6.4Long-termparkingmaintenance
If you do not use A63.7080/81 for a long time, you should turn off the electron gun and
keep it in a shutdown state.
7. Transportation and storage
7.1Attention in hoisting and transportation
In the process of hoisting, handling and transportation, A63.7080/81 should prevent
inversion, rain and strong shock, prevent cold and insolation.
7.2Attention in storage condition and storage period
Before unpacking, A63.7080/81 should be stored in a dry, ventilated house without
corrosive gases, prevent inversion, rain, cold and insolation. The temperature in the
storage room is -10
℃
~ +40
℃
, and the relative humidity is no more than 80%. Since
the interior of the A63.7080/81 electronic optical column is in a vacuum state when it
leaves the factory, it is better to complete the installation within one month after the
arrival of the goods, so as to ensure the vacuum state inside the optical column and
avoid the accumulation of water and gas, which may cause pollution and even
corrosion to the sensitive parts inside and affect the best working state of the
instrument.
8. Unpacking and Inspection
8.1 Notes for unpacking
1. Before unpacking A63.7080/81, carefully check the laboratory conditions to ensure