6
7
Initialization Time
Capnogram, displayed in less than 15 seconds, at
an ambient temperature of 25°C, full specifications
within 2 minutes
CO2 Measurement
Range
0 to 150 mmHg, 0 to 19.7%
CO2 Resolution
0.1 mmHg 0 to 69 mmHg
0.25 mmHg 70 to 150 mmHg
CO2 Accuracy
(
0
~
40) mmHg ± 2 mmHg
(41
~
70) mmHg ± 5% of reading
(71
~
100) mmHg ± 8% of reading
(101
~
150) mmHg ± 10% of reading
Above 80 breaths per minute ±12% of reading
* NOTE: Gas temperature at 25°C
Respiration Rate
Range
0 to 150 breaths per minute (bpm)
Respiration Rate
Accuracy
± 1 breath(s) per minute (bpm)
Voltage
Requirements
DC 5.0V ±5%
Temperature and
Humidity
Operating: 0°C to 40°C, 15% to 85% RH,
non-condensing
Storage: -10°C to 40°C, 15% to 85% RH,
non-condensing
Water Resistance
IPX2 – Splash-proof (sensor head only)