755-00002 Rev P
LifePort Kidney Transporter Operator’s Manual
34
Guidance and Manufacturer’s Declaration — ELECTROMAGNETIC IMMUNITY
LifePort Kidney Transporter is intended for use in the electromagnetic environment specified below. The customer or the
user of LifePort Kidney Transporter should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic
Environment: Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70% UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70% UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of LifePort Kidney Transporter
requires continued operation
during mains power interruptions,
LifePort Kidney Transporter can
be lowered from the internal
battery.
Power frequency (50-60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
UT is the AC mains voltage prior to application of the test level.