Setting the System
262
ZW User's Manual
Calibrating Sensor Head
Operating procedure
When Sensor Head calibration fails
If the Sensor Head is calibrated in an inappropriate environment, an error is displayed on the main display. If this
happens, press the
key
to return to the previous screen and try again. If an error continues even after
calibrating the Sensor Head in an appropriate environment, the fiber connector on the Sensor Head or Sensor
Controller may be stained. Clean the fiber cable or fiber connector referring to p.35.
Steps
Key operation
Display
Description
1
Press and hold the
key for two seconds to enter the FUN mode.
(For details on the functions of the Mode switching key, see p.220.)
2
Press the
/
keys to select "SYSTEM" and press the
key.
3
Press the
/
keys to select "H.CALIB", then press the
key.
4
"OK/CAN" is displayed on the sub-display.
Press the
key to execute calibration or the
key to cancel.
5
Press the
key for two seconds to enter the RUN mode.
Important