Keratron
™
Onda Installation and operating manual
OPTIKON 2000
Cod. 161401EN
7-2
2015-03-27
Rev.C
Figure 7-2: Possibility of repeating the
test
Calibration and inspection of the motors
Once the test has been performed again, if the procedure does not reveal any
errors, the user may proceed by pressing the “Continue” button or waiting for the
software to proceed automatically after approx. 5 seconds (Figure 7-3).
Figure 7-3
: Repetition of the “Motor Diagnosis” test
was successful
If, instead, conditions exist that cause the test to indicate some problem, the
message will be the one shown in Figure 7-4. At this point it is possible for the
operator to repeat the test by means of the “Retry” button, or the “Continue” button
can be pressed to exit the window.
IMPORTANT:
If the “Motor Diagnosis” indicates some problem or error, the software will prevent
the use of the Topoaberrometry, Aberrometry and Accommodation functions for
reasons of safety.