background image

Automatic Blood Pressure Monitor

and Print-Out 

with 

IntelliSense™

Model

HEM-705CP

INSTRUCTION MANUAL

R

Summary of Contents for INTELLISENSE HEM-705CP

Page 1: ...Automatic Blood Pressure Monitor and Print Out with IntelliSense Model HEM 705CP INSTRUCTION MANUAL R ...

Page 2: ......

Page 3: ...to Set Time and Date 9 How to Connect Printer Unit 10 How to Load Printer Paper 11 How to Use the Optional AC adapter 12 Correct Usage 13 How to Apply Arm Cuff 13 14 How to Take a Reading 15 16 How to Use the Memory Function 17 How to Use the Printer Unit 18 19 20 Troubleshooting 21 22 Caution 23 Care and Maintenance 24 Warranty 25 Specifications 26 27 3 ...

Page 4: ...n with every reading Arm discomfort and misreading due to improper cuff inflation are virtually eliminated Your new OMRON monitor uses the oscillometric method of blood pressure measurement This means the monitor detects your blood s movement through your brachial artery and converts the movements into a digital reading At the push of a button the monitor automatically fills the cuff with air star...

Page 5: ... long period of time approximately three months or more take out the batteries Replace the worn batteries with new ones immediately Do not use worn and new batteries together Do not insert the batteries with their polarities in wrong direction Do not force to bend the arm cuff or the air tube excessively When removing the air tube pull at the connector Do not apply strong shock or drop the main un...

Page 6: ...sure ON OFF Button AC Adapter Set Date Time Adjust Date Time Diastolic Blood Pressure Print All Data Print Current Data Stop Printing Printer Cover Air Plug Connects to Air Jack Main Unit Printer Unit Accessories AC Adapter Jack Printer Jack Cuff Jack Printer Cable Plug Power Operation Light Applicable arm cuff for Adults 9 to 13 in not a standard accessory ...

Page 7: ...e cuff is above brachial artery 5 Pull the end of the cuff so that the entire cuff is evenly tightened around your arm and press the hook material firmly against the pile side of the cuff 6 Press the ON OFF button 7 After the Heart Symbol appears on the digital panel press the Start button and remain still until the measurement is complete 8 When measurement is complete the monitor displays your b...

Page 8: ...e and negative polarities match the polarities of the battery compartment as indicated 3 Replace the battery cover Notes If the Low Battery Indicator appears on the display replace all four batteries Long life alkaline batteries are recommended Remove the batteries if the monitor will not be used for an extended period of time Main Unit Printer Unit ...

Page 9: ... button The year digits 2001 will flash 3 Press the ADJUST button to advance the digit s one at a time If you hold down the ADJUST button the digits will advance rapidly 4 Press SET button when desired number is on the display to lock setting 5 Repeat Step 3 4 for month and date 6 Repeat Step 3 4 for hour and minutes ...

Page 10: ...10 HOW TO CONNECT PRINTER UNIT 1 Remove printer cable plug from around printer unit 2 Connect printer cable plug cord to the main unit Arrow on printer cable plug should face down to main unit ...

Page 11: ...ting paper into the paper feeding slot When the edge of the paper protrudes sufficiently release the PAPER FEED button 4 Close the plastic printer cover Make certain that the paper sticks out from the top when the cover is replaced 5 Thermal sensitive paper is supplied with the unit Use only thermal paper of similar size width approx 38mm and roll diameter not to exceed approx 26mm These rolls of ...

Page 12: ...er plug into the AC Adapter jack on the back of the main unit 2 Plug in the AC adapter into a 120V AC outlet 50 60 cycles To remove the AC adapter To remove the AC adapter disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor s jack NOTE The monitor is designed not to draw power from the batteries when the AC adapter is in use 1 2 ...

Page 13: ...of the cuff loop 3 Sit in a chair with your feet flat on the floor and place your arm on a table so that the cuff is at the same level as your heart 4 Put your arm through the cuff loop The bottom of the cuff should be approximately 1 2 above the elbow The Green Marker on the cuff should lie over the brachial artery on the inside of the arm Tube should run down center of arm approximately even wit...

Page 14: ...he sewn hook material FIRMLY against the pile side of the cuff 7 Make certain the cuff fits snugly around your arm The cuff should make good contact with your skin You should be able to fit your index finger between the cuff and your arm easily so you can pull the cuff off and on 8 Relax your arm and turn your palm upward 9 Be sure there are no kinks in the air tubing ...

Page 15: ...ally and measurement is started As the cuff slowly deflates decreasing numbers appear on the display and the Heart Symbol flashes at every heartbeat In rare circumstances a higher inflation may be necessary In those cases the monitor reinflates the cuff up to 30 mmHg higher than initial inflation and restarts the measurement 4 When the measurement is complete the arm cuff completely deflates and y...

Page 16: ...l the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure NOTE Do not apply more pressure than necessary The monitor will not inflate above 300 mmHg If you want to stop the measurement push the ON OFF button The monitor will stop inflating and start deflating rapidly then the monitor will turn off Instructions for special conditions ...

Page 17: ...all stored readings while the Heart Symbol is displayed When twenty eight 28 sets of readings are stored in memory the oldest set will be deleted to store a new set START Press these two buttons at the same time Date and Time of stored readings will be alternately displayed To delete stored readings press the button and the Start button simultaneously then all the readings will be deleted NOTE You...

Page 18: ...nit does NOT have to be ON to operate printer unit Printer unit will automatically shut OFF when printing is completed If you need to stop printing during printing process press DATA STOP button 1 Press DATA STOP to print measurement of most recent reading The numeric print out will look like the following example 7 9 6 36 AM SYS 111mmHg DIA 67mmHg PULSE 77 min Month Date Systolic Reading Diastoli...

Page 19: ...all stored readings Month s The month or months will be printed Example If the blood pressure reading was stored between months March June then the printout will indicate 3 6 Date Only the day s date 1 31 will be printed Example If the blood pressure reading was stored on June 1st then the printout will indicate 1 under the D column Time Hour Only the hour in military time will be printed Example ...

Page 20: ...s a maximum of 28 readings 50 100 150 200 mmHg 3 6 Diastolic Reading Systolic Reading Average systolic reading of stored measurements Average diastolic reading of stored measurements Last reading stored Month s If reading is lower than 50 mmHg or higher than 200 mmHg will appear Most recent reading ...

Page 21: ...and take another measurement Repeat steps listed under How to Take a Reading If EE is displayed again take another measurement by pressing and holding the START button as shown on instruction page 12 under How to Take a Reading instructions for special conditions Your cuff maybe wrapped too tightly around your arm Turn the monitor off carefully read and repeat steps listed under How to Take a Read...

Page 22: ...p 10 Contact Omron Customer Service Take paper jammed from the printer and install again Purchase new printer paper Printer Unit TROUBLESHOOTING Batteries are worn No readings stored Printer paper is not installed correctly Replace with new batteries Take measurement then print Install correctly Batteries are worn Power Operation light is blinking stopped printing Replace with new batteries Nothin...

Page 23: ...on reception which can be determined by turning the product on and off the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and receiver Connect the product into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an ex...

Page 24: ...res humidity moisture and direct sunlight Folding the cuff and tubing tightly Inflating the monitor over 280 mmHg Disassembling the monitor Subjecting the monitor to strong shocks for example dropping the monitor on the floor Do not clean the monitor with volatile liquids THE MONITOR SHOULD BE CLEANED WITH A SOFT DRY CLOTH Do not clean cuff ...

Page 25: ...fects in the materials or workmanship The above warranty is complete and exclusive The warrantor expressly disclaims liability for incidental special or consequential damages of any nature Some states do not allow the exclusion or limitation of incidental or consequential damages so the above warranty may not apply to you Any implied warranties arising by the operation of law shall be limited in d...

Page 26: ...Wave Detection Capacitive pressure sensor Power Source 4 AA batteries or AC adapter Battery Life Approximately 300 uses Operating 50 F to 104 F 10 C to 40 C Temperatures Humidity 30 to 85 RH maximum Storage 4 F to 140 F 20 C to 60 C Temperatures Humidity 10 to 95 RH maximum Main Unit Weight Approximately 13 4 oz 380g not including batteries Main Unit Dimensions Approximately 4 1 2 l x 7 w x 2 4 5 ...

Page 27: ...current data Printer Weight Approximately 6 oz 170 g not including batteries Printer Dimensions Approximately 4 4 5 l x 2 4 5 w x 2 h 123 mm x 72 mm x 50 mm NOTE These specifications are subject to change without notice Optional accessories large cuff fits arms 13 to 17 in circumference and small cuff fits arms 7 to 9 in circumference are sold separately ...

Page 28: ...vice Call Toll Free 1 800 634 4350 Manufactured by OMRON Distributed by OMRON HEALTHCARE INC 300 Lakeview Parkway Vernon Hills Illinois 60061 www omronhealthcare com Copyright 2001 OMRON HEALTHCARE INC 705CPINST Rev A ...

Reviews: