background image

Manual Inflation Blood Pressure Monitor

Model M1 Plus

• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
•     
• Instrukcja
• Kezelési Útmutató

EN

FR

DE

IT

ES

NL

RU

IM-HEM-4011C-E-03-10/2011

1649533-0C

PL

HU

Summary of Contents for M1 Plus

Page 1: ... Monitor Model M1 Plus Instruction Manual Mode d emploi Gebrauchsanweisung Manuale di istruzioni Manual de instrucciones Gebruiksaanwijzing Instrukcja Kezelési Útmutató EN FR DE IT ES NL RU IM HEM 4011C E 03 10 2011 1649533 0C PL HU ...

Page 2: ...des 13 3 Using the Unit 14 3 1 How to Sit Correctly When Taking a Measurement 14 3 2 Applying the Arm Cuff 16 3 3 Taking a Reading 18 3 4 Using the Memory Function 23 4 Quick Reference Guide 26 5 Handling Errors and Problems 27 5 1 Error Messages 27 5 2 Troubleshooting 29 6 Maintenance and Storage 31 7 Optional Parts 34 8 Technical Data 35 9 Some Useful Information about Blood Pressure 37 ...

Page 3: ... measurements in memory and calculates an average reading based on the three most recent measurements taken within 10 minutes of the last reading If there are only two readings in memory for that period the average will be based on two readings If there is only one reading in memory for that period that reading will be displayed as the average value Please read this instruction manual thoroughly b...

Page 4: ...uff Only use four AAA manganese or alkaline batteries with this unit Do not use other types of batteries When using batteries the battery liquid may leak and damage the main unit Please observe the following points Do not insert the batteries with their polarities incorrectly aligned Replace old batteries with new ones immediately When you are not going to use the unit for a long period of time ap...

Page 5: ...5 1 Overview EN 1 Overview Main unit A Display B SET button C Air connector D MEMORY button E I O button F Battery compartment B A C C E D F ...

Page 6: ...arts to leak from the arm cuff please replace with a new one Refer to Chapter 7 Optional Parts Inflation Bulb G Arm cuff Medium cuff arm circumference 22 32 cm H Air Tube I Air Release Button J Air Tube K Air Inflation Bulb H G J K I ...

Page 7: ...easurements P Buzzer symbol Q Battery low symbol R Heartbeat symbol 1 Flashes during measurement 2 If flashing after measurement is complete indicates blood pressure out of recommended range S Pulse T Deflation symbol U Date Time V Irregular heartbeat symbol W Reinflation symbol X Storage case Y Four AAA alkaline LR03 batteries Instruction manual Guarantee card Blood pressure pass Blood Pressure M...

Page 8: ...art of the cover 3 Install or replace four AAA size batteries so that the positive and negative polarities match the polarities indicated on the battery compartment 4 Put the battery cover back in place Slide the battery cover as indicated until it clicks into place Note The measurement values continue to be stored in memory for maximum 30 seconds even after the batteries are replaced ...

Page 9: ...ies with new ones at once Turn the unit off before replacing the batteries Remove the batteries if the unit will not be used for three months or more If the batteries are removed for 30 seconds the Date Time setting will need to be reset See 2 2 Setting the Date and Time for details Dispose of batteries according to applicable local regulations Four new AAA alkaline batteries will last for approxi...

Page 10: ... will be displayed as the average value To make use of the memory and average values functions Set the unit to the correct date and time before taking a measurement for the first time If the batteries have been removed for more than 30 seconds the date and time setting will need to be reset 1 When the I O button is pressed to turn the unit on for the first time after inserting batteries the year d...

Page 11: ...and the month digits flash on the display 4 Repeat steps 2 and 3 to set the month The month is set and the day digits flash on the display 5 Repeat steps 2 and 3 to set the day The day is set and the hour digits flash on the display 6 Repeat steps 2 and 3 to set the hour The hour is set and the minutes digits flash on the display ...

Page 12: ...s to standby mode after the minute setting has been set To adjust the date and time press the SET button twice while the monitor is in standby mode Notes After the buzzer setup is completed the time date setting will flash on the display After each setup is completed please push the I O button to turn off the monitor ...

Page 13: ...d and when measurement is complete 1 Press the SET button once while the unit is in standby mode The buzzer symbol appears on the display 2 Press the MEMORY button to select on or off The setting alternates between on and off with each press of the MEMORY button The default setting is on 3 When you have selected the desired setting press the I O button to turn the unit off Notes If you want to adj...

Page 14: ... take measurement after bathing drinking alcohol or exercising Do not move or talk during measurement Correct Posture Remove tight fitting clothing from your upper arm and any thick clothing such as a sweater Do not place the cuff over thick clothes and do not roll up your sleeve if it is too tight Sit upright with your back straight Place your arm on a table so that the cuff will be at the same l...

Page 15: ...hour after waking up is recommended Incorrect Posture These situations tend to lead to higher blood pressure values due to strain or the arm cuff being lower than the heart If the arm cuff is at a lower position than your heart use cushions etc to adjust the height of your arm Arched back leaning forwards Sitting cross legged Sitting on a sofa or at a low table so that you lean forward ...

Page 16: ...ng a reading 1 Connect the air tube of the arm cuff and the inflation bulb to their respective air connectors 2 Put your left arm through the cuff loop The air tube should run down the inside of your forearm and be in line with your middle finger Note If the cuff is not assembled pass the end of the cuff furthest from the tubing through the metal D ring to form a loop The smooth cloth should be on...

Page 17: ...f should be 1 to 2 cm above the elbow 4 When the cuff is positioned correctly close the fabric fastener FIRMLY 1 The air tube should run down the inside of your forearm and be in line with your middle finger 2 The bottom of the cuff should be approximately 1 to 2 cm above your elbow 3 Apply the cuff to your upper arm so that the coloured marker blue arrow under tube is centered on the middle of yo...

Page 18: ... heartbeat symbol will appear and the current time and 0 is displayed Note If the deflation symbol does not disappear soon press the air release button to release any air in the arm cuff 2 Pump the inflation bulb to inflate the arm cuff 1 Inflate the cuff until it is 30 to 40 mmHg above your expected systolic blood pressure value ex If your expected blood pressure is around 140mmHg inflate the arm...

Page 19: ...mbol flashes Remain still and do not talk during measurement During measurement the buzzer if set to on will beep in rhythm with your heartbeat Note To cancel a measurement press the I O button to turn off the unit and press the air release button to release the air in the arm cuff Common Causes of Inaccurate Results Leaning forwards Arm cuff is too loose Movement or talking during measurement Not...

Page 20: ...emory Function The buzzer will sound if switched on when measurement is completed Notes Self diagnosis of measured results and treatment are dangerous Please follow the instructions of your doctor Wait 2 3 minutes before taking another blood pressure measurement Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement 6 Undo the fastene...

Page 21: ...ludes an irregular heartbeat feature Irregular heartbeats can influence the results of the measurement The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated If the measurement results are affected by irregular heartbeats but the result is valid the result is shown together with the irregular heartbeat symbol If the irregular heartbeats caus...

Page 22: ...ystem that drives the heartbeat Typical symptoms are skipped heartbeats premature contraction an abnormally rapid tachycardia or slow bradycardia pulse This can be caused by heart disease aging physical predisposition stress lack of sleep fatigue etc Arrhythmia can only be diagnosed by a doctor through a special examination Whether the appearance of the irregular heartbeat symbol in the results in...

Page 23: ...ading in memory for that period that reading will be displayed as the average value Note When 21 sets of readings are stored in memory the oldest set will be deleted to store a new set 1 Press the O I button to turn the power on 2 When the heartbeat symbol appears on the display press the MEMORY button If there are three readings stored in memory that were taken within 10 minutes of the last readi...

Page 24: ...or diastolic pressure is outside the standard range the heartbeat symbol will blink when the measurement result is displayed Refer to 3 3 number 7 4 Press the MEMORY button repeatedly to cycle through the previous measurement results Keep the button pressed down to cycle rapidly through the previous results If the irregular heartbeat symbol was displayed at the time of a measurement this is displa...

Page 25: ...rst If the I O button is pressed first the monitor is shut off 2 Press the I O button to turn the unit off If you forget to turn the unit off it will automatically shut itself off after five minutes Notes If the date and time are reset to a time before the most recent measurement the average value will be based on any measurements taken after the date and time were reset However you can still view...

Page 26: ...at the same level as your heart 2 Apply the arm cuff to your upper arm The coloured marker should be centred on the inside of your arm and point down the inside of the arm so that the air tube runs down the inside of your forearm and is in line with your middle finger 3 Secure the cuff around your arm using the fabric fastener strip 4 Press the I O button to turn the unit on then pump the inflatio...

Page 27: ...nected Insert the air tube securely Refer to section 3 2 Arm cuff not applied correctly Apply the arm cuff correctly Refer to section 3 2 Clothing is interfering with the arm cuff Remove any clothing interfering with the arm cuff Refer to section 3 2 Air is leaking from the arm cuff Replace cuff with new one Refer to Chapter 7 The arm cuff was inflated above 299 mmHg Release your hand from the air...

Page 28: ...ontact your OMRON retail outlet or distributor Refer to Chapter 8 Pressure is too low Press the inflation bulb to inflate the arm cuff until the reinflation symbol goes out Or deflate the arm cuff and repeat measurement after checking that the heartbeat symbol has been displayed Refer to section 3 3 Error Display Cause Remedy ...

Page 29: ... air tube is connected securely Refer to section 3 2 Air is leaking from the arm cuff Replace the arm cuff with a new one Refer to Chapter 7 Arm cuff deflates too soon The arm cuff is loose Apply the cuff correctly so that it is firmly wrapped around the arm Refer to section 3 2 Cannot measure or readings are too low or too high Has the arm cuff been inflated sufficiently Inflate the cuff so that ...

Page 30: ...n 2 1 The batteries have been inserted incorrectly Insert the batteries with the correct polarity Refer to section 2 1 Other problems Press the I O button and repeat measurement If the problem continues try replacing the batteries with new ones If this still does not solve the problem contact your OMRON retail outlet or distributor Problem Cause Remedy ...

Page 31: ...ble the unit Do not subject the unit to strong shocks or vibrations for example dropping the unit on the floor Do not use volatile liquids to clean the main unit The unit should be cleaned with a soft dry cloth Use a soft moistened cloth and soap to clean the arm cuff Do not wash the arm cuff or immerse it in water Do not use petrol thinners or similar solvents to clean the arm cuff Do not carry o...

Page 32: ... the unit inspected every two years to ensure correct functioning and accuracy Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature Do not carry out any repairs yourself If a defect occurs or you have doubts about the correct functioning of the device consult your authorised OMRON dealer or the OMRON Customer Service...

Page 33: ... not bend the air tube excessively 3 Place the arm cuff inflation bulb and main unit in the storage case Do not store the unit in the following situations If the unit is wet Locations exposed to extreme temperatures humidity direct sunlight dust or corrosive vapours Locations exposed to vibrations shocks or where it will be at an angle Locations exposed to chemicals or corrosive vapours ...

Page 34: ...Arm circumference 22 32 cm Large Arm Cuff Arm circumference 32 42 cm Small Arm Cuff Arm circumference 17 22 cm Small cuff and bulb combination Arm circumference 17 22 cm Regular bulb CM1 7935058 8 CM 4997086 7 CL 4997065 4 CS 4997076 0 4997099 9 4997965 1 ...

Page 35: ...he device Product Description Manual Inflation Blood Pressure Monitor Model OMRON M1 Plus HEM 4011C E Display LCD Digital Display Measurement Method Oscillometric method Measurement Range Pressure 0 mmHg to 299 mmHg Pulse 40 to 180 min Memory 21 Measurements with date and time Accuracy Pressure r3 mmHg Pulse r 5 of display reading Inflation Manual by inflation bulb Deflation Automatic pressure rel...

Page 36: ... may result in incorrect operation of the unit and create a potentially unsafe situation Recom mendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in accordance with EN60601 1 2 2001 is available at OMRON Healthcare Europe at the address mentioned in this instruction manual Documentation is also available at ww...

Page 37: ...g your blood pressure measured by a doctor can cause anxiety which is itself a cause of high blood pressure As a variety of conditions affect blood pressure a single measurement may not be sufficient for an accurate diagnosis Many factors such as physical activity anxiety or the time of day can influence your blood pressure Thus it is best to try and measure your blood pressure at the same time ea...

Page 38: ...ent departments of hospitals There is no universally accepted definition of hypotension However those having the systolic pressure below 100 mmHg are assumed as hypotensive Example fluctuation within a day male 35 years old mmHg 200 150 100 50 6 12 18 24 Upper curve systolic blood pressure Lower curve diastolic blood pressure Time of day 80 180 160 140 130 120 85 90 100 110 Diastolic blood pressur...

Page 39: ...2132 LR Hoofddorp THE NETHERLANDS www omron healthcare com Production facility OMRON DALIAN CO LTD Dalian CHINA Subsidiary OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 0DG U K OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH John Deere Str 81a 68163 Mannheim GERMANY www omron medizintechnik de OMRON SANTÉ FRANCE SAS 14 rue de Lisbonne 93561 Rosny sous Bois Cedex FRANCE Made in China...

Reviews: