background image

BioShift Pass-Through UV-C Germicidal 

Chamber - Instruction Manual

This manual provides user guidelines and additional operation 

instructions for the Large BioShift Pass-Through UV-C Germicidal Chamber.

(Model 24-0201, 12NC: 929002100323)

For service or parts call your 

sales representative, visit our website: 

www.once.lighting/uv-c

Or contact us at: 

Once Inc.

United States

15255 23rd Ave N

Plymouth, MN 55447

+1 (763) 381-5621

[email protected] 

Germany

Am Südfeld 7
49377 Vechta

+49 (0) 4441 8898-10

[email protected] 

Poland

Wloclawska 167

87-100 Tourn

+48 (56) 699-5330

[email protected]  

Summary of Contents for 24-0201

Page 1: ...hamber Model 24 0201 12NC 929002100323 For service or parts call your sales representative visit our website www once lighting uv c Or contact us at Once Inc United States 15255 23rd Ave N Plymouth MN...

Page 2: ...Screen Components 8 Settings Screen Components 9 Programming 10 Cycle Time 10 Sleep Timer 10 Maintenance 11 13 Alert Notifications Errors 11 Check Lamp Life 12 Order Replacement Lamps and Ballasts 12...

Page 3: ...nation through the import of pathogens on everyday items like cell phones eye glasses tools and protection gear Verified to inactivate common pathogens by a nationally recognized testing laboratory sp...

Page 4: ...decreasing linearly to 50 relative humidity at 104 F 40 C Note Under 41 F 5 C temperature must run warm up cycle before disinfecting See Operating section Storage 20 F 28 C to 140 F 60 C temperature...

Page 5: ...n Overexposure to direct or reflected UV C light can cause severe damage to the eyes and skin Never look into an illuminated UV C lamp without proper eye protection The chamber windows provide a safe...

Page 6: ...dirty start button side door is opened or a new cycle is started If a cycle is interrupted it will start over at the beginning The dirty door will remain locked anytime the clean door is open Pressing...

Page 7: ...t breaker in the building shall be provided as a mains disconnection device 120VAC power systems shall use a circuit breaker with a minimum current rating of 15A 230VAC power systems shall use a circu...

Page 8: ...EEN DISPLAY The touchscreen display is located on the clean side of the BioShift chamber Home Screen Components 1 Cycle Time Display 2 Chamber Start Stop 3 Chamber Status 4 Settings 5 Maintenance 1 3...

Page 9: ...ed under the Maintenance Button Maintenance Screen Components 1 View Lamp Life Remaining 2 Reset Lamp Life 3 View Lamp Replacement Info 4 Run 100mJ cm Dose Test 5 Run 250mJ cm Dose Test 6 Run 1000mJ c...

Page 10: ...9 1 4 2 3 TOUCHSCREEN DISPLAY Located under the Home Screen Settings button Settings Screen Components 1 View Set Lamp Cycle 2 Set Lamp Cycle 3 Set Sleep Time 4 Expired Lamp Life 5 Contact ONCE 5...

Page 11: ...sired lamp life cycle has been chosen select the check mark to set PROGRAMMING ITEMS TO NOTE Controller cannot be programmed if EMERGENCY STOP button s are pressed in Pressing EMERGENCY STOP button s...

Page 12: ...hen one of the following occurs UV C lamps have reached maximum cycle hours Replacement lamps have been installed and Lamp Life needs to be reset Simply follow the red exclamation point to the source...

Page 13: ...052 929002100325 Changing Lamps and Ballasts Remove all shelves and de energize the chamber To replace lamps simply twist the lamp and pull out from its housing Slide the new lamp in and twist to lock...

Page 14: ...ioShift chamber to ensure proper disinfection Every 50 cycles or once per month whichever comes first the operator shall verify that the UV C dose exceeds a minimum of 100 mJ cm2 in a short programmed...

Page 15: ...nitation ADDITIONAL OPERATING FUNCTIONS BioShift chamber doors shall not be opened during radiation cycle BioShift chamber doors cannot be open at the same time under normal operation In case of emerg...

Page 16: ...other acts of God 4 Corrosion due to use of improper solvents or detergents 5 Repair or modification performed by anyone other than ONCE or authorized person 6 Use of spare parts supplied by anyone ot...

Page 17: ...aus s par une catastrophe ou d autres v nements de force majeure 4 Corrosion due l utilisation de solvants ou de d tergents inappropri s 5 R paration ou modification effectu e par une personne autre q...

Page 18: ...ms of failure and 4 Total hours of use and documented cleaning record Upon request ONCE representatives shall be granted access to the defective Product system or application for verification of non c...

Page 19: ...bearing the ONCE name or sub brands ONCE makes no warranty of any kind express or implied including without limitation any warranty of merchantability or fitness for a particular purpose but can make...

Page 20: ...have entered into with any party in relation to the supply provision donation and or otherwise making available of BioShift Chambers ONCE explicitly excludes any and all liability towards any party i...

Page 21: ...es des informations suppl mentaires peuvent tre requises sur demande 1 Num ro de mod le du produit d faillant 2 Date d installation et informations sur la facture 3 Description d taill e du probl me n...

Page 22: ...imit e la p riode de garantie est de deux 2 ans compter de la date de livraison sous r serve d un nettoyage document des produits au moins une fois tous les trois 3 mois au cours de la p riode de deux...

Page 23: ...autre EPI de qualit pour dispositifs m dicaux ONCE d cline toute responsabilit envers toute partie si des chambres BioShift sont utilis es pour d sinfecter les surfaces de dispositifs m dicaux et ou d...

Page 24: ...ice or parts call your sales representative visit our website www once lighting uv c Or contact us at Once Inc Conforms to cETLus LVD CE ETL CDR IEC EN ANSI UL CUL 61010 1 IEC EN ANSI UL CUL 61010 2 0...

Reviews: