Technical data
Product Description:
Compressor Nebulizer
Model:
NE-C802 (NE-C802-UK)
Rating (AC adapter):
100-240V ~ 50/60Hz 15VA
Rating (Compressor
Nebulizer):
6V 0.6A
Operating Temperature/
Humidity:
+10°C to +40°C / 30% to 85%RH
Storage and Transport
Temperature/Humidity/
Air Pressure:
-20°C to +60°C / 10% to 95%RH /
700 - 1060hPa
Weight:
Approx. 190g (compressor only)
Dimensions:
Approx. 85 (W) × 43 (H) ×115 (D) mm
(compressor only)
Contents:
Compressor, Nebulizer Kit, Air Tube (PVC,
100cm), Mouthpiece, Nosepiece, Adult Mask
(PVC), Child Mask (PVC), 5 pcs Spare Air
Filters, AC Adapter, Storage Bag, Instruction
Manual, Warranty Card.
Classification:
Class ll equipment, Type BF applied part
= Class ll
equipment
= Type BF
applied part
Read the instruction
manual carefully
= No operation
= Power on
Technical data for the OMRON NE-C802 Compressor with the Nebulizer kit:
Particle Size:
**MMAD Approx. 3
μ
m
MMAD = Mass Median Aerodynamic Diameter
Medication Tank Capacity:
10ml maximum
Appropriate Medication
Quantities:
2ml minimum - 10ml maximum
Sound:
*Less than 45dB
Nebulization Rate:
*Approx. 0.25ml/min (by weight loss)
Aerosol Output:
**0.56ml (2ml, 1%NaF)
Aerosol Output Rate:
**0.07ml/min (2ml, 1%NaF)
Result of cascade impactor measurements for particle size**
100
90
80
70
60
50
40
30
20
10
0
1
10
100
0.1
Cumulative % particle mass of sodium fluoride undersize
Cumulative Undersize%
Particle Size (
μ
m)
Individual
Mean
*
Measured by OMRON HEALTHCARE Co., Ltd. Measured with saline at ambient
temperature of 23°C and humidity of 40%. It may vary with the type of medication.
**
Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009.
Notes:
• Subject to technical modification without prior notice.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan.
• The device may not work if the temperature and voltage conditions are different to
those defined in the specifications.
• The device fulfils the provisions of the EC directive 93/42/EEC (Medical Device
Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory
therapy equipment - Part1: Nebulizing systems and their components.
• Performance may vary with drugs such as suspensions or high viscosity. See drug
supplier’s data sheet for further details.
• See website of OMRON HEALTHCARE EUROPE to update technical information.
URL: www.omron-healthcare.com
Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is connected to the air connector on the
compressor.
2) Remove the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the
moisture.
Changing the air filter
If the air filter has changed colour, or has been used on average for more
than 60 days, replace it with a new one.
1.
Pull the air filter cover off the compressor.
2.
Change the air filter.
3.
Put the air filter cover back in place.
Notes:
• Use only OMRON air filters designed for this device. Do not operate without air filter.
• Do not attempt to wash or clean the air filter. If the air filter becomes wet, replace it.
Damp air filters can cause blockages.
• There is no front or back orientation for the air filters.
• Check that air filters are clean and free of dust before inserting them.
• Do not boil the air filter cover.
Cleaning and disinfecting
Cleaning
Clean the parts after each use to remove residual medication. This will avoid
inefficient nebulization or infection.
Firstly, make sure that the power plug unplugged from the power outlet.
■
Nebulizer Kit (Inhalation top, Accessory Adapter, Medication tank),
Mask (PVC), Mouthpiece, Nosepiece
Wash them in warm water and mild detergent. Rinse them thoroughly with
clean hot tap water and allow to air dry in a clean place.
Note
: Nebulizer kit should be replaced after using for one year. Replace ONLY with
nebulizer kit designed for NE-C802.
■
Compressor, Air Tube
Wipe clean with a soft cloth moistened with water or mild detergent.
■
Vaporizer Head
Wash it with running water.
Disinfecting
Disinfect the parts once a week. If the parts are heavily stained, replace
them with new ones.
Use one of the following methods:
To select the method, refer to the table below.
A
. Use a commercially available disinfectant. Follow the instructions
provided by the disinfectant manufacturer.
1) Submerge the parts in the disinfectant for the specified period.
2) Rinse them with clean hot tap water and allow to air dry in a clean place.
Note
: Never clean with benzene, thinner or a flammable chemical.
B.
Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake
off excess water and allow to air dry in a clean
environment.
Note
: Do not boil Air tube, Mask (PVC), Air filter and Air filter
cover.
Caution: Handling the Vaporizer Head
• Always wash the vaporizer head after each use.
• Do not use a brush or pin to clean the vaporizer head.
• When disinfecting the parts by boiling, be sure to boil them in plenty of water.
• Do not boil vaporizer head together with objects other than applicable nebulizer
accessories.
Use the table below as a guideline to select a method for disinfecting.
O: applicable ×: not applicable
Parts
Mouthpiece
Nosepiece
Nebulizer Kit
Air Filter
Air Filter Cover
Adult Mask (PVC)
Child Mask (PVC)
Air T
ube
(PVC, 100cm)
Adult Mask (SEBS) (optional)
Child Mask (SEBS) (optional)
Materials
PP
PP
PP
(V
aporizer Head: PC)
Polyester
ABS
Mask: PVC
(Phthalate free)
Band: Rubber
(non-latex)
PVC
(Phthalate free)
Mask: SEBS
Band: Rubber
(non-latex)
Adapter: PP
Boiling
O
O
O
×
×
×
×
O
Alcohol
O
O
O
×
O
O
×
O
Disinfecting
ethanol
Sodium
hypochlorite
O
O
O
×
O
O
×
O
Milton*
(0.1%)
Quaternary
ammonium
O
O
O
×
O
O
×
O
Osvan*
(0.1%)
Chlorhexidine
O
O
O
×
O
O
×
O
Hibitane*
(0.5%)
Amphoteric
Surfactant
O
O
O
×
O
O
×
O
Tego*
(0.2%)
*
an example of commercially available disinfectant.
How to use
1.
Make sure that the power switch is in the off (
) position.
2.
Connect the AC adapter to the compressor and plug the power plug into
a power outlet.
3.
Remove the inhalation top together with
the accessory adapter and vaporizer
head from the medication tank.
4.
Add the correct amount of prescribed
medication to the medication tank.
Reattach the vaporizer head to the
medication tank.
5.
Put the inhalation top and accessory
adapter back inside the medication tank.
6.
Attach the mask, mouthpiece or
nosepiece to the nebulizer kit tightly.
Note
:
Make sure an arrow mark on the nebulizer
kit is facing toward the patient. Otherwise,
medication may be sprayed on patient’s skin
or clothes.
7.
Attach the air tube.
Push the air tube plugs firmly
into the air tube connectors.
8.
Hold the nebulizer kit as
indicated on the right.
Follow the instructions of your
doctor or respiratory therapist.
Exhale as normal through
mouthpiece, mask or through
the nose.
Caution:
Do not tilt the nebulizer kit at an
angle of greater than 45 degrees
in all directions. Medication may
flow into the mouth or it may result
in not nebulizing effectively.
9.
Press the power switch to the on (
) position. As the compressor
starts, nebulization begins and aerosol is generated.
Inhale the medication.
10.
When treatment is complete, turn the power off. Unplug the AC adapter
from the power outlet and disconnect it from the compressor.
Note
:
Securely hold and press the power plug
down to disconnect the AC adapter.
Caution:
Do not pull the power cord. Damage
may be caused.
Troubleshooting
Check the following if the device should fail during operation.
Problem
Cause
Remedy
Nothing happens when the
power switch is pressed.
Is the AC adapter connected
correctly to a power outlet and
the compressor?
Check that the AC adapter is
connected to a power outlet and
the compressor. Unplug then
reinsert the plug if necessary.
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication tank.
Is there too much/little
medication in the medication
tank?
Is the vaporizer head missing
or not assembled correctly?
Attach the vaporizer head
correctly.
Is the nebulizer kit assembled
correctly?
Assemble the nebulizer kit
correctly.
Is the nozzle blocked?
Make sure that the nozzle is free
of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer kit is
not tilted at an angle of more than
45 degrees.
Is the air tube connected
correctly?
Make sure that the air tube
is correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube does
not contain kinks.
Is the air tube blocked?
Make sure that the air tube is free
of blockages.
Is the air filter dirty?
Replace the air filter with a new
one.
The compressor is
abnormally loud.
Is the air filter cover attached
correctly?
Attach the air filter cover correctly.
The compressor is very hot.
Is the compressor covered?
Do not cover the compressor with
any type of cover during use.
Note:
If the suggested remedy does not solve the problem, do not try to repair the device
- no parts of the device are user serviceable. Return the device to an authorized
OMRON retail outlet or distributor.