Battery Information 

The logger contains a 3V CR2032 battery (HRB-TEMP). Expected 
battery life varies based on the ambient temperature where 
the logger is deployed, the logging or sampling interval, 
frequency of offloading to the computer, number of channels 
that are active, if burst or statistics logging modes are active, 
and battery performance. A new battery typically lasts 1 year 
with logging intervals greater than 1 minute. Deployments in 
extremely cold or hot temperatures, a logging interval faster 
than 1 minute, or a sampling interval faster than 15 seconds 
can impact battery life. Estimates are not guaranteed due to 
uncertainties in initial battery conditions and operating 
environment. 

The logger can also be powered by the USB cable when the 
remaining battery voltage is too low for it to continue logging. 
Connect the logger to the computer, click the Readout button 
on the toolbar, and save the data as prompted. Replace the 
battery before launching the logger again. 

To replace the battery: 

1.

 

Holding the logger with the LCD screen facing up, pull the 
battery tray out of the logger housing. 

 

2.

 

Remove the old battery from the tray. 

3.

 

Place the new battery in the tray with the positive side 
facing down. 

 

4.

 

With the LCD screen still facing up, slide the tray back into 
the logger. The LCD should display “HOBO” briefly after 
correctly installing the battery. 

 

WARNING:

 Do not cut open, incinerate, heat above 85°C 

(185°F), or recharge the lithium battery. The battery may 
explode if the logger is exposed to extreme heat or conditions 
that could damage or destroy the battery case. Do not dispose 
of the logger or battery in fire. Do not expose the contents of 
the battery to water. Dispose of the battery according to local 
regulations for lithium batteries. 

HOBOware provides the option of recording the current battery 
voltage at each logging interval, which is disabled by default. 
Recording battery life at each logging interval takes up memory 
and therefore reduces logging duration. It is recommended you 
only record battery voltage for diagnostic purposes. 

CR2032 battery being 
placed in the tray, 
positive side facing 
down 

Battery tray removed 
from logger 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for HOBO UX100-014M

Page 1: ...F thermocouple probe accuracy 0 08 C 0 15 F S 50 to 1 720 C 58 to 3 128 F 2 2 C 3 96 F thermocouple probe accuracy 0 08 C 0 15 F B 550 to 1 820 C 1 022 to 3 308 F 2 5 C 4 5 F thermocouple probe accura...

Page 2: ...button for 1 second to record an internal event see Recording Internal Logger Events or to turn the LCD screen on if the option to turn off the LCD has been enabled see Setting up the Logger Battery T...

Page 3: ...atistics in HOBOware Press the Alarm Stats button for 1 second to cycle through each of the available statistics and then back to the current temperature or to the alarm value if applicable This is th...

Page 4: ...g the oldest This option is not available if Burst is selected for the Logging Mode Push Button Logging will end once you press the Start Stop logging button for 3 seconds Note that if you also choose...

Page 5: ...e course of the deployment is equal to the selected duration If you select Consecutive then the alarm will trip when the time the sensor is continuously out of range is equal to the selected duration...

Page 6: ...ogger window Click the Edit button next to Logging Mode in the Launch Logger window to make additional changes Notes Burst logging is available on the thermocouple channel only it is not available for...

Page 7: ...OBOware or connect it to a HOBO U Shuttle U DT 1 firmware version 1 18m030 or higher and then offload the data files from the U Shuttle to HOBOware Refer to the HOBOware Help for details Recording Int...

Page 8: ...r to the computer click the Readout button on the toolbar and save the data as prompted Replace the battery before launching the logger again To replace the battery 1 Holding the logger with the LCD s...

Reviews: