The ambient temperature range of the Orbit X is
-20°C to + 45°C
The Orbit X shall be protected from contamin-
ation by liquid and dust; this is normally
achieved by the operator carrying the device on
their person.
The USB port may be used for charging the
equipment when in the non-hazardous area
only, using one of the following chargers
specifically supplied for use with the unit:
Deltaco part number USB-AC35M
(ambient for charging is 0ºC to 45ºC)
Ansmann part number 1001-0007
(ambient for charging is 0ºC to 45ºC)
GlobTek model number GT-41078-0505-USB
(ambient for charging is 0ºC to 45ºC)
GlobTek model number
GT-41078-0506-0.4-USB
(ambient for charging is 0ºC to 45ºC)
Glob Tek model number GT*-46101-*05*-USB
(ambient for charging is 0ºC to 40ºC)
Glob Tek model number GT*-46101-*06*-USB
(ambient for charging is 0ºC to 35ºC).
The maximum input voltage (Um) from the
charger between the lines is 5.8 V
.
The USB port is also used for data download.
The port has been assessed with a Um o
f 5.8
V
and shall be installed in accordance with
clause 12.2.1 of IEC 60079-14:2007/EN
60079-14:2008
Conditions For Use (ATEX & IECEx)