Installation instructions
When the cabinet is being built in ventilation must be provided.
The ventilation hole above the cabinet must be at least 50 mm or one side must be open. Alternatively
a 50 x 500 mm ventilation hole can be made in the panel above the cabinet. See the sketch.
If the cabinet is installed in a high unit, the 50 x 500 mm ventilation hole must cut at the top of the
panels to achieve a chimney effect. See the sketch.
Cleaning and maintenance
•
Clean visible components using a damp cloth.
•
Use cleaning products generally used in healthcare.
•
Wipe clean using a dry cloth.
•
Do not clean the medication cabinet by pouring water on it.
•
Check the defrost water container (the large bottle), empty as necessary.
•
Check that the defrost water hose running from the unit down to the defrost water container is
clean.
•
The small bottle contains glycerol and helps to measure the temperature in the cabinet and
must not be emptied.
•
The refrigerator has power cables, check that these are intact. Never use cables that are
defective or in poor condition. If power cables are damaged they must be replaced by
authorised personnel or by the manufacturer.
•
The refrigerator is factory calibrated.