Polarization-maintaining Fiber (PMF) 

 

 

 

 

Damaged: Replacement required 

Dirty: Cleaning required 

Clean: No cleaning required 

 

4.2.2

 

Cleaning Optical Connector Sides, Receptacles, Adaptors 

 
Dust or particles can adhere to the insides of receptacles and adaptors or the sides 
of the optical connector ferrule. Their presence can affect the alignment of the optical 
fiber connectors and increase connection loss. To properly clean optical connectors, 
receptacles,  and  adaptors  utilizing  a 

swab  and  alcohol

  follow  the  procedure  as 

follows: 
 
1.  Disable  the  optical  output  and  turn  off  unit  to  prevent  accidental  exposure  or 

damage to the optical connector by optical emission. 

2.  Moisten the swab by placing it on top of the alcohol dispenser and push down to 

saturate the swab. 

3.  For receptacles, adapters, or other connection points, insert the moistened swab 

and rotate the tip 1/2 turn clockwise and counter-clockwise 6 times while applying 
light but firm pressure.  

4.  For fiber connectors, rotate the tip of the  moistened swab 5 revolutions around 

the connector while applying light but firm pressure. 

5.  Visually inspect the end face of the connector with an optical microscope to verify 

cleanliness. Clean end-face as needed. 

 
 

5. 

Service and Support 

 

5.1  Service and Calibration 

 
Your  product  has  been  designed  to  provide  years  of  trouble-free  operation.  No 
internal service is required provided that the equipment is properly handled, operated 
and kept away from contamination. For any questions regarding the operation and 
performance of the product, please contact Optilab, LLC. 

Summary of Contents for EYDFA-33-PM-R

Page 1: ...THER THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE THE USER MUST READ THIS MANUAL BEFORE OPERATING THE PRODUCT OPERATIONS OTHER THAN THOSE DESCRIBED IN THIS MANUAL MAY RESULT...

Page 2: ...00 08 08 2018 Manual introduced Optilab LLC 2018 All rights reserved Information in this document is subject to change without notice All pictures and diagrams are for illustration purposes only and t...

Page 3: ...__________________6 3 2 Operation Instructions ______________________________________________________8 3 2 1Preparation Procedure ___________________________________________________8 3 2 2Start up Pro...

Page 4: ...report There MUST be an input signal present before turning on the optical output Do NOT use this product in HIGH PEAK PULSE POWER 10W applications unless with the consent from the manufacturer Faili...

Page 5: ...AL COMPONENTS CONNECTED TO THE OPTICAL OUTPUT MUST BE RATED FOR HIGH POWER USE It is highly recommended that these components be properly cooled to dissipate heat FOR CONTINUED PROTECTION AGAINST FIRE...

Page 6: ...erformance Standards for Light Emitting Products 5 US Department of Labor OSHA Publication 8 1 7 Guidelines for Laser Safety and Hazard Assessment 6 Laser Safety Equipment Laurin Publishing 7 Laser sa...

Page 7: ...and up to 33 dBm saturation power at 1550 nm It is a self contained turn key rackmount designed to supply high and stable gain across the specified range of the wavelength window 2 2 Features Up to 33...

Page 8: ...put Power Stability 0 2 dB over 8 hours Optical Input Output Fiber Type 8 125 NA 0 12 PM Optical Input Output Connector FC APC fast axis aligned to slow axis Mechanical Specifications Operating Temper...

Page 9: ...The Touch LCD display shows the parameters and optical settings of the EYDFA rackmount unit This is a touchscreen type display see section 3 2 3 for more information Output Power Enable Switch This sw...

Page 10: ...remote control and monitoring through a PC workstation Ventilation Fans The ventilation fans ensure proper ventilation inside the unit The back panel and the two sides of this equipment should be plac...

Page 11: ...off and make sure the key switch on the front panel is also off Interlock should be shorted using the provided accessory or connected to a compatible interlock device After plugging in the appropriat...

Page 12: ...ight a parameter to change it will turn yellow When the parameter is highlighted use the INC button to cycle through the options For the pump current adjust each digit then press SAVE to engage the ne...

Page 13: ...ry to short this port or connect it to a compatible interlock device If the interlock is open during normal operation the pump lasers will be turned off To re enable the pump laser output turn the out...

Page 14: ...the command afterwards Syntax is CMD XX case sensitive READ This command displays system status including the current settings of each pump and maximum current setting Example of Response Optilab LLC...

Page 15: ...input or optical input power is too small S Check whether optical input is present or check optical input power is correct according to the test report C Optical connectors are dirty S Disable optica...

Page 16: ...moist wipe removes dust particles oil and contaminants that may damage or blot the end face of the connector during connection The dry wipe removes residual alcohol that may be ignited by optical emis...

Page 17: ...posure or damage to the optical connector by optical emission 2 Moisten the swab by placing it on top of the alcohol dispenser and push down to saturate the swab 3 For receptacles adapters or other co...

Page 18: ...effect all repairs under terms of the warranty providing the Product is returned to Optilab LLC as specified in the warranty statement The purchaser shall bear the freight charges incurred in returnin...

Reviews: