background image

Caution

Please read the instructions for use.

Do not reuse

Manufacturer

BF type applied part

Battery Recycling

Paper Recycling

The CE mark and Notified 

Body Registration Numbers, 

the requirement of Annex II 

from Medical Device 

Directive 93/42/EEC are 

met.

Indicates this device is subject to the 

Waste Electrical and Electronic

Equipment Directive in the European 

Union. To protect the environment, 

dispose of useless device at appropriate 

collection sites according to national or 

local regulations.

Authorized representative 

in the European 

community.

Guidance and manufacturer’s declaration – electromagnetic emissions

The TH8xyz series is intended for use in the electromagnetic environment specified below. The customer or the user of the 

TH8xyz series should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions

CISPR 11

RF emissions

CISPR 11

Group 1

Class B

The TH8xyz series uses RF energy only for its internal function. Therefore, 

its RF emissions are very low and are not likely to cause any interference 

in nearby electronic equipment. 

Harmonic emissions

IEC 61000-3-2

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Not applicable

Not applicable

The TH8xyz series is suitable for use in all establishments, including domestic 

establishments and those directly connected to the public low-voltage power 

supply network that supplies buildings used for domestic purposes.

Symbol Descriptions

Guidance and manufacturer’s declaration – electromagnetic immunity

The TH8xyz series is intended for use in the electromagnetic environment specified below. The customer or the user of the 

TH8xyz series should assure that it is used in such an environment.

Immunity test

Compliance 

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

Not 

applicable

3  V/m

IEC 60601 

test level

3 Vrms

150 kHz 

to 80 MHz

3 V/m

80 MHz 

to 2,5 GHz

Portable and mobile RF communications equipment should be used no closer to 

any part of the TH8xyz series, including cables, than the recommended separation 

distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

where 

P

 is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer and 

d

 is the recommended separation 

distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic 

site survey, a should be less than the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with the following 

symbol:

80 MHz to 800 MHz

800 MHz to 2,5 GHz

d = 1,2

d = 1,2

d = 2,3

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people.

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 
    amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
    electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
    field strength in the location in which the TH8xyz series is used exceeds the applicable RF compliance level above, the TH8xyz 
    series should be observed to verify normal operation. If abnormal performance is observed, additional measures may be 
    necessary, such as reorienting or relocating the TH8xyz series.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Guidance and manufacturer’s declaration – electromagnetic immunity

The TH8xyz series is intended for use in the electromagnetic environment specified below. The customer or the user of the 

TH8xyz series should assure that it is used in such an environment.

Immunity test

Compliance level

Electromagnetic environment – guidance

Electrostatic          

discharge (ESD)

IEC 61000-4-2

Electrical fast

transient/burst

IEC 61000-4-4

6 kV contact

8 kV air

6 kV contact

8 kV air

Not applicable

Floors should be wood, concrete or ceramic tile. If floors are covered 

with synthetic material, the relative humidity should be at least 30 %.

Mains power quality should be that of a typical commercial or 

hospital environment.

Mains power quality should be that of a typical commercial or 

hospital environment.

Surge

IEC 61000-4-5

Not applicable

IEC 60601 

test level

2 kV for power 

supply lines

1 kV for input/

output lines

1 kV line(s) to 

   line(s)

2 kV line(s) to 

   earth

interruptions 

and voltage 

variations on 

power supply 

input lines

IEC 61000-4-11

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

<5 % UT

(>95 % dip in UT)

for 5 sec

Mains power quality should be that of a typical commercial or 

hospital environment. If the user of the TH8xyz series requires 

continued operation during power mains interruptions, it is 

recommended that the TH8xyz series be powered from an 

uninterruptible power supply or a battery.

Power frequency magnetic fields should be at levels characteristic of 

a typical location in a typical commercial or hospital environment.

Power frequency 

(50/60 Hz)

magnetic field

IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.

Not applicable

3  A/m

3  A/m

Recommended separation distances between portable and 

mobile RF communications equipment and the ME EQUIPMENT or ME SYSTEM

The TH8xyz series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 

The customer or the user of the TH8xyz series can help prevent electromagnetic interference by maintaining a minimum 

distance between portable and mobile RF communications equipment (transmitters) and the TH8xyz series as recommended 

below, according to the maximum output power of the communications equipment.

Rated maximum output 

power of transmitter

                W

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in meters (m) can 

be estimated using the equation applicable to the frequency of the transmitter, where 

P

 is the maximum output power rating 

of the transmitter in watts (W) according to the transmitter manufacturer. 

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects and people.

Separation distance according to frequency of transmitter

                                          m

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

80 MHz to 800 MHz

150 kHz to 80 MHz

800 MHz to 2,5 GHz

d = 1,2

d = 1,2

d = 2,3

Reviews: