Setting Menus
ZR-RX70 User’s Manual
3
SETTINGS AN
D
MEASUREMENT
91
(9)-7 Level Settings
The Level Settings are the same as the Trigger Level Settings except that the Combinations include only edge
operations (and no level operation) and the Alarm Output can be set to Rising (from Cleared to Generated) and
Falling (from Generated to Cleared).
See the “Trigger Level Settings/Alarm Level Settings” p. 72. This section describes only how to specify
the Alarm Output.
(9)-8 Next Search
Moves the cursor to a position after the current cursor position where the search conditions are met (Set the
search conditions in Section (9)-7 Level Settings).
(9)-9 Prev. Search
Moves the cursor to a position before the current cursor position where the search conditions are met (Set the
search conditions in Section (9)-7 Level Settings).
Setting
Selections available
(1) Alarm CH
Sets the alarm output to be used for search.
Off: No alarm output is used for search.
Output 1 to Output 4: The specified output is used for search.
(2) Mode
Sets the search mode.
↑
Rising : Finds an edge at which the alarm output changes from Cleared to Generated.
↓
Falling : Finds an edge at which the alarm output changes from Generated to Cleared.
Summary of Contents for ZR-RX70
Page 1: ...User s Manual Portable Multi Logger ZR RX70 Cat No Z283 E1 02...
Page 12: ...10 ZR RX70 User s Manual MEMO...
Page 16: ...14 ZR RX70 User s Manual MEMO...
Page 108: ...WEB Server Function 106 ZR RX70 User s Manual MEMO...
Page 118: ...External Dimensions 116 ZR RX70 User s Manual MEMO...
Page 119: ...5 APPENDIX APPENDIX Error Messages and Countermeasures 118 Index 121 Revision History 124...
Page 122: ...Error Messages and Countermeasures 120 ZR RX70 User s Manual MEMO...
Page 125: ...Index ZR RX70 User s Manual 123 5 APPENDIX MEMO...