Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF communications equipment
should be used no closer to any part of T830/T840,
including cables, than the recommended separation dis-
tance calculated from the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance:
d = 1.2
d = 1.2
for 80 MHz to 800 MHz
d = 2.3
for 80 MHz to 2.5 GHz,
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation dis-
tance in metres (m).
Field strengths from fixed RF transmitters, as determ-
ined by an electromagnetic site survey,
a
should be less
than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with this symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
Note 1
: At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
Note 2
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which T830/T840 is used exceeds the applicable RF com-
pliance level above, the T830/T840 should be observed to verify normal operation. If abnormal performance is observed, additional measures might be neces-
sary, such as reorienting or relocating T830/T840.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and T830/T840
The T830/T840 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
T830/T840 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the T830/T840as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
27
User Guide
Inmedico A/S - T830/T840