TECHNICAL SPECIFICATION FOR SURROSENSE
®
PROTECT
Size
2.75in x 1.18in x 0.47in (L x W x H)
Weight
0.7oz. (20g)
Power Source
160mAh Rechargeable LiPo battery
Frequency Band
ISM 2.4 Ghz, 2400 to 2483.5 MHz
Wireless Technology
Frequency Hopping Spread Spectrum
Number of Operation Modes
2 – ON / Sleep
Storage Temperature
Temperature -4 to 131 Fahrenheit / -20 to +55
Celsius / 95%humidity
Operation
Temperature -4 to 131 Fahrenheit / -20 to +55
Celsius / 95%humidity
Water Resistance
No. Do not submerge
SHOE POD INDICATOR LIGHTS
There are red, yellow, green and orange lights inside each shoe pod to indicate system status. The
meaning of the lights is as follows:
INDICATOR LIGHT
STATUS
Flashing green while plugged in
Charging.
Solid green while plugged in
Charging is complete.
All colors (Red, Yellow, Green,
Orange) flashing simultaneously
The system has been reset and is initializing.
Constant light of all colors (Red,
Green, Orange)
Sensor error.
Plug the shoe pod into charger, wait 30 seconds
then remove. If the error re-appears after a