background image

8

amplitude of the resulting output pulse, E

out

, is

measured in volts with an oscilloscope. The noise
spread can then be calculated from the formula

where  E

rms

 is output noise in volts on the 3400A

meter, E

in

 is the input signal in keV particle energy,

and E

out

 is the output signal in volts corresponding to

the above input. If the gain of the shaping amplifier
is adjusted so that the output pulse height is 2.35 V
for an input of 1 MeV equivalent input charge, then
the rms meter will be calibrated to read directly in
energy (1 mV = 1 keV).

7.  The noise performance of the preamplifier, as
measured by these methods, should not differ
significantly from that given in the Specifications in
Section 2.

8. If, during testing of the preamplifier and detector,
the noise performance of the preamplifier has been
verified as outlined in the previous steps or is
otherwise not suspected, a detector may be tested
to some extent by duplicating the noise performance
tests with the detector connected in place and with
normal operating bias applied. The resulting
combined noise measurement, made by either the
analyzer or voltmeter method, indicates the sum in
quadrature of the separate noise sources of the
amplifier and the detector. In other words, the total
noise is given by

9.  Each quantity is expressed in keV FWHM. The
quantity  N

det

 is known as the “noise width” of the

detector and is included as one of its specified

parameters. By use of the above equation and with
a knowledge of the noise of the preamplifier, the
noise width of the detector can be determined. The
significance of this noise width in evaluating the
detector is subject to interpretation, but generally
the actual resolution of the detector is related
directly to the noise width in its normal applications.
The most useful purpose for determining this quality
for the detector is to occasionally monitor the
detector noise width to verify that its characteristics
have not undergone any significant changes during
use.

10.  Use an ORTEC 419 Precision Pulse Generator
with a matched charge termination to measure the
risetime of the 142IH through the T or E output
connector. Connect the 419 output through the
charge terminator to the Input of the 142IH and use
an oscilloscope with a fast risetime (1 ns if
possible). The risetime of the preamplifier can then
be computed by

(Total risetime)

2

 = (Preamplifier risetime)

2

 + 

(Pulser risetime)

2

 + (Oscilloscope risetime)

The risetime of the 419 is typically 3 ns.

5.2.  FACTORY REPAIR

This instrument can be returned to ORTEC for
service and repair at a nominal cost. The standard
procedure for repair ensures the same quality
control and checkout that are used for a new
instrument. Always contact Customer Services at
ORTEC (865) 482-4411, before sending in an
instrument for repair to obtain the necessary
shipping instructions and so that the required
Return Authorization Number can be assigned to
the unit. Write this number on the address label and
on the package to ensure prompt attention when it
reaches the factory.

Summary of Contents for 142IH

Page 1: ...Model 142IH Preamplifier Operating and Service Manual Printed in U S A ORTEC Part No 717590 1202 Manual Revision C...

Page 2: ...n so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of t...

Page 3: ...TS 3 2 4 CONNECTORS 3 2 5 ELECTRICAL AND MECHANICAL 3 3 INSTALLATION 3 3 1 CONNECTION TO DETECTOR 3 3 2 CONNECTION TO AMPLIFIERS 4 3 3 INPUT POWER 4 3 4 TEST PULSE 4 3 5 DETECTOR BIAS INPUT 4 4 OPERAT...

Page 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Page 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Page 6: ...vi...

Page 7: ...ry before connecting a low impedance a cable or any other capacitive device to the Input connector on the preamplifier 2 Discharge the bias circuitry before making any connections to the Input connect...

Page 8: ...resistance on the input so that relatively long cable lengths can be used between the detector and the preamplifier without disrupting the system stability A Test Pulse connector with built in charge...

Page 9: ...t polarity is opposite from the input pulse polarity output pulse polarity is the same as bias polarity 2 4 CONNECTORS INPUT and BIAS Type SHV TEST PULSE E and T Type BNC POWER CABLE 10 ft 3 m captive...

Page 10: ...se for energy calibration can be accepted through the Test Pulse input connector on the 142IH without the use of an external charge terminator The test input of the preamplifier has an input impedance...

Page 11: ...n ORTEC 114 Preamplifier Power Supply Figure 4 1 is a simplified block diagram of the circuits in the 142IH Preamplifier The complete circuit is shown in schematic 142IH 0201 S1 included at the back o...

Page 12: ...amplifier Input 4 4 TIMING OUTPUT The T output connector provides an alternate path for the same output pulses that are furnished through the E output connector The intent is to provide both connectio...

Page 13: ...t shown for zero external capacity In any case the Input connector and capacitors when used should be completely shielded electrically A wrapping of aluminum foil around the Input connector or a shiel...

Page 14: ...d parameters By use of the above equation and with a knowledge of the noise of the preamplifier the noise width of the detector can be determined The significance of this noise width in evaluating the...

Reviews: