3.4 Turning “OFF” Bluetooth

Turn off the 

Bluetooth

 in your monitor in the following areas where use of 

wireless equipment is prohibited.
• On aircraft  • In hospitals  • While abroad
Your blood pressure monitor comes with the 

Bluetooth

 already turned on.

1.

Turn your monitor power off if it is turned on.

2.

Press and hold the connection button for more than 10 
seconds.

3.

Release the connection button when “

” appears on 

the display.

4.

Press the [START/STOP] button to turn the monitor off.

Note: The monitor will automatically turn off after 2 minutes.

How to turn “ON” Bluetooth

Operate the following step when the power has been off.
Press and hold the connection button for more than 2 seconds and 
then release the connection button when 

 appears on the 

display.

3.5 Restoring Your Monitor to the Default Settings

To delete all the information stored in your monitor, follow the instructions 
below. Make sure that your monitor is turned off. It is off when the screen is 
not showing numbers or symbols. 

1.

While holding the connection button down, the Connection 
symbol “ ” and “P” appear on the display, and then press 
the [START/STOP] button for more than 5 seconds.

2.

When the “Clr” symbol appears on the display, release the 
connection button and [START/STOP] button.

The “Clr” will appear on the display when your monitor settings have 
reverted back to their original.

3.

Press the [START/STOP] button to turn the monitor off.

Notes:
• The monitor will automatically turn off after 2 minutes.
• Reverting to initial settings of this monitor will not delete the information 

in the app.

• When using this monitor again, please delete the registration of the 

monitor in the app first, and then pair your monitor.

4. Error Messages and Troubleshooting

4.1 Error Messages

4.2 Troubleshooting

5. Maintenance and Storage

5.1 Maintenance

To protect your device from damage, please follow the directions below:

• Store the device and the components in a clean, safe location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components or immerse them in water.
• Do not use gasoline, thinners or similar solvents to clean the device.

• Use a soft dry cloth, or a soft cloth moistened with neutral soap to clean 

on the monitor and the arm cuff, and then wipe it with a dry cloth.

• Changes or modification not approved by the manufacturer will void the 

user warranty. Do not disassemble or attempt to repair the device or 
components. 

Calibration and Service

• The accuracy of this blood pressure monitor has been carefully tested 

and is designed for a long service life.

• It is generally recommended to have the device inspected every 2 years 

to ensure correct functioning and accuracy. Please consult your local 
OMRON representatives.

5.2 Storage

Do not store the device in the following situations:

•  If the device is wet.
•  Locations exposed to extreme temperatures, humidity, direct sunlight, 

dust or corrosive vapors such as bleach.

•  Locations exposed to vibrations, shocks or where it will be at an angle.

6. Specifications

Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, K5 was used on 85 subjects for determination of 

diastolic blood pressure.

• This device is clinically investigated according to the requirements of 

ISO81060-2:2013.

• This device has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by enclosures in accordance 

with IEC 60529. This device is protected against solid foreign objects of 12 mm 
diameter and greater such as a finger, and against oblique falling water drops 
which may cause issues during a normal operation.

• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device 

Directive). 

• This blood pressure monitor is designed according to the European Standard 

EN1060, Non-invasive sphygmomanometers Part 1: General Requirements 
and Part 3: Supplementary requirements for electromechanical blood pressure 
measuring systems.

• This OMRON product is produced under the strict quality system of OMRON 

HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood 
pressure monitors, which is the Pressure Sensor, is produced in Japan.

7. Trademarks

The 

Bluetooth

®

 

word mark and logos are registered 

trademarks owned by Bluetooth SIG, Inc. and any use of 
such marks by OMRON HEALTHCARE Co., Ltd. is under 
license. Other trademarks and trade names are those of 
their respective owners.

Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. 

and other countries. App Store is a service mark of Apple Inc.

Android, the Google Play logo are trademarks of Google Inc.

8. Some Useful Information about Blood Pressure

What is Blood Pressure?

Blood pressure is a measure of the force of blood flowing against the walls of the 
arteries. Arterial blood pressure is constantly changing during the course of the 
heart’s cycle.
The highest pressure in the cycle is called the Systolic Blood Pressure; the 
lowest is the Diastolic Blood Pressure. Both pressures, the Systolic and 
Diastolic, are necessary to enable a doctor to evaluate the status of a patient’s 
blood pressure.

What is Arrhythmia?

Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in 

the bio-electrical system that drives the heartbeat. Typical symptoms are 
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or 
slow (bradycardia) pulse.

Why is it a Good Thing to measure Blood Pressure at 

Home?

Many factors such as physical activity, 
anxiety, or the time of day, can influence 
your blood pressure. A single 
measurement may not be sufficient for an 
accurate diagnosis.
Thus it is best to try and measure your 
blood pressure at the same time each 
day, to get an accurate indication of any 
changes in blood pressure. Blood pressure is typically low in the morning and 
increases from afternoon to evening. It is lower in the summer and higher in the 
winter.

Error Display

Cause

Solution

Irregular heartbeat is 

detected.

Remove the arm cuff. Wait 2 

- 3 minutes and then take 

another measurement. 

Repeat the steps in section 

3.1. If this error continues to 

appear, contact your doctor.

Movement during 

measurement.

Carefully read and repeat the 

steps in section 3.3.

Arm cuff is applied 

too loosely.

Apply the arm cuff tighter.

Refer to section 3.1.

The batteries are 

low.

Recommend replacement of 

all 4 batteries with new ones.

Refer to section 2.1.

The batteries are 

depleted or battery 

polarities are not 

properly aligned. 

Immediately replace 4 

batteries with new ones.

Refer to section 2.1.

Confirm the batteries are 

properly inserted with 

polarities correctly aligned. 

Refer to section 2.1.

Connection failure. 

Data is not being 

transmitted.

Refer to “Connection failure. 

Data is not being 

transmitted.” in section 4.2.

Arm cuff is applied 

too loosely.

Apply the arm cuff tighter.

Refer to section 3.1.

The arm cuff was 

inflated exceeding 

the maximum 

allowable pressure, 

and then deflated 

automatically.

Do not touch the arm cuff 

while taking a measurement. 

Refer to section 3.3.

Movement during 

measurement.

Repeat measurement. 

Remain still and do not talk 

during measurement.

Refer to section 3.3.

Clothing is interfering 

with the arm cuff.

Remove any clothing 

interfering with the arm cuff.

Refer to section 3.1.

If the number is 

shown except 1 to 5.

Device error.

Contact your local OMRON 

representative.

(2 sec.)

(8 sec.)

Problem

Cause and Solution

No power.

No display appears on the 

monitor.

Replace all batteries with new ones.

Check the battery installation for proper 

placement of the battery polarities.

Refer to section 2.1.

Measurement values appear 

too high or too low.

Blood pressure varies constantly. Many 

factors including stress, time of day, and how 

you apply the arm cuff, may affect your blood 

pressure. 

Review the section, 3.2 and 3.3.

Connection failure. / 

Data is not being transmitted.

The blood pressure monitor might not be 

properly placed within the smart device's 

transmission range and is too far from the 

smart device. If there are no causes of data 

transmission interference found near the 

blood pressure monitor, move the blood 

pressure monitor within 5 m (16 ft.) of the 

smart device and try again.

The 

Bluetooth

 feature of the smart device is 

turned off. Turn on the 

Bluetooth

 feature of 

the smart device and your monitor, and try 

sending the data again.

The 

Bluetooth

 feature of the blood pressure 

monitor is turned off. 

Refer to “To turn “ON” the 

Bluetooth

” in “3.4 

Turning “OFF” the 

Bluetooth

”.

The blood pressure monitor did not pair 

successfully to the smart device. Try to pair 

the devices again. 

Refer to “Pair Your Monitor with a Smart 

Device” in section 2.2.

For more detail, go to “Help” in the “OMRON 

connect” app.

The application on the smart device is not 

ready. Check the application then try sending 

the data again. 

Refer to “Pair Your Monitor with a Smart 

Device” in section 2.2. If the “Err” symbol still 

displays after checking the application, 

contact your local OMRON representative. 

Confirm your smart device compatibility with 

this monitor at omronconnect.com

Product category 

Electronic Sphygmomanometers

Product description

Upper Arm Blood Pressure Monitor

Model

HEM-7600T

Display

OLED display

Cuff pressure range

Pressure: 0 to 299 mmHg

Measurement range

Pressure: 40 to 260 mmHg
Pulse: 40 to 180 beats / min.

Accuracy

Pressure: ±3 mmHg
Pulse: ±5% of display reading

Inflation

Fuzzy-logic controlled by electric pump

Deflation

Automatic rapid deflation

Measurement method

Oscillometric method

Transmission method

Bluetooth

 

low energy technology

Wireless communication

Frequency range: 2.4 GHz (2400 - 2483.5 MHz)
Modulation: GFSK
Effective radiated power: <20 dBm

IP classification

IP 22

Power source

4 “AAA” batteries 1.5V

Battery life

Approximately 300 measurements

(using new alkaline batteries)

Durable period (Service life)

Monitor: 5 years

Operating conditions

10°C to 40°C / 15 to 90% RH (non-condensing) / 
800 to 1060 hPa

Storage / Transport 
conditions

-20°C to 60°C / 10 to 90% RH (non-condensing) 

Weight

Monitor: Approximately 240 g

(not including batteries)

Dimensions

Monitor: Approximately 85 mm x 120 mm x 20 mm

(not including the arm cuff)

Arm circumference

22 to 42 cm

Contents

Monitor, battery set, instruction manual, setup instructions

Applied part

Type BF

Protection against electric 
shock 

Internally powered ME equipment 

®

Symbol descriptions

Please consult this instruction manual

Applied part - Type BF

Degree of protection against electric shock (leakage 

current)

To indicate generally elevated, potentially hazardous, 

levels of non-ionizing radiation, or to indicate 

equipment or systems e.g. in the medical electrical 

area that include RF transmitters or that intentionally 

apply RF electromagnetic energy for diagnosis or 

treatment.

Temperature limitation

Humidity limitation

Atmospheric pressure limitation

Artery mark

About a wireless communication interference

This Product operates in the unlicensed ISM band at 2.4GHz. In case this 
Product is used around the other wireless devices including microwave and 
wireless LAN, which operate same frequency band of this Product, there is a 
possibility that interference occurs between this Product and such other 
devices. If such interference occurs, please stop the operation of other 
devices or relocate this Product before using this Product or do not use this 
Product around the other wireless devices.

Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates 
that it should not be disposed of, with other household wastes 
at the end of its working life. To prevent possible harm to the 
environment or human health from uncontrolled waste 
disposal, please separate this product from other types of 
wastes and recycle it responsibly to promote the sustainable reuse of material 
resources.

Household users should contact either the retailer where they purchased this 
product, or their local government office, for details of where and how they 
can return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and 
conditions of the purchase contract. This product should not be mixed with 
other commercial waste for disposal. 

or

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, KYOTO,
617-0002 JAPAN

EU-representative

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp, 
THE NETHERLANDS

Asia Pacific HQ

OMRON HEALTHCARE SINGAPORE PTE LTD.

438A Alexandra Road, #05-05/08,

Alexandra Technopark, Singapore 119967

www.omronhealthcare-ap.com

Production facility

OMRON HEALTHCARE Co., Ltd.
Matsusaka Factory

1855-370, Kubo-cho, Matsusaka-shi,
Mie, 515-8503 Japan

Made in Japan

Example: fluctuation within a day (male, 35 years old)

mmHg

200

150

100

50

6

12

18

24

Upper curve: systolic blood pressure
Lower curve: diastolic blood pressure

Time of day

HEM-7600T-AP3_A_M03_170119.fm  2 

  

Summary of Contents for HEM-7600T

Page 1: ...otion trembling shivering may affect the measurement reading Consult your doctor before using this monitor if you have had a mastectomy Consult your doctor before using the monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising Do not take measurements more often than necessary because bruising due to blood flow interference may occur Remove the arm ...

Page 2: ...he data When the SYNC symbol appears on the display transfer the data before the measurements are deleted If the memory is full the monitor will be delete the oldest value Maximum internal memory is for 100 measurements 2013 ESH ESC Guidelines for the management of arterial hypertension Definitions of hypertension by office and home blood pressure levels These are from statistical values for blood...

Page 3: ...smart device will set the date and time on this monitor Once you have paired with the OMRON connect app the date and time is now setup on this monitor all future measurements will include the date and time when you transfer If you do not pair this monitor to set the date and time anyࠉ measurements you take with this monitor will not include the date and time If you use an app other than the OMRON ...

Page 4: ...being transmitted in section 4 2 for more detail Notes When the SYNC symbol appears on the display transfer the data immediately before the readings are deleted When the Connection OFF symbol appears on the display turn on the Bluetooth feature of your monitor Refer to section 3 4 Once you send the your readings to the app the data cannot be resent to app or smart device 4 After your readings have...

Page 5: ...settings of this monitor will not delete the information in the app When using this monitor again please delete the registration of the monitor in the app first and then pair your monitor 4 Error Messages and Troubleshooting 4 1 Error Messages Error Display Cause Solution Irregular heartbeat is detected Remove the arm cuff Wait 2 3 minutes and then take another measurement Repeat the steps in sect...

Page 6: ...d pressure monitor within 5 m 16 ft of the smart device and try again The Bluetooth feature of the smart device is turned off Turn on the Bluetooth feature of the smart device and your monitor and try sending the data again The Bluetooth feature of the blood pressure monitor is turned off Refer to To turn ON the Bluetooth in 3 4 Turning OFF the Bluetooth The blood pressure monitor did not pair suc...

Page 7: ...he medical electrical area that include RF transmitters or that intentionally apply RF electromagnetic energy for diagnosis or treatment Temperature limitation Humidity limitation Atmospheric pressure limitation Artery mark About a wireless communication interference This Product operates in the unlicensed ISM band at 2 4GHz In case this Product is used around the other wireless devices including ...

Page 8: ...es the heartbeat Typical symptoms are skipped heartbeats premature contraction an abnormally rapid tachycardia or slow bradycardia pulse Why is it a Good Thing to measure Blood Pressure at Home Many factors such as physical activity anxiety or the time of day can influence your blood pressure A single measurement may not be sufficient for an accurate diagnosis Thus it is best to try and measure yo...