7 - PROBLEMS, MAINTENANCE AND WARRANTY

ERROR SIGNALS

The instrument display is used to view instrument error 

conditions by displaying the following messages:

 "E1" 

- Tr probe 

error (E1) interrupted or short-circuited.

 "o1" - "u1" 

- Tr (1) 

probe input in overrange (o) or in underrange (u). In these 

cases, verify the correct connection of the probes with the 

instrument and then proceed to verify them.

"EE" 

- Memory error, in this case check, and if necessary, 

reprogram the operating parameters.

MAINTENANCE

It is recommended to avoid using abrasive detergents or 

solvents that can damage the instrument.

WARRANTY AND REPAIR

The instrument is warranted for 12 months against construction 

faults or material defects. The warranty is limited to the repair 

or replacement of the product. The eventual opening of the 

casing, the manipulation of the instrument or the use and 

installation not in accordance with the instructions will 

automatically void the warranty. In the event that the product is 

defective within or outside the warranty period, contact OSAKA 

to obtain authorization for the shipment. Send the defective 

product, accompanied by the indications of the defect found, to 

the OSAKA establishment, except for different agreements.

8 - HOW TO MODIFY THE DECIMAL POINT IN THE (OI 12 

NTC, PT100, J / K / S, mA and V)

To change the decimal point, first remove the front of the 

equipment by inserting a screwdriver and making a slight turn 

in the small slots shown in the figure of the equipment casing:

Press the key "

1

”For about 2 seconds until“ OPEr ”appears on 

the display. Press the key "

1

”And“ InP ”will appear. Press the key 

"

1

”To enter the folder and use the keys

 

2

" Y "

3

”To scroll 

through the parameters. To modify the decimal point, go to the 

parameter "dP" and press the key "

1

”. Use the keys " 

2

" Y "

3

”To 

increase or decrease the value of said parameter ( 

on = 

decimal 

point on) and (

oF = 

decimal point off). Press the key "

1

" to 

confirm. To exit, press and hold the “

2

" Y "

3

 

Until leaving.