E.2 Material Properties
Material POM (pins/blocks):
1,4 gr/cm³
Material PC (screws):
1,2 gr/cm³
Material Glass Fibre (bar):
1,8 gr/cm³
This product is not magnetically attracted, does not have an
effect on the susceptibility of the magnetic field, does not
create significant noise in the MR scan and it does not warm up
during scanning. This product is labelled as ‘MR safe’.
Warning: keep metal wrenches which might be supplied in the
package out of the MRI room.
E.3 Dosimetric Properties
Attenuation factor (± 0.15 %)
6 MV
2.67 %
15 MV
2.00 %
H
2
O equivalence
7.72 mm
Note: Use these numbers as a guidance only. Perform the
measurements again in your department to verify these results.
F.
MAINTENANCE AND WASSTE MANAGEMENT
This product can be cleaned and disinfected by means of soapy
water or an isopropanol based disinfectant, applied with a soft
cloth. If unsure about the cleaning fluid, do not use.
Never use
aerosol sprays, corrosive cleaning agents, solvents or abrasive
detergents.
Periodic checks of the product should be done to insure the
parts are not worn and require repair or replacement.
Do not
attempt to make repairs yourself
. Contact your distributor if
there are any questions or concerns.
The product can be disposed of with household waste.
G.
ADDITIONAL INFORMATION
For additional information such as distributor contact
information, product brochures, Safety Data Sheets and
regulatory information, please visit our website