Adjustment
25
During adjustment a progress bar is
displayed.
When completed, the display returns
to the measurement value display.
$GMXVW
㸰
7XUE7&'
6SDQ$GMXVWPHQW
2-point adjustment
Display current values of the 2 points as any adjustment values,
respectively.
Performing 2-point adjustment changes the offset and span based on the
adjusted values.
For 4-20 mA output (“6 4-20 mA Setting” P.15) and alarm output (“Alarm
output setting” P.17), the adjusted values are applied.
Measurement
item [Display]
Measurement item setting range
(display value/adjusted value)
Factory default
setting
Turbidity
[Turb.]
[Point 1]: 0.0 to (2nd point value - 20)
0.0
[Point 2]: (1st point value + 20) to 3000 3000
Caution
If offset, span, or 2-point adjustment has already been performed,
be sure to initialize the adjustment before performing the 2-point
adjustment.
[Point 1]
Adjusted value
Measurement value
Current display
value
Adjusted
value
Display value
after 2-point
adjustment
Current display
value
[Point 2]
Press
.
The measurement item selection display is displayed.
1