Troubleshooting 

Occasionally, the Hand Control screen may appear blank or locked in the “Waiting for Data” screen for 
too long.  If the screen is blank or the “Waiting for Data” message persists, first try simply unplugging 
and re-plugging the coiled handset cable in the bottom of the hand control.  If this does not solve the 
issue try adding a ferrite core choke or contact Optec support.  When changes are made in FocusLynx 
Commander set up dialogs, the cable may need to be unplugged and then re-plugged to re-establish 
communication between the hub firmware and FLHC firmware.  

Contact Optec Technical Support (

[email protected]

with any other issues concerning the 

FocusLynx Hand Control option. 

Upgrades 

If the FocusLynx Hand Control firmware needs to be updated, a PIC chip replacement must be ordered.  
Specify Optec #19695-FW to order a hand control firmware update.  The firmware is provided in a 28-
pin DIP (dual in-line package) programmable integrated circuit 
with the version number marked on the IC.   

First unplug the coiled cable from the hand control.  Remove the 
FLHC back cover to expose the circuit board and IC socket.  Note 
the position of the notch indicating Pin 1.  Gently pry the existing 
firmware chip up and out of the socket and replace with the new 
firmware PIC chip.  Ensure that Pin 1 is in the same orientation as 
the original.   Re-attach the back cover and coiled cable.   Connect 
to the hub and watch for the firmware version number to appear.  

 

 

 

Fine Focus Knob Upgrade 

Owners of the first generation FocusLynx Hand Controllers may wish to upgrade to the second 
generation which includes the Fine Focus Knob.  This second generation upgrade requires returning the 
hand control to Optec for addition of the rotary encoder and associated circuitry.  Optec offers our 
service item #17696 for the upgrade which includes the #19695-FW firmware update to the most 
current version.  Contact Optec for details and pricing.   