Orthofix CONSTRUX Mini PTC Instructions For Use Manual Download Page 2

2

English

EN

INSTRUCTIONS FOR USE

Important Information – Please Read Prior to Use

Device System Name:

CONSTRUX

®

 Mini PTC Spacer System 

Description:

The CONSTRUX Mini PTC Spacer System is comprised of a variety of implants that have a PEEK 

core with integrated porous titanium end plates. CONSTRUX Mini PTC spacers are implanted in 

the cervical intervertebral disc space and are intended to facilitate vertebral fusion by stabilizing 

adjacent vertebrae, maintaining disc height and preventing the collapse of one vertebra onto 

another. 

The CONSTRUX Mini PTC Spacer System is not intended to be used as a stand-alone device and 

must be used with supplemental fixation. The implants are used singly and are implanted using 

an anterior approach.

The CONSTRUX Mini PTC implants are provided sterile.

CONSTRUX Mini PTC implants are design to be used with CONSTRUX Mini PEEK Spacer System 

instrumentation. The CONSTRUX Mini PTC implants are not compatible with components or 

metal from any other manufacturer’s system.

Indications for Use:

The CONSTRUX Mini PTC Spacer System is indicated for spinal fusion procedures at one or 

two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with 

degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with 

degeneration of the disc confirmed by patient history and radiographic studies.

The CONSTRUX Mini PTC Spacer System is intended for use with autograft and/or allograft 

comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system 

(i.e., an anterior cervical plate such as the Orthofix Hallmark Anterior Cervical Plate System).

Patients must have undergone a regimen of at least six weeks of non-operative treatment prior 

to being treated with the CONSTRUX Mini PTC Spacer System in the cervical spine.

Contraindications:

The CONSTRUX Mini PTC Spacer System, as with other orthopedic implants, is contraindicated 

for use in patients with:

1.  Active infections in which the use of an implant could preclude adequate and appropriate 

treatment of the infection.

2.  Rapidly progressive joint disease or bone absorption syndromes such as Paget’s disease, 

osteopenia, osteoporosis, or osteomyelitis which may prevent adequate fixation.

3.  Conditions that may place excessive stresses on bone and implants such as severe obesity, 

pregnancy or degenerative diseases. The decision to use this system in such conditions 

must be made by the physician taking into account the risks versus the benefits to the 

patient.

4.  Prior fusion at the level to be treated.

Potential Adverse Events:

Potential adverse events include, but are not limited to:

1.  Failure of the device to provide adequate mechanical stability.

2.  Loss of fixation of the implant.

3.  Device component failure.

4.  Migration or bending of the device.

5.  Loss of bony alignment.

6.  Non-union.

7.  Fracture of bony structures.

8.  Resorption without incorporation of any bone graft utilized.

9.  Immunogenic response to the implant materials.

Note: As with any major surgical procedure, there are risks involved in orthopedic surgery. 

Infrequent operative and postoperative complications known to occur are: early or late infection, 

which may result in the need for additional surgeries, damage to blood vessels, spinal cord 

or peripheral nerves, pulmonary emboli, loss of sensory and/or motor function, impotence, 

permanent pain and/or deformity. In rare instances, some complications may be fatal.

Warnings and Precautions:

The surgeon should be aware of the following when using implants:

1.  The correct selection of the implant is extremely important. The potential for success 

is increased by the selection of the proper size, shape, and design of the implant. No 

implant can be expected to withstand the unsupported stresses of full weight bearing. 

The size, shape and condition of human bones are also contributing factors to the success 

of the surgery. 

2.  DO NOT USE DAMAGED IMPLANTS. The correct handling of the implant is extremely 

important. Implants should not be bent, notched or scratched. These operations can 

produce defects in surface finish and may cause internal stress concentrations which may 

become the focal point for eventual failure of the device.

3.  CONSTRUX Mini PTC implants are provided STERILE. Do not use if the package is opened 

or damaged or if the expiration date has passed.

4.  DO NOT re-sterilize these implants as this could result in injury or require reoperation due 

to breakage.

5.  SINGLE USE ONLY – CONSTRUX Mini PTC Spacer System implants are intended for Single 

Use Only. No surgical implants should be reused. Any implant once used should be 

discarded. Even though the device appears undamaged, it may already have small defects 

and internal stress patterns that may lead to fatigue failure. Reuse could result in injury or 

require reoperation due to breakage or infection.

6.  Non-Sterile; the CONSTRUX Mini PEEK Spacer System instruments are provided non-

sterile, and therefore, must be thoroughly cleaned and sterilized before initial use and 

after each subsequent use.

7.  Postoperative care is important. The patient should be instructed in the limitations of the 

implant and should be cautioned regarding weight bearing and body stress on the device 

prior to secure bone healing.

8.  Based on dynamic testing results, the physician should consider the levels of implantation, 

patient weight, patient activity level and other patient conditions that may impact the 

performance of the intervertebral body fusion device.

9.  The implantation of the intervertebral body fusion device should be performed only by 

experienced spinal surgeons with specific training in the use of the device because it is a 

technically demanding procedure presenting a risk of serious injury to the patient.

MRI Compatibility Information:

The CONSTRUX Mini PTC Spacer System has not been evaluated for safety and compatibility in 

the Magnetic Resonance (MR) environment.  It has not been tested for heating, migration, or 

image artifact in the MR environment. The safety of CONSTRUX Mini PTC Spacer System in the 

MR environment is unknown. Scanning a patient who has this device may result in patient injury.

Cleaning:

CONSTRUX Mini PTC Spacer System implants are provided STERILE. Do not use the implant if the 

package is opened or damaged or if the expiration date has passed. Please discard all open and 

unused implants. Do not re-sterilize an opened and unused implant.

The CONSTRUX Mini PEEK Spacer System instruments are compatible for use with the CONSTRUX 

Mini PTC Spacer System. All instruments must be thoroughly cleaned and sterilized after each 

use. Cleaning may be done using validated hospital methods or following the validated cleaning 

processes described below.

None of the instruments require disassembly prior to cleaning.

From Point of Use:

Whenever possible, do not allow blood, debris or body fluids to dry on instruments. For best 

results and to prolong the life of the surgical instrument reprocess immediately after use. 

1.  Remove excess body fluids and tissue from instruments with a disposable, non-shedding 

wipe. Place instruments in a basin of purified water or in a tray covered with damp 

towels. Do not allow saline, blood, body fluids, tissue, bone fragments or other organic 

debris to dry on instruments prior to cleaning.

2.  For optimal results, instruments should be cleaned within 30 minutes of use or after 

removal from solution to minimize the potential for drying prior to cleaning.

3.  Used instruments must be transported to the central supply in closed or covered 

containers to prevent unnecessary contamination risk.

Note:

 Soaking in proteolytic enzymatic detergents or other pre-cleaning solutions facilitates 

cleaning, especially in instruments with complex features and hard-to-reach areas (e.g. 

cannulated and tubular designs, etc.). These enzymatic detergents as well as enzymatic foam 

sprays break down protein matter and prevent blood and protein based materials from drying 

on instruments. Manufacturer’s instructions for preparation and use of these solutions should be 

explicitly followed.

Preparation for Cleaning:

1.  All instruments with moving parts (e.g., knobs, triggers, hinges) should be placed in the 

open position to allow access of the cleaning fluid to areas that are difficult to clean.

2.  Soak the instruments for a minimum of 10 minutes in purified water prior to the manual 

or automated cleaning process.

3.  Use a soft cloth or a soft plastic bristle brush to remove any visible soil from the 

instruments prior to manual or automated cleaning. Use a soft plastic bristle brush 

or a pipe cleaner to remove soil from any inner lumens. You can also use a syringe (if 

appropriate) for hard to reach areas.

Orthofix Inc.

3451 Plano Parkway

Lewisville, Texas 75056-9453 U.S.A.

1-214-937-3199

1-888-298-5700

www.orthofix.com

Medical Device Safety Service 

(MDSS)

Schiffgraben 41  

30175 Hannover

Germany

+49 511 6262 8630

www.mdss.com

Australian Sponsor 

Emergo Australia 

Level 20, Tower II 

Darling Park 

201 Sussex Street 

Sydney, NSW 2000 

Australia

Summary of Contents for CONSTRUX Mini PTC

Page 1: ...6 7 ZH TW Click directory below for desired language Orthofix Inc 3451 Plano Parkway Lewisville Texas 75056 9453 U S A 1 214 937 3199 1 888 298 5700 www orthofix com Medical Device Safety Service MDS...

Page 2: ...jury or require reoperation due to breakage 5 SINGLE USE ONLY CONSTRUX Mini PTC Spacer System implants are intended for Single Use Only No surgical implants should be reused Any implant once used shou...

Page 3: ...rilization Do not re sterilize Sterilization in Orthofix Cases with Blue Wrap The CONSTRUX Mini PEEK Spacer System instruments are supplied NON STERILE and are compatible for use with the CONSTRUX Min...

Page 4: ...lantes no deben doblarse marcarse con una muesca ni rayarse Estas manipulaciones pueden producir defectos en el acabado de su superficie y podr an causar concentraciones de las tensiones internas que...

Page 5: ...ible Nota Ciertas soluciones de limpieza como las que contienen sosa c ustica formol glutaraldeh do lej a u otras soluciones de limpieza alcalinas podr an da ar los instrumentos Estas soluciones no de...

Page 6: ...X Mini PTC 4 5 CONSTRUX Mini PTC 6 CONSTRUX Mini PEEK 7 8 9 CONSTRUX Mini PTC CONSTRUX Mini PTC CONSTRUX Mini PTC CONSTRUX Mini PEEK CONSTRUX Mini PTC 1 2 30 3 ZH TW Orthofix Inc 3451 Plano Parkway Le...

Page 7: ...C 15 4 30 4 30 Orthofix 59 1 cm x 28 6 cm 132 C 4 4 30 ANSI AAMI ST79 ANSI AAMI ST79 Comprehensive guide to steam sterilization and sterility assurance in health care facilities 10 6 Orthofix CONSTRUX...

Page 8: ...r zerkratzt werden Eine solche Behandlung kann Defekte in der Oberfl chenbearbeitung und interne Belastungskonzentrationen verursachen die sp ter zum Ausgangspunkt eines Implantatversagens werden k nn...

Page 9: ...sichtbaren Verschmutzungen entfernt sind Hinweis Bestimmte Reinigungsl sungen wie z B solche die Natronlauge Formalin Glutaraldehyd Bleichmittel und oder andere alkalische Reinigungsmittel enthalten k...