radEAGLE
User Manual
Using the
radEAGLE
2.3 Starting Up the
radEAGLE
1. Position the instrument in a low-radiation environment.
2. Position the supplied KCl (pottasium chloride) box (Fig.
, p.
) in front of the detec-
tor cap if your
radEAGLE
has no built-in source.
Figure 17:
Box with potassium chloride (KCl) for stabilization and calibration
3. Press and hold down the
key (Fig.
, p.
). (Ensure the battery slide switch behind
the latch (Fig.
, p.
) is in the top postion “ON”.)
4. After a couple of seconds the instrument begins to boot when the alarm LEDs (Fig.
p.
) begin to flash.
5. You are welcomed by the start-up screen with dots appearing from left to right
(Fig.
, p.
Figure 18:
Starting the
radEAGLE
6. The
radEAGLE
begins an initial stabilization (Fig.
, p.
, see
, p.
for details).
7. When the stabilization source is identified it is shown (Fig.
, p.
).
© innoRIID GmbH • 2019-04-11
Software 3.2.12 • Document 3.3.0o
29/147
Summary of Contents for radEAGLE
Page 1: ...radEAGLE User Manual Software 3 2 12 Document 3 3 0o 2019 04 11 ...
Page 20: ...20 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 40: ...40 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 106: ...106 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 118: ...118 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...
Page 134: ...134 147 Software 3 2 12 Document 3 3 0o innoRIID GmbH 2019 04 11 ...