Direction for Use AcceleDent® Aura
Electromagnetic Immunity – Part 1
The Activator is intended for use in the electromagnetic environment
specified below. The customer or the user of the Activator should assure
that it is used in such an environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
± 6 kV contact
± 8 kV air
± 2 kV
for power supply
lines
± 1 kV
for input/output
lines
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
3 Alm
3 Alm
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
± 6 kV contact
± 8 kV air
Not Applicable
Not Applicable
Not Applicable
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Not Applicable
Not Applicable
Not Applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Immunity Test
Electromagnetic
environment-guidance
Compliance
Level
IEC 60601
Test Level