Connecting by No-protocol Communications
354
ZW User's Manual
Setting Serial Data Output
Setting the data output destination p.231
Operating procedure
Steps
Key operation
Display
Description
1
Press and hold the
key for two seconds to enter the FUN mode.
2
Press the
/
keys to select "I/O" and press the
key.
3
Press the
/
keys to select "COM.OUT" and press the
key.
4
Select the setting item:
OUTPUT: Data output destination
FORMAT: Output data type
INT.NUM: Digits of integer
DEC.NUM: Digits of decimal
ZEROSP: Zero suprression
FIELD: Field separator
RECORD: Record separator
In this example, press the
/
keys, select "OUTPUT" and press
the
key.
5
Automatically Following the Range of Measurement Area2
The current setting value is displayed on the sub-display.
Press
/
key to enter editing mode and the sub-display blinks.
Select the data length, and press the
key.
6
Press the
key to return to the previous menu.
7
Select the save count.
Press the
/
keys to select "FORMAT" and press the
key.
8
Repeat steps 5 to 7 to set other items.
9
Press and hold the
key for two seconds to enter the RUN mode.
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......