Intended use
Medical Purpose

This product is intended to be used for inhaling medication for respiratory 

disorders.

Intended User

• Legally certified medical experts, such as doctor, nurse and therapist.

• Caregiver or patient under the guidance of qualified medical experts for 

home treatment.

• The user should also be capable of understanding general operation of 

NE-C101 and the content of instruction manual.

Intended Patients

This product should not be used on patients, who are unconscious or are 

not breathing spontaneously.

Environment

This product is intended for use in a medical facility, such as Hospital, clinic 

and doctor’s office, and in a room of general household.

Durable period

Durable periods are as follows, provided the product is used to nebulize 

saline 3 times a day for 10 minutes each time at room

temperature (23°C).
Durable period may vary depending on usage environment.
Frequent use of the product may shorten the durable period.
 

Compressor (Main unit) 

5 years

 

Nebulizer Kit 

1 year

 

Mouthpiece 

1 year

 

Nosepiece 

1 year

 

Air Tube (PVC) 

1 year

 

Air Filter 

60 days

 

Mask (PVC) 

1 year

Precautions for use

Warnings and cautions described in the instruction manual should be 

observed.

NE-C101-AP_C_M03_180222.pdf

Important safety instructions

Read all the information in the instruction manual and any other literature included in 

the box before using the device.

 

Warning:

Indicates a potentially hazardous situation which, if not avoided, could result in 

serious injury.

(Usage)

•  For type, dose, and regime of medication follow the instructions of your doctor or 

respiratory therapist.

•  If you experience any allergic reactions or other difficulties during use, stop using the 

device immediately and consult your doctor.

•  Do not use only water in the nebulizer for inhaling purposes.

•  Keep the device out of reach of unsupervised infants and children. The device may 

contain small pieces that can be swallowed.

•  Do not store the air tube while there is moisture or medication remaining inside.

•  Do not use or store the device where it may be exposed to noxious fumes or volatile 

substances.

•  Do not use the device where it may be exposed to flammable gas.

•  Do not cover the compressor with a blanket, towel, or any other type of cover during 

use. The compressor could get hot and there is a possible risk of a burn if touched.

•  Always dispose of any remaining medication after use, and use fresh medication each time.

•  Do not use in anaesthetic or ventilator breathing circuits.

•  Do not block the ventilation slots. Never place the device where the ventilation slots 

may be obstructed during operation. The compressor could get hot and there is a 

possible risk of a burn if touched.

•  Store the parts in a clean location to avoid infection.

(Risk of electrical shock)

•  Never plug or unplug the device with wet hands.

•  The compressor and power cord are not waterproof. Do not spill water or other liquids 

on these parts. If liquid does spill on these parts, immediately unplug the power cord 

and wipe off the liquid with gauze or other absorbent material.  

•  Do not immerse the compressor in water or other liquid.

•  Do not use or store the device in humid locations.

•  Do not operate the device with a damaged power cord or plug.

•  Keep the power cord away from heated surfaces.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these 

rules may result in damage, inefficient nebulization or infection. For the instructions, 

refer to “Cleaning and disinfecting” section.

•  Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using 

them when as follows:

- the first time after purchase

- if the device has not been used for a long period of time

- if more than one person uses the same device

•  Be sure to wash or wipe the parts after use, and ensure that they are thoroughly 

disinfected and dried, and stored in a clean location.

•  Do not leave the cleaning solution to dry on the parts.

 

Caution:

Indicates a potentially hazardous situation which if not avoided, may result in 

minor or moderate injury, or physical damage.

(Usage)

•  In order to prevent strangulation by the power cord or air tube, or injuries caused by 

mishandling, please provide close supervision when this device is being used by, on 

or near children or invalids.

•  Make sure that the parts are attached correctly.

•  Make sure that the air filter is correctly attached.

•  Make sure that the air filter is clean. If the air filter has been used for more than 

60 days, replace it with a new one.

•  Do not tilt the nebulizer kit at an angle of greater than 30 degrees in all directions or 

shake it while in use.

•  Do not use or store the device while the air tube is creased.

•  Use only original nebulizing parts, air tube, air filter and air filter cover.

•  Do not add more than 12 ml of medication to the medication tank.

•  Do not carry or leave the nebulizer kit while the medication tank contains medication.

•  Do not leave the device unattended with infants or persons who cannot express their 

consent.

•  Do not subject the device to any strong shocks such as dropping on the floor.

•  Do not distort the vaporizer head, or jab the nozzle of the medication tank with a pin 

or any sharp object.

•  Do not insert fingers or objects inside the compressor.

•  Do not disassemble or attempt to repair the compressor or power cord.

•  Do not leave the device in extreme hot or cold temperature, or under direct sunlight.

•  Do not block the air filter cover.

•  Do not use the device while sleeping or if drowsy.

•  Approved for human use only.

•  The compressor may become hot during operation.

•  Do not touch the compressor for other than necessary operation such as turning off 

the power while nebulizing.

•  To avoid the medication residue on the face, be sure to wipe the face after removing 

the mask.

•  To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of 

the nose.

•  Do not block the slit between the cap and the inhalation air inlet.

•  Make sure that the vaporizer head is correctly installed before use.

•  Do not use a damaged nebulizer kit, mouthpiece or nosepiece.

•  While using this device, make sure that no mobile phone or any other electrical 

devices that emit electromagnetic fields is within 30cm. This may result in 

degradation of performance of the device.

(Risk of electrical shock)

•  Always unplug the device from the power outlet after use and before cleaning.

•  Plug the device into the appropriate voltage outlet. Do not overload power outlets or 

use extension cords.

•  Do not misuse the power cord.

•  Do not wind the power cord around the compressor.

•  Changes or modifications not approved by OMRON HEALTHCARE will void the user 

warranty.

•  Do not pull the power cord strongly.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these 

rules may result in damage, inefficient nebulization or infection. For the instructions, 

refer to “Cleaning and disinfecting” section.

Thank you for purchasing the NE-C101.
This product was developed in conjunction with respiratory therapists 

for the successful treatment of asthma, chronic bronchitis, allergies 

and other respiratory disorders. 

This is a medical device. Operate the device only as instructed by 

your doctor and/or respiratory therapist.

How the nebulizer kit works

The medication that is pumped up 

through the medication channel is 

mixed with compressed air which is 

generated by a compressor pump. 

The compressed air mixed with 

medication is turned into fine particles 

and sprayed when in contact with the 

baffle. 

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, KYOTO, 617-0002 JAPAN

EU-representative

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS

www.omron-healthcare.com

Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.

Binh Duong Province, VIETNAM

Asia Pacific HQ

OMRON HEALTHCARE SINGAPORE PTE LTD.

438A Alexandra Road, #05-05/08

Alexandra Technopark, Singapore 119967

www.omronhealthcare-ap.com

Made in Vietnam

•  Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.

•  Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.

•  When disinfecting parts by boiling, make sure that the container does not boil dry. 

Otherwise it may also result in a fire.

General Safety Precautions:

•  Inspect the device and parts before using them each time, and check that there are 

no problems. In particular, be sure to check the following:

- That the nozzle or air tube are not damaged.

- That the nozzle is not blocked.

- That the compressor operates normally.

•  When using this device, there will be some noise and vibration caused by the pump in 

the compressor. There will also be some noise caused by the emission of compressed 

air from the nebulizer kit. This is normal and does not indicate a malfunction.

•  Operate the device only as intended. Do not use the device for any other purpose.

•  Do not use the device at temperatures greater than +40°C.

•  Make sure that the air tube is securely attached to the compressor and nebulizer kit, 

and does not come loose. Twist the air tube plug slightly when inserting it into the air 

tube connectors to avoid the air tube disconnecting during use.

•  To completely isolate the device from the power source unplug the plug from the 

power source.

•  Using the device longer than the standard operating time for which it was designed 

may result in accidents such as fire or injury due to deterioration over time.

Keep these instructions for future reference.

Technical data

Product Category:

Nebulizers

Product Description:

Compressor Nebulizer

Model:

NE-C101

Rating:

230V  50Hz

Power Consumption:

150VA

Current Fuse:

T1.6AL250V

Operating Mode:

Continuous Use

Operating Temperature/

Humidity/Air Pressure:

+5°C to +40°C / 15% to 85%RH (non-condensing) 

700 to 1060hPa

Storage and Transport

Temperature/Humidity:

-20°C to +60°C / 5% to 95%RH (non-condensing) 

Weight:

Approx. 1.05kg (compressor only)

Dimensions:

Approx. 145 (W) × 124 (H) ×182 (D) mm 

(compressor only)

Contents:

Compressor, Nebulizer Kit, Air Tube (PVC, 

100cm), Mouthpiece, Nosepiece, Adult Mask 

(PVC), Child Mask (PVC), 2 pcs Spare Air Filters, 

Instruction Manual, Quick Start Guide.

Classification:

Class ll (Protection against electric shock),  

Type BF (Applied part); Mouthpiece, Nosepiece, 

Mask

IP21 (Ingress Protection)

 

Technical data for the NE-C101 Compressor:
Particle Size:

**MMAD Approx. 3.90µm

MMAD = Mass Median Aerodynamic Diameter

Medication Tank Capacity:

12ml maximum

Appropriate Medication 

Quantities:

2ml minimum - 12ml maximum

Noise level  

(at 1m distance): 

*Less than 63dB

Nebulization Rate:

*Approx. 0.3ml/min (by weight loss)

Aerosol Output:

0.259ml (2ml, 1%NaF)

Aerosol Output Rate:

0.05ml/min (2ml, 1%NaF)

Result of cascade impactor measurements for particle size**

Cumulative % particle mass of sodium fluoride under size

Particle Size Dp (µm)

Cumulative Undersize %

0

10

20

30

40

50

60

70

80

90

100

0.1

1

10

100

Individual tests
Mean

MMAD: 3.9μm

*

   Measured by OMRON HEALTHCARE Co., Ltd.

**  Measured values reflect internal data obtained by NEXT GENERATION IMPACTOR 

(NGI), according to EN13544-1:2007 + A1:2009 and ISO27427:2013.

Notes:

•  Subject to technical modification without prior notice.

•  This OMRON product is produced under the strict quality system of OMRON 

HEALTHCARE Co., Ltd., Japan.

•  The device may not work if the temperature and voltage conditions are different to 

those defined in the specifications.

•  Ensure that this device has acclimated to room temperature before nebulizing the 

medication. Nebulizing the medication after an extreme temperature change could 

lead to less nebulization. OMRON recommends waiting for approximately 2 hours 

for the device to warm up or cool down when the device is used at an environment 

within the temperature specified as operating conditions after it is stored either at 

the maximum or at the minimum storage temperature. For additional information of 

operating and storage/transport temperature, refer to Technical data.

•  The device fulfils the provisions of the EC directive 93/42/EEC (Medical Device 

Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory 

therapy equipment - Part1: Nebulizing systems and their components.

•  Performance may vary with medication such as suspensions or high viscosity. See 

medication supplier’s data sheet for further details.

•  IP classification is degrees of protection provided by enclosures in accordance with 

IEC 60529. The device is protected against solid foreign objects of 12.5mm diameter 

and greater such as finger and against vertically falling water drops which may cause 

issues during a normal operation.

Symbols description

Need for the user to consult the instructions for use

Applied part - Type BF

Degree of protection against electric shock (leakage current)
Class ll

Protection against electric shock

IPXX

Ingress protection degree provided by IEC 60529

CE Marking

Serial number

Temperature limitation

Humidity limitation

Atmospheric pressure limitation

Power OFF

Power ON

Alternating current

Product production date is integrated in the Serial number, which placed on the 

product and/or sales package: the first 4 digits mean year of production, the next 

2 digits mean month of production.

 

Correct Disposal of This Product

 

(Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates that it should not be disposed 

of, with other household wastes at the end of its working life. To prevent possible harm to the 

environment or human health from uncontrolled waste disposal, please separate this product 

from other types of wastes and recycle it responsibly to promote the sustainable reuse of 

material resources.

Household users should contact either the retailer where they purchased this product, or 

their local government office, for details of where and how they can return this item for 

environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the 

purchase contract. This product should not be mixed with other commercial waste for disposal.

Know your unit

Optional Medical Accessories

(within the scope of EC Medical Device Directive 93/42/EEC)

Product description

Model

Nebulizer Kit

NEB-NK-11AP

Nebulizer Kit & Adult Mask & Air Tube Set

NEB-NSET4-11AP

Nebulizer Kit & Child Mask & Air Tube Set

NEB-NSET5-11AP

Adult Mask (PVC)

NEB-MSLP-AP

Child Mask (PVC)

NEB-MSMP-AP

Mouthpiece

NEB-MP-AP

Nosepiece

NEB-NP-AP

Air Tube (PVC, 100cm)

NEB-TP-L10AP

Other Optional/Replacement Parts

Product description

Model

Air Filters x 5

NEB-AFR-30AP

Compressor Nebulizer

Model 

NE-C101

Instruction Manual

Baffle

Aerosol

Medication 

Channel

Medication 

Aerosol

Nozzle

Medication 

Compressed Air

2829198-6C

12M3148

Nebulizer Kit x 1

Nosepiece x 1

Mouthpiece x 1

Adult Mask (PVC) x 1

Child Mask (PVC) x 1

Air Tube 

(PVC, 100cm) x 1

Instruction Manual
Quick Start Guide

Spare Air Filters x 2

Compressor (Main Unit)

Power Switch

Air Tube 

Connector

Air Filter Cover 

Including air filter

Nebulizer Kit 

Holder

Ventilation Slots

Rear View

Power Cord

Ventilation Slots

Power Plug

Air Tube Connector

Inhalation Top

Vaporizer Head

Medication Tank

Nozzle

Reviews: