9-44
INTEGRA Operations Revision A
7/26/00
20-0200-231 (CR), 20-0200-232 (Std), 20-0200-251 (CD)
.
Wafer EFFECT (Brush Conditioning)
The third set of buttons (see illustration on following page) are
specifically designed to prepare (chemically dose) the brushes in both
Brush Stations either:
1)
before
processing the first wafer
after
having sat idle (1
st
WAFER)
or
2)
before
processing all (every) wafer (ALL WAFERS).
During continuous wafer cycling, the brush pH stays constant when
HPD PORs are used. Use the 1
st
WAFER wafer dosing options to
correct the process pH when continuous wafer cycling is interrupted.
When wafer cycling is interrupted and DI LOW FLOW is ON, then the
brush pH levels are neutralized. When normal processing resumes,
the 1
st
WAFER “effect” option doses the brushes to re-establish brush
pH, thus ensuring wafer-to-wafer process uniformity.
Select the DOSE MODE for Brush Station #1 and Brush Station #2 to
select either the 1
st
WAFER or ALL WAFERS mode of operation.
The DOSE MODE is a toggle that displays the current mode on the
DOSE MODE button. 1
st
WAFER mode will
only
dose the brushes
with chemical
before
processing the
first
wafer. ALL WAFERS mode
doses the brushes with chemical before processing for
every
wafer,
not just the first.
STANDARD
Summary of Contents for Synergy Integra
Page 8: ......
Page 24: ...1 4 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Page 38: ...2 14 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Page 222: ...9 62 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Page 314: ...10 76 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...