Automatic Digital Blood Pressure Monitor

with 

IntelliSense™

Model 

HEM-757

INSTRUCTION MANUAL

SYSTOLIC

DIASTOLIC

PULSE

/min

DIGITAL BLOOD PRESSURE MONITOR 

WITH FUZZY LOGIC HEM-757

READY TO 

MEASURE

REPLACE 

BATTERIES

PROGRESS

ON/OFF

START

R

Summary of Contents for HEM-757

Page 1: ...tal Blood Pressure Monitor with IntelliSense Model HEM 757 INSTRUCTION MANUAL SYSTOLIC DIASTOLIC PULSE min DIGITALBLOODPRESSUREMONITOR WITHFUZZYLOGICHEM 757 READY TO MEASURE REPLACE BATTERIES PROGRESS ON OFF START R ...

Page 2: ...llent sensing technology the monitor can achieve accurate measurement Introduction 1 Know Your Unit 2 Quick Reference Guide 3 Battery Installation Replacement 4 Suggestions Before Blood Pressure Measurement 5 How to Apply the Arm Cuff 6 7 How to Take a Reading 8 9 10 How to Use the Memory Function 11 How to Use the Optional AC Adapter 12 Troubleshooting 13 Caution 14 Care and Maintenance 15 Five Y...

Page 3: ...comfort and misreading due to improper cuff inflation are virtually eliminated Your new OMRON monitoruses the oscillometric method of blood pressure measurement This means the monitor detects your blood s movement through your brachial artery and converts the movements into a digital reading At the push of a button the monitor automatically fills the cuff with air and displays your systolic diasto...

Page 4: ...TO MEASURE REPLACE BATTERIES PROGRESS ON OFF START START Button Arm Cuff ON OFF Button Air Jack Air Plug Connects to the air jack Green Cuff Marker Diastolic Blood Pressure Display Systolic Blood Pressure Display Carrying Case IntelliSense Progress Indicator 2 Air Tube Pulse Display Memory Call Button ...

Page 5: ...mediately deflate the cuff bladder and readjust the cuff if skin becomes pinched 4 Sit in a chair with your feet flat on the floor and place your left arm on a table so that the cuff is at the same level as your heart 5 Press the ON OFF button 6 After the Heart Symbol appears next to a zero on the digital panel press the START button 7 Remain still throughout the entire reading 8 When measurement ...

Page 6: ... and negative polarities match the polarities of the battery compartment as indicated 3 Replace the battery cover 4 If the Low Battery Indicator appears on the display replace all four batteries Long life alkaline batteries are recommended 5 Remove the batteries if the monitor will not be used for an extended period of time Low Battery Indication ...

Page 7: ...he cuff is at the same level as your heart 5 Remain still and do not talk during the measurement 6 Keep a record of your blood pressure and pulse measurements for your doctor Remember a single measurement does not provide an accurate indication of your true blood pressure You need to take and record several measurements over a period of time Try to measure your blood pressure at the same time each...

Page 8: ... loop 3 If the cuff is assembled correctly the sewn hook material material will be on the outside of the cuff loop and the metal D ring will not touch your skin 4 Put your left arm through the cuff loop The bottom of the cuff should be approximately 1 2 above the elbow The cuff tubing should be positioned along the middle of the inside of your arm as shown 5 Pull the cuff so that the top and botto...

Page 9: ... cuff should be wrapped so that it fits snugly and stays in place You should be able to fit one finger between the cuff and your arm 8 Sit in a chair with your feet flat on the floor and rest your arm on a table so that the cuff is at the same level as your heart 9 Relax your arm and turn your palm upward 10 Be sure there are no kinks in the air tubing ...

Page 10: ... When the monitor completes the necessary preparations before measurement the Heart Symbol appears next to a zero NOTE Wait for the Heart Symbol before taking a measurement 2 Press the START button As the cuff begins to inflate the IntelliSense monitor automatically determines your ideal inflation level During this process the IntelliSense symbol will be flashing on the display Because this monito...

Page 11: ... process A mark indicates the measurement process During the inflation a mark turns on During the deflation a mark turns on and one each mark turns off with the progress of approximately every 1 3 of the process 4 When the measurement is complete a Heart Symbol flashes and the arm cuff deflates 5 At the end of the measurement your blood pressure and pulse are alternately displayed 6 Press the ON O...

Page 12: ...higher than the previous inflation value Reinflation occurs only once 2 If your systolic pressure is known to be more than 220 mmHg push and hold the START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure NOTE Do not apply more pressure than necessary 3 If you want to stop the measurement push the ON OFF button The monitor will stop inflating and start d...

Page 13: ... measurement values can be called out For the fifteenth measurements and after the values from the oldest measurement will be deleted START Previous value Value before the previous measurement Value of fourteen measurements before Push these two buttons at the same time Display of memory deletion If you want to delete all the stored data You can delete all the stored data while the mark is display...

Page 14: ...r 3 Insert the AC adapter plug into a 120V AC outlet 50 60 cycles 4 To remove the AC adapter disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor s jack To avoid possible damage to the monitor use only the exclusive AC adapter Model HEM ADPT1 available from Omron Healthcare Inc Other adapters may vary in output voltage and polarities AC Adapter Output ...

Page 15: ...g Battery voltage is excessively low Wait 10 minutes before taking another measurement Repeat steps listed under How to Take a Reading If EE is displayed again take another measurement by pressing and holding the START button as listed in step 2 under How to Take a Reading Carefully read and repeat steps listed under How to Take a Reading Replace all four AA batteries with new batteries Trouble ca...

Page 16: ...r television reception which can be determined by turning the product on and off the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and receiver Connect the product into an outlet on a circuit different from that to which the receiver is connected Consult the deale...

Page 17: ...temperatures humidity and direct sunlight Folding the cuff and tubing tightly Inflating the monitor over 280 mmHg Disassembling the monitor Subjecting the monitor to strong shocks for example dropping the monitor on the floor Do not clean the monitor with volatile liquids THE MONITOR SHOULD BE CLEANED WITH A SOFT DRY CLOTH ...

Page 18: ...rge any parts necessary to correct defects in the materials or workmanship The above warranty is complete and exclusive The warrantor expressly disclaims liability for incidental special or consequential damages of any nature Some states do not allow the exclusion or limitation of incidental or consequential damages so the above warranty may not apply to you Any implied warranties arising by the o...

Page 19: ...ric method Pulse Wave Detection Capacitive pressure sensor Power Source 4 AA batteries or AC adapter sold separately Battery Life Approximately 300 uses Operating Temperature Humidity 50 F to 104 F 10 C to 40 C 30 to 85 RH maximum Storage Temperature Humidity 4 F to 140 F 20 C to 60 C 10 to 95 RH maximum Console Weight Approximately 19 oz 530 g not including batteries Outer Dimensions Approximatel...

Page 20: ...ustomer Service Call Toll Free 1 800 634 4350 Manufactured by OMRON Distributed by OMRON HEALTHCARE INC 300 Lakeview Parkway Vernon Hills Illinois 60061 www omron com ohi Copyright 2000 OMRON HEALTHCARE INC 000000 0X ...