-21-
4.8
Save User Configuration to Flash
Camera can boot using a specific user configuration stored in its flash memory. At power up, the camera uses
any user configuration previously saved. If no user configuration is there, the camera boots using the default
factory configuration.
User configuration can be saved by calling command
“
SaveToFlashWithRebootAndWithoutAutoStart
”.
A second command “
SaveToFlashWithRebootAndWithAutoStart
” save user configuration but also force
live/acquisition start without requiring any frame grabber command. Beware that if frame grabber sends
Connection Reset command (Default CXP behavior), the stream will be stopped and
this “AutoStart” feature
won’t be used.
To boot with default camera configuration, user configuration must be deleted by calling XML command
“
FactoryFlash
”.
User configuration in camera flash is controlled by XML entries:
•
SaveToFlashWithRebootAndWithAutoStart
•
SaveToFlashWithRebootAndWithoutAutoStart
•
FactoryFlash