INSTRUCTIONS FOR USE  

 

 
 

RAYCAST

®

 HIGH PRECISION 

INDEXING SYSTEM FOR 
iBEAM

®

-TYPE COUCH TOPS 

(MRI COMPATIBLE) 

 
Article No. : 

32126

 

A.

 

GENERAL PRODUCT INFORMATION 

The product referred to in this instructions is a fixation device to 
attach Raycast High Precision base plates to iBEAM® couch tops. 

B.

 

PRODUCT DESCRIPTION 

The  indexing  pins  described  in  these  instructions  are  fixation 
devices to position a 32124 Pelvicast

TM

 base plate and an HP PRO 

base plate on a couch with holes in its surface as indexing points 
(indexed  couch  tops  like  the  Medical  Intelligence  iBEAM

®

Tomotherapy HiArt

®

, or Brainlab Novalis

®

). 

C.

 

USAGE 

The  32124  Pelvicast

TM

  base  plate  has  4  holes  into  which  the 

indexing pins can be screwed. The HP PRO base plate has 4 holes 
on the  underside  of  the  board on  the  reinforced  sections.  The 
32126 set contains four of the described Indexing Pins that are 
screwed into these holes. The holes are positioned in such a way 
that the pins will index to the holes on the side of the couch top. 
 

 
The indexing pins can be mounted in 2 different positions: 
 
 

TYPE 1: UNLOCKED POSITION 

Normal position when the base plate is fully supported by the 
couch top. 

 

 

-

 

Turn the Indexing Pin with a clockwise motion into the holes 
situated on the bottom of the base plate. 

-

 

In  the  unlocked  position  the  largest  diameter  of  the  pin  is 
against the Base Plate surface (A). 

 

Remarks: 

-

 

Always use the entire thread length. 

-

 

Mount the pins before placing the base plate onto the couch. 
It cannot be done afterwards. 

-

 

Turn  manually.  Only  use  power  tools  if  absolutely  needed 
and only rotate counter clockwise. 

 

TYPE 2: LOCKED POSITION 

The base plate is locked for use in overhang position on the 
couch top. 

 

 

-

 

First position the base plate onto the couch. 

-

 

Turn the Indexing Pin with a clockwise motion into the 
holes situated on the bottom of the Base Plate. 

-

 

The largest diameter of the pins (collar) is opposite of the 
Base Plate surface (A). 

 

Remarks: 

-

 

Always use the entire thread length. 

-

 

Mount the pins after placing the base plate onto the couch. 

-

 

Turn manually. Only use power tools if absolutely needed 
and only rotate counter clockwise. 

Cannot be used on couch tops with blind indexing holes 
(Example: Tomotherapy 6-60091-01). 
 
 