Safety Precautions

Warning Indications

Precautions for Safe Use

(1) 

Do not use or store the Product in the following locations:
 • Outdoor or locations subject to direct sunlight
 • Locations subject to rain and wind damage
 • Locations subject to excessive vibration or shock
 • Locations subject to rapid temperature changes
 • Locations prone to icing and dew condensation
 • Locations subject to water or oil
 • Locations subject to dust or corrosive gases 
   (particularly sulfurizing gases, ammonia, etc.)
 • Locations subject to influence of static electricity and noise
 • Locations subject to bugs and small animals

(2) A 

switch or circuit breaker should be provided close to this unit.

The switch or circuit breaker should be within easy reach of 
the operator, and must be marked as a disconnecting means 
for this unit.

(3) 

Mount the Product in the correct direction for installation.

(4) 

Be sure to use power terminals carefully, because power 
supply terminals have hazardous voltage.

(5) 

Use the wire given in this manual.

(6) 

Use the power supply voltage within the range of the 
specifications and rated values. Use the input voltage within 
the range of the specifications and rated values.

(7) Make 

sure the crimp terminals for wiring are of the specified size.

(8) Do 

not connect anything to terminals that are not being used.

(9) 

Confirm the wiring the input and output terminals correctly 
before power is supplied.

(10)  The terminal block may be damaged if you insert a 

flat-blade screwdriver in the release hole with excessive 
force. When inserting a flat-blade screwdriver into the 
release holes, operate with a force of 15·N or less.

(11) K7GE-MG may be subject to radio disturbances. Do not install 

the Product near equipment that generates high frequencies or 
surges.

(12)  The maximum terminal temperature is 80°C. Use wires with a 

heat resistance of 80°C min to wire the terminals.

(13)  Make sure the LCD and the LEDs for output indicators 

operate correctly. Depending on the operating environment, 
the Product parts may deteriorate faster than expected, 
causing the indicators to fail.

(14) Use 

the 

cable within the length that is rated in the specification 

requirements for the wiring between the sensor and the 
Product. As for the requirements on the cable distance, refer to 
K7GE-MG User's Manual (Cat. No. N224).

(15) In 

order to prevent inductive noise, wire the lines connected to 

the Product separately from power lines carrying high voltages 
or currents. Also, do not wire in parallel with or on the same 
cables as power lines. Other measures for reducing noise are 
to separate from ducts including noisy lines.

(16)  Do not continue to use the Product if the front surface peels.
(17) The 

alarm output function is a function for the output of an alarm 

when the set threshold value is below. Do not use this function for 
control, etc.

(18)  Use this Product inside the control panel to prevent external noise.
(19) When 

discarding the Product, properly dispose of it as industrial 

waste.

(20)  Never touch the charging terminal of the load while the 

K7GE-MG is in measurement operation.

(21)  K7GE-MG cannot be used for legal inspection. Be sure to 

use a periodically calibrated measuring instrument for legal 
inspection.

(22) When wiring, wire by enough length.
(23)  Check terminal polarity when wiring and wire all connections 

correctly. Do not wire the input and output terminals incorrectly.

(24) K7GE-MG provides 50 VDC of megger voltage. Do not use on 

equipment that may be damaged by this voltage.

(25)  Use and store the Product in a location where the ambient 

temperature and humidity are within the specified ranges. If 
applicable, provide forced cooling.

(26)  Please read and understand this Instruction Manual before 

using K7GE-MG.

Observe the following operating methods to prevent failure and 
malfunction.
(1)  During periodic inspection, installation of an additional 

sensor, or adjustment of sensor position, use the Product 
after ensuring that correct operation can be performed.

(2)  When mounting K7GE-MG on a DIN Track, follow the 

installation method shown on the right to install it correctly.

(3)  Confirm that wire does not stick up after wiring of stranded 

cable.

(4)  In case of crossover wiring, install these by 10 A per 1 

terminal because when Products are connected more than 
one in parallel, quite many electric currents to be called off.

(5)  The terminal block may be damaged if specialized tool is 

not used. Use a recommended flat-blade screwdriver to 
inserted into a release hole on the terminal block.

(6)  This Product is designed for use by qualified personnel with 

a knowledge of electrical systems. Read this Instruction 
Manual carefully before using the Product.

(7)  Use the power supply voltage, input power, and other 

power supplies and transformers with suitable capacities 
and rated outputs.

(8)  Use a power supply that will reach the rated voltage within 

1 second after the power is turned ON.

(9) Do 

not 

install the Product close contact with the heating element.

(10)  Do not install the Product near equipment that generates 

high frequencies or surges.

(11) Install the Product so that the load doesn't span the Product 

body.

(12)  If an error occurs during K7GE-MG operation, stop 

operation immediately and make suitable corrections such 
as replacement.

(13)  Do not use K7GE-MG for safety devices or applications that 

could result in loss of life.

(14)  Make sure to connect Main Unit and Probe units before use.
(15)  Make sure that the number of additional Probe Units is 

within the specified range.

(16)  Be sure to install a magnetic contactor, etc. between the 

power supply and the load, and wire K7GE-MG on the 
secondary side of it.

(17)  Do not turn ON power to the load while K7GE-MG is 

measuring.

(18)  Set the "Motor stop waiting time" setting value to at least 

the time from when the contactor is turned off until the load 
completely stops.

(19)  If you accidentally drop K7GE-MG, the inside of the Product 

may be damaged, so do not use it.

(20)  Do not wire anything to the release holes.
(21)  Do not tilt or twist a flat-blade screwdriver while it is inserted 

into a release hole on the terminal block. The terminal block 
may be damaged.

(22)  Insert a flat-blade screwdriver into the release holes at an 

angle. The terminal block may be damaged if you insert the 
screwdriver straight in.

(23)  Do not allow the flat-blade screwdriver to fall out while it is 

inserted into a release hole.

(24)  Do not bend a wire past its natural bending radius or pull on 

it with excessive force. Doing so may cause the wire 
disconnection.

(25)  Do not insert more than one wire into each terminal 

insertion hole.

(26)  Do not use any liquids such as paint thinner, similar 

solvents or alcohol to clean the Product. Clean it with a soft, 
dry cloth.

• For wiring from the Probe Unit to the load, use a Class CC, Class J, or Class T fuse with a rated current of 7A or less.
• The protection provided by the device may be impaired if the device is used in a manner that is not specified by the manufacturer. 
•To use the Product, install it as an embedded device within a control panel.
• The table below shows the nominal voltage and measurement circuit connections available for each measurement category in the 

Main Power Supply System Configurations. Do not use the device under conditions that exceed this category and conditions.

This is a class A product. In residential areas it may cause radio 
interference, in which case the user may be required to take 
adequate measures to reduce interference.

Suitability for Use

Omron Companies shall not be responsible for conformity with 
any standards, codes or regulations which apply to the 
combination of the Product in the Buyer’s application or use of 
the Product.
At Buyer’s request, Omron will provide applicable third party 
certification documents identifying ratings and limitations of 
use which apply to the Product. This information by itself is not 
sufficient for a complete determination of the suitability of the 
Product in combination with the end product, machine, system, 
or other application or use. Buyer shall be solely responsible 
for determining appropriateness of the particular Product with 
respect to Buyer’s application, product or system. Buyer shall 
take application responsibility in all cases.
NEVER USE THE PRODUCT FOR AN APPLICATION 
INVOLVING SERIOUS RISK TO LIFE OR PROPERTY OR IN 
LARGE QUANTITIES WITHOUT ENSURING THAT THE 
SYSTEM AS A WHOLE HAS BEEN DESIGNED TO ADDRESS 
THE RISKS, AND THAT THE OMRON PRODUCT(S) IS 
PROPERLY RATED AND INSTALLED FOR THE INTENDED 
USE WITHIN THE OVERALL EQUIPMENT OR SYSTEM.

Minor electric shock, fire, or malfunction may 
occasionally occur. Do not allow metal objects, 
conductors, or cuttings from installation work to 
enter the Product.

Minor injury due to explosion may occasionally occur. 
Do not use the Product where subject to flammable or 
explosive gas.

Minor electric shock, fire, or malfunction may 
occasionally occur. Do not disassemble, modify, or 
repair the Product or touch the interior of the Product.

If used the Product with incorrect wiring, it may cause 
property damage due to ignition. Make sure the cable 
is connected properly when the power supply is turned 
ON.

If the wiring material is inserted in a shallow position, 
it may cause property damage due to ignition. When 
wiring, make sure that the wiring material is properly 
inserted all the way into the terminal block.

Minor injury due to electric shock may occasionally 
occur. Do not touch the Product except for any 
buttons (keys) while power is being supplied.

Always connect the protective earthing terminal (     ) 
to a ground. Use AWG 16 for the protective conductor.

Perform periodic inspection to the Product.
A malfunction in the Product may occasionally 
prevent monitoring impossible or alarm outputs, 
resulting in property damage to connected equipment 
or devices. To maintain safety in the event of 
malfunction of the Product, take appropriate safety 
measures, such as installing a monitoring device on a 
separate line.

CAUTION

Indicates a potentially hazardous situation 
which, if not avoided, may result in minor 
or moderate injury, or property damage.

Electrical Shock Caution

Indicates possibility of electric shock under specific 
conditions.

General Prohibition

Indicates non-specific general prohibitions.

Disassembly Prohibition

Indicates prohibitions when there is a possibility of 
injury, such as from electric shock, as the result of 
disassembly.

General Caution

Indicates non-specific general cautions, warnings, 
and dangers.

Precaution for 

Safe Use

Precautions on what to do and what not to 
do to ensure safe usage of the product.

Precaution for 

Correct Use

Precautions on what to do and what not to do 
to ensure proper operation and performance.

Meaning of Product Safety Symbols

Mounting to DIN Track

Dismounting from the DIN Track

Pull out the DIN Track 
mounting hook with a 
flat-blade screwdriver 
and lift the Unit from the 
bottom to remove it. 
The K7GE-MG can be 
easily installed and 
removed if the distance 
to other devices is 30 
mm or more.

(1) Connect the Units together.

(5) Install an End Plate (sold 

separately) on each side of the 
connected K7GE-MG without 
any gap. Up to one Main Unit 
and eight Probe Units can be 
installed between End Plates.

Applicable DIN Track

PFP-100N (100 cm)
PFP-50N (50 cm)

Conformance to Safety Standards

Conformance to EN/IEC Standards

Insulation Resistance 
Monitoring Device

Precaution for Correct Use

Thank you for purchasing the OMRON Product.
Please observe the following items when using the Product.

Before using the Product, thoroughly read and understand 

this manual to ensure correct use.

This Product is designed for use by qualified personnel with 

a knowledge of electrical systems.

Be sure to read the K7GE-MG User's Manual (Cat. No. 

N224) before using the Product.

Keep this manual in a safe location so that it is available for 

reference when required.

The appearance in which the Main Unit and the Probe Unit 

are connected is called K7GE-MG.

If you have any questions, consult your dealer or OMRON 
representative.

Mounting method

(2) Pull out all DIN Track 

mounting hooks on the 
bottom of the K7GE-MG.

(3) To mount the K7GE-MG to 

a DIN Track, hook the upper 
hook of K7GE-MG onto the 
DIN Track and press the 
Units in the direction of the 
arrow.

(4) Raise all DIN Track mounting 

hook and fix the K7GE-MG 
to the DIN Track.

©All Rights Reserved

5641253-5A (Side-B)

CAUTION

End
Plate

End
Plate

Probe Unit

(K7GE-MG1)

Main Unit

(K7GE-MGM)

Before you mount on the DIN Track, be sure to connect the 
units.

30 mm

above

Main Power Supply System Configurations

TT

3-phase, 4-wire type
(neutral point grounding)

Phase voltage/
line voltage
277 V/ 480 V

347 V/ 600 V in
addition to the above

480 V in addition to
the above

480 V in addition to
the above

480 V in addition to
the above

Phase voltage/
line voltage
240 V/ 480 V

Line voltage
300 V

Same as above

Line voltage
300 V

Line voltage
240 V

3-phase,
3-wire type
(no grounding)

Single-phase,
2-wire type
AC or DC

Single-phase
(split phase),
2-wire type
AC or DC

3-phase,
3-wire type
(single-phase
 grounding)

TN-C-S

P1
P2

N

P3

E

E

L1
L2

N

L3

PE

PEN

P1

P2
P3

E

E

P1

P2
P3

E

E

L1

L2

E

E

L1

L2

E

E

N

CAT III

CAT II

TT:
Connect E in the diagram 
above to the No. 7 terminal 
(PE) of the K7GE-MGM. 
Connect P1 and P2 to the 
No. 1 and No. 3 terminals 
of the K7GE-MG1.

Connect E in the diagram above to the No. 7 terminal 
(PE) of the K7GE-MGM.
Connect P1 and P2 to the No. 1 and No. 3 terminals 
of the K7GE-MG1.

Connect E in the diagram above to the No. 7 terminal 
(PE) of the K7GE-MGM.
Connect L1 and L2 to the No. 1 and No. 3 terminals of 
the K7GE-MG1.

TN-C-S:
Connect PE in the diagram 
above to the No. 7 terminal 
(PE) of the K7GE-MGM. 
Connect L1 and L2 to the 
No. 1 and No. 3 terminals 
of the K7GE-MG1.

Fixed
equipment

Distribution
board

CAT IV

CAT III

CAT II

CAT I

Power

outlet

Internal wiring

Service
entry wire

The measurement category classifies the places and equipment 
which you can connect to the measurement terminals, as 
prescribed in EN/IEC61010-2-030.
Each category is as follows.
CAT I:  Equipment to connect to circuits where measures are 

taken to limit transient overvoltages to low levels

CAT II:  Energy-consuming equipment with an energy supply from 

fixed wiring equipment (such as a power outlet)

CAT III:  Equipment in fixed wiring equipment that particularly 

demands equipment reliability and effectiveness

CAT IV:  Equipment to use at the electrical service entry

OMRON EUROPE B.V. (Importer in EU)

Wegalaan 67-69, NL-2132 JD Hoofddorp The Netherlands
 Phone 31-2356-81-300  FAX 31-2356-81-388

OMRON ELECTRONICS LLC

2895 Greenspoint Parkway, Suite 200 Hoffman Estates, IL 60169 U.S.A.  
 Phone 1-847-843-7900  FAX 1-847-843-7787

OMRON ASIA PACIFIC PTE. LTD.

No. 438A Alexandra Road # 05-05/08 (Lobby 2),
Alexandra Technopark, Singapore 119967
 Phone 65-6835-3011

 FAX 65-6835-2711

OMRON Corporation (Manufacturer)

Shiokoji Horikawa, Shimogyo-ku, Kyoto 600-8530 JAPAN

Wire type

Wiring

material

Recommended

wires

Stripping length

without ferrules

Solid or 
Stranded wires

Copper

0.25 to 1.5 mm²

AWG 24 to AWG 16

8 mm

Connection to measurement circuit

INSTRUCTION MANUAL

EN

OMRON Corporation

K7GE-MG

Measurement category

Specifications

Power supply voltage and frequency

K7GE-MGMA: 100 to 240 VAC, 50/60 Hz
K7GE-MGMD: 24 VAC, 50/60 Hz, 24 VDC

Operating voltage range

85% to 110% of the rated voltage

Operating frequency range

45 to 65 Hz

Power consumption

Maximum Unit configuration: one Main Unit and eight Probe Units
12.9 VA max. (100 to 240 VAC)/
7.8 VA max. (24 VAC)/ 4.7 W max. (24 VDC)
Minimum Unit configuration: one Main Unit and one Probe Unit
8.4 VA max. (100 to 240 VAC)/
4.6 VA max. (24 VAC)/ 2.5 W max. (24 VDC)

Ambient operating temperature

-10 to 55°C (with no condensation or icing)

Ambient operating humidity

25% to 85% (with no condensation)

Storage temperature

-20 to 65°C

Altitude

2,000 m max.

Recommended fuse

T2A, time delay, high-breaking capacity
(for Main Unit operating power supply)
Tripping current: 7 A max., fast-blow
(for Probe Unit voltage input)

Vibration resistance Frequency

10 to 55 Hz, 0.35-mm single amplitude, acceleration 50 m/s², 
10 sweeps of 5 min each in X, Y, and Z directions

Shock resistance

100 m/s², 3 times each in X, Y, and Z axes, 6 directions 

Degree of protection 

IP20

Terminal block type

Push-In Plus

Exterior color

Black (Munsell N 1.5)

Mounting

DIN Track mounting

Weight

Main Unit: Approx. 156 g
Probe Unit: Approx. 63 g

Measurement range

0.1 to 99.9 M

Ω

 (0.0 M

Ω

 for less than 0.1 M

Ω

)

Measurement accuracy

±5% rdg. ±1 digit
(at ambient temperature -10 to 55°C and ambient humidity 
25% to 65%)

Installation environment

Operation voltage:
EN/IEC61010-1, Overvoltage category II, Pollution Degree 2
Measurement circuit:
EN/IEC61010-2-030, Pollution Degree 2
Measurement category conforms to Conformance to Safety 
Standards.

Industrial electromagnetic environment

EN/IEC61326-1

Wiring material