10
Setup
Locating the SCA
To maximize ease of use and operational safety, make sure the following preparations
of the physical site are met:
•
Place the SCA on a sturdy horizontal surface or cart capable of supporting 200 lbs
(the weight when filled with water).
•
The selected location should be isolated from unintentional contact from those
working in the area.
•
The electrical outlet must be located within 2 m (6 ft) of the SCA. Do not use an
extension cord or power strip with the system.
•
The SCA should be near a grounded wall outlet such that the power cord does not
pose a hazard to people or equipment passing by.
•
The grounded electrical outlet (US: 110VAC, 20A, 60Hz, single phase, Europe:
220VAC, 16A, 50Hz)
•
The power cord plug-in receptacle is a disconnecting device that should be easily
accessible at all times. Position the unit so that the plug can be easily reached.
•
It is strongly recommended that the unit be placed on a circuit with its own GFCI
breaker.
•
Place the unit in a well-ventilated area.
•
The operating environmental temperature must be between
5⁰C
-
40⁰C (41⁰F –
104⁰F); and operating environmental humidity between 0%
- 80% RH.
Preparations for draining and disposal
•
Consult your local regulations regarding disposal of the cleaning solution effluent
prior to use. A permit or form of pre-treatment may be required.
•
The SCA relies on gravity to drain the cleaning tank. Locating the system near a drain
or sink will make draining the unit much easier. Alternatively, the SCA can be kept on
a cart for easy transport to a sink or drain.
•
Use a 1” inner diameter hose made of PVC or equivalent on the drain tube to
facilitate draining. Ensure the tubing or hose can withstand the temperature and
caustic properties of the cleaning solution.