MX3 Plus

English

Important information regarding
Electro Magnetic Compatibility (EMC)

With the increased number of electronic
devices such as PC’s and mobile
(cellular) telephones, medical devices in
use may be susceptible to
electromagnetic interference from other
devices. Electromagnetic interference
may result in incorrect operation of the
medical device and create a potentially
unsafe situation.
Medical devices should also not interfere
with other devices.

In order to regulate the requirements for
EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product
situations, the IEC60601-1-2 standard
has been implemented. This standard
defines the levels of immunity to
electromagnetic interferences as well as
maximum levels of electromagnetic
emissions for medical devices.

This medical device manufactured by
OMRON Healthcare conforms to this 
IEC60601-1-2:2001 standard for both
immunity and emissions.
Nevertheless, special precautions need
to be observed:
Do not use mobile (cellular) telephones
and other devices, which generate strong
electrical or electromagnetic fields, near
the medical device. This may result in
incorrect operation of the unit and create
a potentially unsafe situation.
Recommendation is to keep a minimum
distance of 7 m. Verify correct operation
of the device in case the distance is
shorter.

Further documentation in accordance
with IEC60601-1-2:2001 is available at
OMRON Healthcare Europe at the
address mentioned in this instruction
manual.
Documentation is also available at
www.omron-healthcare.com.

Correct Disposal of This Product

(Waste Electrical & Electronic
Equipment)                                              
This marking shown on the product or its
literature, indicates that it should not be
disposed with other household wastes at
the end of its working life. To prevent
possible harm to the environment or
human health from uncontrolled waste
disposal, please separate this from other
types of wastes and recycle it responsibly
to promote the sustainable reuse of
material resources.

Household users should contact either
the retailer where they purchased this
product, or their local government office,
for details of where and how they can
take this item for environmentally safe
recycling.

Business users should contact
their supplier and check the
terms and conditions of the
purchase contract. This product
should not be mixed with other
commercial wastes for disposal.

This product does not contain any
hazardous substances.

26032701 MX3Plus GB 01-12  14-04-2006  15:31  Pagina 10

Summary of Contents for MX 3 Plus

Page 1: ...weisung Seite 13 Gebruiksaanwijzing pagina 25 Mode d emploi page 37 Manuale de instructone pagina 49 Manual de instrucciones pagina 61 Automatic Blood Pressure Monitor DIA m m Hg SYS m m Hg MX3 Plus PULSE min MX3 Plus IM HEM 742 E 2 04 06 ...

Page 2: ... are stored in the memory of the OMRON MX3 Plus The OMRON MX3 Plus includes a standard cuff for arm circumferences ranging from 22 32 cm Optional large cuffs 32 42 cm can be applied Introduction 1 How to obtain meaningful blood pressure readings Table of contents General The OMRON MX3 Plus is not suitable for measuring the frequency of cardiac pacemakers Consult your doctor during pregnancy arrhyt...

Page 3: ...s measure on the same arm Wrap the cuff around your arm before start of measurement Caution Operating temperature between 10 C and 40 C Operating environment must be free from excessive vibrations shocks magnetic fields electrical noise etc Stay out of sunlight during measuring Keep portable phones 5 meters away during measurement Do not wrap the cuff around any object other than your arm Do not f...

Page 4: ...SYS mmHg DIA mmHg M MX3 Connect cuff 1 Insert air plug in air jacket L G Fit cuff Caution Do not inflate the cuff when not wrapped around the arm 1 Bare your arm Caution Push up your sleeve avoid constricting the blood flow G Air tube L Ring 3 Preparation 2 Pass the end of the cuff through the ring Caution Keep the air tube on the outside 3 Put your arm through the loop Note keep the tube even wit...

Page 5: ... functions B Memory recall button C Start button D On off button Measurement M Systolic blood pressure in mmHg N Heart symbol flash measurement takes place light up finish measurement P Diastolic blood pressure in mmHg Q Memory display values in memory R Pulse heartbeats per minute S Battery exhausted batteries are weak exhausted T Deflation cuff deflates and measurement finishes U Inflation cuff ...

Page 6: ...urement PULSE min DIA mmHg SYS mmHg M MX3 Plus PULSE min DIA mmHg SYS mmHg M MX3 Plus PULSE min DIA mmHg SYS mmHg M MX3 Plus 3 Press on off button Caution Wait for zero and the heart symbol to continue 4 Press start button Note the unit inflates the cuff to 170 mmHg If required the unit automatically increases the pressure by 40 mmHg Note press and hold the start button to set pressure higher than...

Page 7: ...es in your personal blood pressure pass 3 Press on off button to turn off power Note power automatically shuts of in 5 minutes PULSE min DIA mmHg SYS mmHg M MX3 Plus PULSE min DIA mmHg SYS mmHg M MX3 Plus PULSE min DIA mmHg SYS mmHg M MX3 Plus PULSE min DIA mmHg SYS mmHg M MX3 Plus 1 Read the values of blood pressure and pulse from the display Note display shows blood pressures and heartbeats alte...

Page 8: ...tbeats 3 Press on off button to turn off power PULSE min DIA mmHg SYS mmHg M MX3 Plus M PULSE min DIA mmHg SYS mmHg M MX3 Plus PULSE min DIA mmHg SYS mmHg M W Erase memory W Memory position of latest and previous values 1 14 1 Press on off button to turn on power 2 Press the memory recall button for 2 3 seconds and then whilst continuing to press the recall button press the start button Caution Al...

Page 9: ... repairs of any kind yourself If a defect occurs consult your OMRON retail outlet or distributor as mentioned on the packaging Calibration It is recommended to have the OMRON MX3 Plus inspected every two years to ensure correct function and accuracy Consult your OMRON retail outlet or OMRON distributor as mentioned on the packaging Replace batteries Remove the battery out of the battery 6 Error ca...

Page 10: ... circumference 22 cm to 32 cm Accessories Cuff instruction manual guarantee card blood pressure pass Optional parts Large cuff for arm circumferences between 32 42 cm AC adapter Note Subject to technical modification without prior notice Type B 0197 Manufacturer OMRON HEALTHCARE Co Ltd 24 Yamanouchi Yamanoshita cho Ukyo ku Kyoto 615 0084 Japan EU representative OMRON HEALTHCARE EUROPE B V Kruisweg...

Page 11: ...is may result in incorrect operation of the unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in accordance with IEC60601 1 2 2001 is available at OMRON Healthcare Europe at the address mentioned in this instruction manual Documentation is also available at ...

Page 12: ...dy fat and high levels of choles terol LDL which sticks to the inside in blood vessels reduces elasticity of these vessels Hypertension accelerates arterial sclerosis which can lead to very serious conditions such as stroke and myocardial infarction For these reasons it is very important to know whether the blood pressure is within a healthy range Blood pressure fluctuates from minute to minute th...

Page 13: ...ry nervousness and tension may cause your blood pressure to be higher than in familiar circumstances at home This is known as White Coat Hypertension Treatment of high blood pressure If your blood pressure reaches upper values of 140 160 mmHg and lower values of 90 95 mmHg in repeated measurements over several days you should consult your doctor for detailed medical examination You can assist the ...

Reviews: