements 13 Handling and storage 13 Maintenance 15 16 Troubleshooting 17 Specifications Know your symbols Manufacturer Do not iron Foot end Tumble dry CE marked Do not dry clean REF Catalogue number Resistant to ignition REP EC Authorized EC Representative Wipe down Machine wash at 95 Recycling Drip dry UKCA marked 1 CL 1 Chlorine Medical Device Airflo Xi...

Page 3: ...a properly grounded outlet only Do not disassemble repair or rebuild the product without permission otherwise may has risk of fire and danger in life Grounding Before any connection is made make certain that this product shall be connected to a protective earth conductor via the three wire power cord the three blade plug shall be inserted only into a socket outlet that provides a protective earth ...

Page 4: ...on for 3 seconds to lock the panel press again for 3 seconds to unlock the panel Power Switch 1 The switch is at the right side of the pump Turn ON OFF the power and press Power button on the panel the pump will start stop operation Press Cycle Time 2 The cycle time can be selected from the panel The cycle time value options are 10 15 20 or 25 minutes Auto weight 3 Auto Press the button the pump w...

Page 5: ... under each end of your bed Ensure buckles are securely fastened and straps are pulled tight c IMPORTANT On a profiling bed secure the side straps around the moveable sections of the bed base DO NOT secure to the side rails or main bed frame irreparable damage could be caused to the bed and mattress d Confirm there are no sharp objects in the immediate area which may risk damage to the Mattress Re...

Page 6: ...le pressure please refer to page 8 for the hand check procedure CPR function When there is an emergency requirement to perform CPR on the patient press the CPR button at the head section of the mattress and disconnect the air hose from the pump to release the air quickly from the mattress Pressure set up The pump unit can automatically set the pressure according to patient s weight Other functions...

Page 7: ...e to disconnect all the air cells one by one and spray the disinfectant on all sides including all the connecting tubes and hoses Let it sit for least 10 minutes If there is a base after you remove all the air cells the base has to be sprayed down with the disinfectant inside and outside Let it sit for the required incubation time and wipe down with a cloth Repeat the process with the tubing set s...

Page 8: ...Change air cell Troubleshooting guide Problems Reasons Maintenance Mattress has pillow function but air cell fail to inflate 1 One way valve is assembled reversely 1 Assemble the one way valve in correct direction Pump is working but synchronous motor does not work thus mattress does not alternate and alternate failure alarm is activated 1 Synchronous motor is out of order 2 Wires inside synchrono...

Page 9: ...able Nylon PVC 840D Case Material Flame Retardant ABS 20 x 8 Cell on Cell air cells Max firm mode CPR valve for emergency procedures Alarm Low pressure Power failure Alternate failure Pillow function Alternate Static Seat inflate Auto weight mode Quick connectors Size 31cm L x 12 5cm W x 21cm H Maximum Weight Capacity 200 kg Weight 2 5 kg Note The above specifications are also applicable to those ...

Page 10: ...Series3 Auto User Guide 19 Notes Notes ...

Page 11: ...Series3 Auto User Guide Edward House 5 Penner Road Havant PO9 1QZ 02394 318 318 ask oska uk com oska uk com Pressure Care Experts ...

Reviews: