FluxGage User Manual
36
8
Calibration
8.1
Calibration of the spectrum measurement and the total flux
measurement
For this calibration an LED calibration source is needed. A tungsten lamp cannot be used
because its IR radiation is detected by the solar panels but not by the spectrometer.
Follow these steps:
To calibrate the FluxGage click ‘Calibration’ on the top banner.
Position the calibration source over the FluxGage.
In the setting area, enter the total luminous flux of the calibration source in the
luminous flux box.
Click ‘Open Light File’ and select the spectral calibration file of the calibration source.