Failure and their possible causes

Incorrect measurement is indicated by 

1. A correct reading could not be obtained

because measurement was disturbed by
movement of the body.

2. The cuff is not fitted correctly.

3. Your clothing has constricted your blood flow.

4. There is still air in the cuff when the monitor

is switched off.

The start button was pressed before the
(

) symbol was displayed

The blood pressure values displayed are ex-
tremely low or high, or they are implausible.

The display does not light up when the (

)

button is pressed.

1. The batteries are exhausted.

2. The +/- poles of the batteries have been

reversed.

3. The battery contacts are dirty.

The batteries are weak or exhausted.

The cuff pressure does not rise although
the pump motor can be heard.

Err  problem with mem. function

Failure, causes and rectification

Rectification

1. Repeat the measurement keeping

perfectly still. Do not move your arm and
do not speak.

2. Check that the cuff is correctly fitted. Then

repeat the measurement.

3. Remove the item of clothing which cau-

sed the constriction.

4. The unit may be defective. Please have it

examined and, if necessary, repaired by
OMRON Customer Services.

Wait for the (

) symbol indicating readin-

ess before you press the start button.

Refer to the instructions and then repeat the
measurement.

1. Check the batteries and, if necessary, fit 4

new batteries.

2. Reinsert the batteries with the +/- poles

the right way round.

3. Clean the battery contacts with a dry cloth.

Fit new batteries.

Check that the air tube is properly connec-
ted to the monitor. Push the tube connector
firmly into the socket.

Contact Omron Customer Service.

• Before measuring, stretch your back and sit up straight,

breathe normally 5-6 times, relax your shoulders, arms
and entire body. 

• Slightly bend and support your elbow ( e.g. on a table ). 
• Do not move, talk or touch the device during measure-

ment. 

• Stress influences your blood pressure, so try to relax

before taking a measurement. 

• To compare results, measurements must always be

taken from the same arm.

• Do not place the cuff over thick clothes and do not roll up

your sleeve if it is too tight.

• Align the cuff with the height of your heart (breast

height). If necessary use a pillow or cushion.

• Coldness raises your blood pressure. It is recommended

to measure at room temperature ( +20°C).

• Relax before measurement. If necessary go to the

bathroom before measurement.

• Do not measure directly after bathing.
• If you wish to interrupt measurement prematurely, press

the (

) button. Inflation is stopped and the air is auto-

matically released from the cuff.

• If you wish to repeat the measurement, follow the same

procedure as before starting at step 4 (Inflating the cuff).

• Blood pressure measurements taken in a moving vehicle

may be influenced by vibration.

• If an error has occurred during measurement  is dis-

played. Please refer to the next chapter.

• The monitor should be switched off after measurement

to avoid running down the batteries unnecessarily. If you
forget to do so, the monitor switches itself off automati-
cally after 5 minutes.

Hints on taking
blood pressure
readings

8

9

E

Summary of Contents for M4-I

Page 1: ...nte UE EU Vertegenwoordiger OMRON Healthcare Europe BV Kruisweg 577 NL 2132 NA Hoofddorp Gebrauchsanweisung Seite 28 Istruzioni per l uso Pagina 54 Gebruiksaanwijzing Pagina 15 Instrucciones de uso Página 67 Instruction manual Page 2 Mode d emploi Page 41 The World Hypertension League SYS DIA PULSE min M4 I START mmHg mmHg M4 I ...

Page 2: ... the evening after work or as advised by your doctor Please remember Self measurement is not the same as medical treatment You should never change the dose of medicines prescribed by your doctor Important instruc tions to help you obtain meaningful readings Thank you for purchasing an OMRON product You can be sure you have made a wise choice By buy ing the OMRON M4 I blood pressure monitor you hav...

Page 3: ...ompact fully automatic blood pressure monitor Monitor operating on the oscillometric principle It measures your blood pres sure and pulse simply and quickly It contains an intelligent fuzzy logic system for controlled inflation known as Intelli sense This is an advanced method of oscillo metric measurement and does not require pressure pre setting or re inflation OMRON is the world lea ding manufa...

Page 4: ...s to flash When the monitor no longer detects your pulse whilst the cuff pressure is dropping your systolic and diastolic blood pressures upper and lower values will be displayed The deflation symbol will now flash to indicate the end of the measurement When all the air has been released the symbol appears on the display and your blood pressure and pulse will be displayed on the screen for 5 minut...

Page 5: ...y contacts with a dry cloth Fit new batteries Check that the air tube is properly connec ted to the monitor Push the tube connector firmly into the socket Contact Omron Customer Service Before measuring stretch your back and sit up straight breathe normally 5 6 times relax your shoulders arms and entire body Slightly bend and support your elbow e g on a table Do not move talk or touch the device d...

Page 6: ...pplementary requirements for electrome chanical blood pressure measuring systems For cleaning the monitor only use a soft lightly moi stened cloth Do not use petroleum spirits thinners or similar solvents Stains on the cuff can be carefully removed using a moist cloth and soapy water Do not wash the cuff or make it wet Keep your blood pressure monitor in the storage case to protect it from dust an...

Page 7: ...ical activity anxiety or the time of day can influence your blood pressure Blood pressure is typically low in the morning and increases from afternoon to eve ning It is lower in the summer and higher in the winter Blood pressure is measured in millimetres of mercury mmHg and measurements are written with the systolic pressure before the diastolic e g A blood pressure written as 140 90 is referred ...

Page 8: ...fect blood pressure a single measurement may not be sufficient for an accurate diagnosis The blood pressure measured first thing in the morning after getting up before eating and while at rest is known as the fundamental blood pressure In practice it is rather difficult to record the fundamental blood pressure but it is important to get readings in an environment that is as close as possible to th...

Reviews: