Troubleshooting
Symptom
18
Cause
Correction
No display appears
when the START/STOP
button is pressed.
Batteries are flat.
Replace with new
batteries.
Batteries were inserted
incorrectly.
Insert the batteries with
the correct [+] [-]
polarity.
Cannot measure or
readings are too high.
Are you holding the
wrist cuff at heart level?
Measure while in the
correct posture.
Is the cuff wrapped
snugly around the wrist? Wrap the cuff correctly.
Are your arms and
shoulders tense?
Relax and try taking the
measurement again.
Have you been talking
or moving your hands
during measurement?
Keep still and do not talk
during measurement.
The blood pressure is
different each time. The
reading is extremely
low (or high).
Blood pressure readings constantly vary with time
of measurement and psychological condition. Take
deep breaths to relax before taking a
measurement.
EN[IW2]-2 3/14/05 5:22 PM Page 18