EN4
EN
Display/Problem
Possible Cause
Solution
appears
The monitor has malfunctioned.
Press the [START/STOP] button again. If “Er” still appears,
contact your OMRON retail outlet or distributor.
flashes
Batteries are low.
Replacing all 4 batteries with new ones is recommended.
Refer to section 3 of instruction manual
.
appears or the
monitor is turned off
unexpectedly during
a measurement
Batteries are depleted.
Immediately replace all 4 batteries with new ones. Refer to
section 3 of instruction manual
.
Nothing appears on the display
of the monitor.
Battery polarities are not properly aligned. Check the battery installation for proper placement. Refer to
section 3 of instruction manual
.
Readings appear too high or
too low.
Blood pressure varies constantly. Many factors including stress, time of day, and/or how you apply the
arm cuff, may affect your blood pressure. Review section 2 of instruction manual
.
Any other problem occurs.
Press the [START/STOP] button to turn the monitor off, then press it again to take a measurement. If the
problem continues, remove all batteries and wait for 30 seconds. Then re-install batteries. If the problem
still persists, contact your OMRON retail outlet or distributor.
Troubleshooting for Afib indicator function:
What is different between the
Afib indicator function and
ECG?
The Afib indicator function and ECG use completely different technologies. An ECG measures the
electrical activity of the heart and can be used to diagnose Afib. The Afib indicator function detects
irregular heartbeat and can suggest the possibility of Afib with a sensitivity of 95.5 % and specificity of
93.8 %. Refer to section 11 for details.
If the “
” symbol does not
appear, it means there is no
possibility of Afib?
Even if the “
” symbol does not appear, there is still a possibility of Afib.
Should I consult with my
physician if the “
” symbol
appears?
We recommend you to consult with your physician because there is a possibility of Afib. However, the
“
” symbol may be displayed for other reasons, such as other heart arrhythmias.
What is different between Afib
indicator function and irregular
heart beat function?
The irregular heartbeat function detects irregularities in the pulse waves in one measurement. The Afib
indicator function suggests the possibility of Afib when blood pressure is measured 3 consecutive times.
What should I do if the “
”
symbol sometimes appears?
Afib does not always have symptoms. We recommend you to consult with and follow the directions of
your physician.