EN8
EN
6. Specifications
Product description
Automatic Upper Arm Blood Pressure Monitor
Product category
Electronic Sphygmomanometers
Model (code)
M3 (HEM-7154-E)
Display
LCD digital display
Cuff pressure range
0 to 299 mmHg
Pulse measurement range 40 to 180 beats / min.
Blood pressure measurement range SYS: 60 to 260 mmHg / DIA: 40 to 215 mmHg
Accuracy
Pressure: ±3 mmHg / Pulse: ±5% of display reading
Inflation
Automatic by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
Operating mode
Continuous operation
IP classification
Monitor: IP20 / Optional AC adapter: IP21
Rating
DC6 V 4.0 W
Applied part
Type BF (arm cuff)
Power source
4 “AA” batteries 1.5 V or optional AC adapter (INPUT AC 100 - 240 V 50 - 60 Hz
0.12 - 0.065 A)
Battery life
Approximately 1000 measurements (using new alkaline batteries)
Durable period (Service life)
Monitor: 5 years / Cuff: 5 years / Optional AC adapter: 5 years
Operating conditions
+10 to +40 °C / 15 to 90% RH (non-condensing) / 800 to 1060 hPa
Storage / Transport conditions
-20 to +60 °C / 10 to 90% RH (non-condensing)
Contents
Monitor, arm cuff (HEM-RML31), 4 “AA” batteries, storage case, Instruction Manual and
Protection against electric shock
Internally powered ME equipment (when using only batteries)
Class II ME equipment (optional AC adapter)
Weight
Monitor: approximately 310 g (not including batteries) / Arm cuff: approximately 170 g
Dimensions (approximately value) Monitor: 105 mm (W) × 87 mm (H) × 153 mm (L)
Arm cuff: 145 mm × 594 mm (air tube: 750 mm)
Memory
Stores up to 60 readings per user
Note
• These specifications are subject to change without notice.
• This monitor is clinically investigated according to the requirements of ISO 81060-2:2013. In the clinical validation study, K5 was used on
85 subjects for determination of diastolic blood pressure.
• IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are
protected against solid foreign objects of 12.5 mm diameter and greater such as a finger. The optional AC adapter is protected against vertically
falling water drops which may cause issues during a normal operation.