background image

1

ORTEC MODEL 9306

1-GHz PREAMPLIFIER

1.  DESCRIPTION

The ORTEC Model 9306 1-GHz Preamplifier is
optimized for fast timing and counting applications
with detectors that deliver pulses with ultra-fast rise
times. An output rise time of 350 ps and a non-
inverting gain of 100 make the Model 9306 ideal for
use with microchannel-plate photomultipliers,
microchannel plates, channeltrons, silicon diodes,
fast photomultiplier tubes, and electron multipliers.
The compact preamplifier case with captive power
cord permits close coupling to the detector to
minimize sensitivity to environmental noise.

To preserve the ultra-fast rise time, the Model 9306
is designed to accept and deliver signals on high-
quality, 50-ohm, coaxial cables with SMA
connectors and 50-ohm terminations. The input is

ac-coupled with a 50-ohm input impedance, and is
protected to a maximum of ±1 V. Two identical
outputs are provided for convenient, simultaneous
connection to two different instruments. Both
outputs are ac-coupled, short-circuit protected, and
capable of driving pulse amplitudes from 0 to 

!

2 V

into 50-ohm loads.

The 1-GHz Preamplifier derives its +24

!

V dc power

from a NIM module or power supply via the captive
power cord and standard, 9-pin, D connector. The
ORTEC Model 4002P Portable Power Supply and
most NIM amplifiers provide the required power on
a compatible preamplifier power connector.

2.  SPECIFICATIONS*

2.1.  PERFORMANCE

All specification are measured with a pulser having
a pulse width of 2 ns FWHM, and a rise time of 150
ps. Where significant, the measurement is
corrected for the rise times of the pulser, coaxial
cable, and oscilloscope. The specifications are
identical for OUTPUTS 1 and 2.

GAIN  Nominally 100 (50 to 150), non-inverting,
into a 50-ohm output load.

OUTPUT RISE TIME  Typically 350 ps.

BANDWIDTH (3 dB)  Typically 100 kHz to 1 GHz.

NOISE  <100 

:

V rms equivalent input noise over

the full bandwidth.

2.2.  INPUTS AND OUTPUTS

INPUT  SMA input connector with 50 ohms input
impedance (ac-coupled). Input protected to a
maximum of ±1 V.

*Specifications subject to change without notice.

OUTPUT 1  SMA output connector provides a
linear output range from 0 to 

!

1.75 V and a

maximum output of 

!

2 V into a 50-ohm load.

Output impedance is 50 ohms, ac-coupled, and
short-circuit protected. The unused output must be
terminated with a 50-ohm load for proper operation
of the other output. An optional SMA 50-ohm
terminator is available for this purpose.

OUTPUT 2 Identical to OUTPUT 1

POWER Input power supplied through a captive
cable (length: 3 m) with a standard preamplifier
power connector (9-pin D type). Compatible with
ORTEC instruments that provide a preamplifier
power connector.

2.3.  ELECTRICAL AND MECHANICAL

POWER REQUIRED  +24 V at 200 mA. Captive
power cord with standard 9-pin D connector derives
power from any ORTEC instrument equipped with
the standard Preamplifier Power plug (e.g.,
spectroscopy amplifiers, 4002P Portable Power
Supply, etc.).

Summary of Contents for 9306

Page 1: ...Model 9306 1 GHz Preamplifier Operating and Service Manual Printed in U S A ORTEC Part No 761700 1202 Manual Revision F ...

Page 2: ...rn so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting the unit repaired The ORTEC standard ...

Page 3: ...RUCTIONS AND SYMBOLS iv SAFETY WARNINGS AND CLEANING INSTRUCTIONS v 1 DESCRIPTION 1 2 SPECIFICATIONS 1 2 1 PERFORMANCE 1 2 2 INPUTS AND OUTPUTS 1 2 3 ELECTRICAL AND MECHANICAL 1 2 4 WEIGHT 2 3 INSTALLATION AND OPERATION 2 4 MAINTENANCE 2 ...

Page 4: ...s a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product In addition the following symbol may appear on the product ATTENTION Refer to Manual...

Page 5: ...the instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug the instrument from the ac power supply Remove loose dust on the outside of the instrument with a lint free cloth Remove remaining dirt with a lint free cloth dampened in a general purpose detergent ...

Page 6: ...vi ...

Page 7: ...r power connector 2 SPECIFICATIONS 2 1 PERFORMANCE All specification are measured with a pulser having a pulse width of 2 ns FWHM and a rise time of 150 ps Where significant the measurement is corrected for the rise times of the pulser coaxial cable and oscilloscope The specifications are identical for OUTPUTS 1 and 2 GAIN Nominally 100 50 to 150 non inverting into a 50 ohm output load OUTPUT RISE...

Page 8: ...to note that it is not possible to provide input over voltage protection at such high frequencies Therefore the magnitude of the input signal must be less than 1 V Excessive input voltages will damage the 9306 To utilize the high frequency performance of the 9306 careful signal handling is required The input and output cables for the 9306 should be short high quality 50 ohm cables since only a few...

Reviews: