Hearing level offsets
The calibration software currently calibrates to RETSPL values according to
IEC 60645-1:2012 or ANSI S3.6-2010. To change between IEC and ANSI
values, open the
Settings
menu and change in the
General
tab.
Click on the
ISO
button to see the table, if you need
to replace or change RETSPL values, you are able
to do this in the
Config.xml
file that is placed in the
program folder.
<!--RetSPL values 0.0 to 200.0-->
<RetSPL_TDH39_NameOfStandart>ISO 389-
1:1998</RetSPL_TDH39_NameOfStandart>
<RetSPL_TDH39_125>45</RetSPL_TDH39_125>
<RetSPL_TDH39_250>25,5</RetSPL_TDH39_250>
<RetSPL_TDH39_500>11,5</RetSPL_TDH39_500>
<RetSPL_TDH39_750>7,5</RetSPL_TDH39_750>
<RetSPL_TDH39_1000>7</RetSPL_TDH39_1000>
<RetSPL_TDH39_1500>6,5</RetSPL_TDH39_1500>
<RetSPL_TDH39_2000>9</RetSPL_TDH39_2000>
<RetSPL_TDH39_3000>10</RetSPL_TDH39_3000>
<RetSPL_TDH39_4000>9,5</RetSPL_TDH39_4000>
<RetSPL_TDH39_6000>15,5</RetSPL_TDH39_6000>
<RetSPL_TDH39_8000>13</RetSPL_TDH39_8000>