9.2 DIAGRAM ELECTRIC 
 

 

 
9.3.- EXPLODED VIEW 

 
See page 12 & 13 
 
 
10.- TROUBLESHOOTING 
 

FAILURE OR MALFUNCTION 

CAUSE 

SOLUTION 

Equipment starts up but doesn’t 

aspirate. 

Regulation control is open. 

Close control. 

Suction tube wrongly inserted.. 

Insert correctly.

 

Cap improperly adjusted.

 

Adjust properly.

 

Equipment starts up but suction is not 

enough or inexistent.

 

Bacterial filter is saturated.

 

Change filter.

 

Failure continues.

 

Entry of liquid inside the pump.

 

Disassembly and cleaning of pump 

head by Technical Services.

 

Equipment doesn’t start up nor the   

luminous switch lights up.

 

E

xamine the fuse possible deterioration and 

the power plug.

 

Replace the fuse or the plug.

 

Failure continues. 

Contact Technical Services.

 

 
 
 

11.- TECHNICAL SPECIFICATION, EMC INFORMATION 
 

 

Guidance and manufacturer’s declaration – electromagnetic emissions

 

The system is intended for use in the electromagnetic environment specified below. 

The customer or the user of the equipment should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment 

– 

guidance 

RF emissions CISPR 11 

Group 1 

The equipment uses RF energy only for its internal function. Therefore, 
its RF emissions are very low and are not likely to cause any 
interference in nearby electronic equipment. 

RF emissions CISPR 11 

Class B 

The equipment is suitable for use in all establishments, including 
domestic establishments (not industrially) and those directly connected 
to the public low-voltage power supply network that supplies buildings 
used for domestic purposes. 

Harmonic emissions IEC61000-3-2 

Class A 

Voltage fluctuations/ flicker 
emissions IEC61000-3-3 

Complies 

 
 
 

Guidance and manufacturer’s declaration – electromagnetic immunity

 

The equipment is intended for use in the electromagnetic environment specified below. 

The customer or the user of the equipment should assure that it is used in such an environment. 

I

MMUNITY 

test 

IEC 60601 
test level 

Compliance level 

Electromagnetic environment 

 

guidance 

Electrostatic 
discharge (ESD) 
IEC61000-4-2 

±6kV contact 
±8kV air 

±6kV contact 
±8kV air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %. 

Electrical fast 
transient/burst 
IEC 61000-4-4 

±2kV for power supply lines 
±1KV for input/output lines 

±2kV for power supply 
lines 
 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge 
IEC 61000-4-5 

±1kV line(s) to line(s) 
±2kV line(s) to earth 

±1kV line(s) to line(s) 
±2kV line(s) to earth 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Summary of Contents for ASPIT

Page 1: ...are i ntre inere Maintenance and Functioning manual Mom 09 Edici n 12 Fecha edici n 2019 11 04 Tecnologia 2 Nave 7 08780 Palleja Barcelona Spain 34 93 334 01 12 FAX 34 93 440 25 64 comercial ordisi co...

Page 2: ...or directly with ORDISI S A ORDISI S A Tecnologia 2 08780 Palleja Barcelona Espa a Tel 34 93 334 01 12 Fax 34 93 440 25 64 e mail comercial ordisi com 2 MANUFACTURER S LIABILITIES O OR RD DI IS SI I...

Page 3: ...rflowing and penetration of liquid in the pump interior will thus be prevented In this case we advise you to remove the pump lathe head for to clean it by an Authorized Technical Service 2 Excessively...

Page 4: ...has the same indications The variation is that the inlet diameter of the plug is higher 8 5 mm 5 CLEANING AND STERILIZING 5 1 BODY Before undertaking any operation disconnect the equipment from electr...

Page 5: ...rements 340x150x300 Weight 4 600 Kg Power 105 VA Voltage 230V 50 Hz Fuses 2 of 2A type F Consult us for other voltage Equipment of I Class B type Average life 10 years 8 SPECIAL ACCESSORIES ON OFF ped...

Page 6: ...re not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The equipment is suitable for use in all establishments including domestic establishments not indus...

Page 7: ...continued operation during power mains interruptions it is recommended that the equipment be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000...

Page 8: ...P g 8 de 9 ASPIT...

Page 9: ...P g 9 de 9 ASPIT G...

Reviews: