background image

Introduction

Thank you for purchasing the OMRON EVOLV Automatic Upper Arm Blood 
Pressure Monitor.
Your new automatic upper arm blood pressure monitor uses the oscillometric 
method of blood pressure measurement. This means the monitor detects your 
blood movement through your brachial artery and converts the movements into a 
digital reading. An oscillometric monitor does not need a stethoscope so the 
monitor is simple to use.

Intended Use

This device is a digital monitor intended for use in measuring blood pressure and 
pulse rate in adult patient population. The device detects the appearance of 
irregular heartbeats during measurement and indicates this via a symbol with 
readings. It is mainly designed for general household use. 

Important Safety Information

Please read the Important Safety Information in this instruction manual 
before using the unit.

Please follow this instruction manual thoroughly for your safety.
Please keep for future reference. For specific information about your own blood 
pressure, CONSULT YOUR PHYSICIAN.

Warning

Indicates a potentially hazardous situation which, if not 
avoided, could result in death or serious injury.

• DO NOT adjust medication based on measurement results from this blood 

pressure monitor. Take medication as prescribed by your physician. Only a 
physician is qualified to diagnose and treat high blood pressure.

• DO NOT use this monitor on an injured arm or an arm under medical treatment.
• DO NOT wear the arm cuff on your arm while on an intravenous drip or blood 

transfusion.

• DO NOT apply or inflate the arm cuff around any body parts other than your 

upper arm.

• DO NOT use this monitor in areas containing high frequency (HF) surgical 

equipment, magnetic resonance imaging (MRI) equipment, computerized 
tomography (CT) scanners or in oxygen rich environments. This may result in 
incorrect operation of the device and/or cause an inaccurate reading.

• DO NOT use this monitor on infants, toddlers, children or persons who cannot 

express themselves.

• Consult your physician before using this monitor if you have common arrhythmias 

such as atrial or ventricular premature beats or atrial fibrillation; arterial sclerosis; 
poor perfusion; diabetes; pregnancy; pre-eclampsia; or renal disease.

• Always consult your physician. Self-diagnosis of measurement results and 

self-treatment are dangerous.

• Contains small parts that may cause a choking hazard if swallowed by children, 

toddlers or infants.

• Keep the BATTERY out of the reach of children, toddlers and infants.
• This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this 

product in locations where RF is restricted, such as on an aircraft or in hospitals.

Caution

Indicates a potentially hazardous situation which, if not 

avoided, may result in minor or moderate injury to the 

user or patient or damage to the equipment or other 

property.

• Stop using this monitor and consult your physician if you experience skin 

irritation or discomfort.

• Consult your physician before using this monitor on an arm with an 

arterio-venous (A-V) shunt.

• Note that PATIENT motion, trembling, shivering may affect the measurement 

reading.

• Consult your physician before using this monitor if you have had a mastectomy.
• Consult your physician before using the monitor if you have severe blood flow 

problems or blood disorders as cuff inflation can cause bruising.

• Do not take measurements more often than necessary because bruising, due to 

blood flow interference, may occur.

• Remove the arm cuff if it does not start deflating during the measurement.
• Do not use this monitor for any purpose other than measuring blood pressure.
• During measurement, make sure that no mobile device or any other electrical 

device that emits electromagnetic fields is within 30 cm of this monitor. This may 
result in incorrect operation of the monitor and/or cause an inaccurate reading.

• Do not use the device with other medical electrical (ME) equipment 

simultaneously. This may result in incorrect operation of the device and/or cause 
an inaccurate reading.

• Do not disassemble or attempt to repair the monitor or components. This may 

cause an inaccurate reading.

• Do not use in a location where there is moisture or a risk of water splashing the 

monitor. This may damage the monitor.

• Do not use this monitor in a moving vehicle such as in a car or on an airplane.
• Do not drop, or subject this monitor to strong shocks or vibrations.
• Do not use this monitor in places with high or low humidity or high or low 

temperatures. See the “6. Specifications” section of the manual.

• Please check, by observation of the arm being used, that the monitor is not 

causing any impairment of patient blood circulation.

• Do not use this monitor in high-use environments such as medical clinics or 

physician offices.

• Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 

minutes before taking a measurement.

• Rest for at least 5 minutes before taking the measurement.
• Remove tight-fitting or thick clothing from your arm while taking a measurement.
• Remain still and do not talk while taking a measurement.
• Read and follow the “Important information regarding Electromagnetic 

Compatibility (EMC)” in the “6. Specifications”.

• Read and follow the “Correct Disposal of This Product” in “6. Specifications” 

when disposing of the device and any used accessories or optional parts.

• If this monitor is stored at the maximum or minimum storage and transport 

temperature and is moved to an environment with a temperature of 20°C, we 
recommend waiting for approximately 2 hours before using the monitor.

Data Transmission

• Do not replace the battery while your reading is being transferred to your smart 

device. This may result in the incorrect operation of your monitor and failure to 
transfer your blood pressure data.

• Do not place integrated circuit cards, magnets, metal objects, or other devices 

that emit electromagnetic fields near this monitor while your reading is being 
transferred to your smart device. This may result in the incorrect operation of 
your monitor and failure to transfer your blood pressure data.

Battery Handling and Usage

• Do not insert the batteries with their polarities incorrectly aligned.
• Use only 4 “AAA” alkaline batteries with this monitor. Do not use other types of 

batteries. Do not use new and used batteries together.

• Remove the batteries if the monitor will not be used for a long period of time.
• If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Consult a physician immediately.

• If battery fluid should get on your skin, wash your skin immediately with plenty of 

clean, lukewarm water. If irritation, injury or pain persists, consult a physician.

• Do not use batteries after their expiration date.
• Periodically check your batteries to ensure they are in good working condition.
• Only use batteries specified for this monitor. Use of unsupported batteries may 

damage and/or may be hazardous to the monitor.

1. Know Your Device

Contents:

Monitor, battery set, instruction manual, setup instructions, soft case

Monitor:

1.1 Display symbols

Irregular Heartbeat Symbol (

)

When the monitor detects an irregular rhythm 
two or more times during the measurement, the 
irregular heartbeat symbol will appear on the 
display with the measurement values. 
An irregular heartbeat rhythm is defined as a 
rhythm that is 25% less or 25% more than the 
average rhythm detected while the monitor is 
measuring the systolic and diastolic blood 
pressure. 
If the irregular heartbeat symbol displays with your measurement results, we 
recommend you consult your physician. Follow the directions of your 
physician.

Movement Error Symbol (

)

The movement error symbol is displayed if you move your body during the 
measurement. Please remove the arm cuff, and wait 2 - 3 minutes. 
Take another measurement and remain still during measurement.

Cuff Wrap Guide Symbol (

 / 

)

If the cuff was wrapped too loosely, it may cause unreliable results. If the 
wrapping of cuff is too loose, the cuff wrap guide symbol “

” appears. 

Otherwise “

” appears. This is the function which is used as an aid in 

determining if the cuff is wrapped snugly enough. 

SYNC Symbol (

)

• When the SYNC symbol appears on the display, make sure that your 

monitor has been pairing with the smart device or transfer the data.

• When the SYNC symbol appears on the display, transfer the data before 

the measurements are deleted. Maximum internal memory is for 
100 measurements. The oldest value will be deleted when this is reached 
before downloading.

2013 ESH/ESC Guidelines for the management of arterial hypertension

Definitions of hypertension by office and home blood pressure levels

These are from statistical values for blood pressure.

1.2 Before Taking a Measurement

To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and 

eating for 30 minutes before taking a measurement. 

2. Rest for at least 5 minutes before taking the measurement.
3. Stress raises blood pressure. Avoid taking measurements during stressful 

times.

4. Measurements should be taken in a quiet place.
5. Remove tight-fitting clothing from your arm.
6. Keep a record of your blood pressure and pulse readings for your 

physician. A single measurement does not provide an accurate indication 
of your true blood pressure. You need to take and record several 
readings over a period of time. Try to measure your blood pressure at the 
same time each day for consistency.

2. Preparation

2.1 Battery Installation

1.

Remove the battery cover.

Gently slip into the gap of the hook 
and pull downward.

2.

Insert 4 “AAA” batteries as 
indicated, into the battery 
compartment.

3.

Replace the battery cover.

When the batteries have been securely placed in this device, the message 
and SYNC symbol below will appear on the screen:

Notes:
• When the depleted battery symbol “

” appears on the display, turn the 

monitor off and remove all the batteries. Replace with 4 new batteries at 
the same time. Long life alkaline batteries are recommended.

• The measurement values continue to be stored in the memory even after 

the batteries are replaced.

• The supplied batteries may have a shorter life than new batteries.
• Disposal of used batteries should be carried out in accordance with the 

national/local regulations for the disposal of batteries.

2.2 Pairing Your Monitor with a Smart Device

The date and time on your monitor will automatically be set when you pair it 
with your smart device.
To begin using the “OMRON connect” app for the first time, please visit 
omronconnect.com/setup for the initial set-up instructions.

1.

Turn on the 

Bluetooth

®

 feature of the smart device.

2.

Download and install the free “OMRON connect” app on your 
smart device.

* If you already have the “OMRON connect” app, go to:

Menu > Devices > Add Device

3.

Open the app on your smart device and follow the set-up and 
pairing instructions.

4.

Confirm the monitor is connected successfully.

If the monitor is connected successfully to the smart device, “OK” will 
appear on the display.

If “Err” appears, refer to “Connection failure. / Data is not being 
transmitted.” in section 4.2 for more detail.

Notes: 
• If you use any other app instead of “OMRON connect”, the date and time 

may not be automatically set.
If you need to set the date and time, pair your monitor or transfer the data 
by using “OMRON connect” app.

• The date/time will only be set for future measurements taken after the 

measurement that was transferred, it will not be saved to the measurement 
that was just transferred.

• Please be aware that OMRON will not be responsible for the loss of data 

and/or information.

• If you use the monitor without pairing, refer to the “Using your Monitor 

without Pairing” in the “3.3 Taking a Measurement”.

3. Using the Device

3.1 Applying the Arm Cuff

Remove tight-fitting clothing from your left upper arm. 
Do not place the arm cuff over thick clothes.

1.

Apply the arm cuff to your left upper 
arm.

The bottom edge of the arm cuff should be 1 to 
2 cm above the elbow. Positioning the display in 
clear view of the user makes it easier to read 
the results.

2.

Securely close with the fabric 
fastener.

Notes:
• The blood pressure can differ between 

the right arm and the left arm, and the 
measured blood pressure values can be 
different. Omron recommends to always use the same arm for 
measurement. If the values between both arms differ substantially, please 
check with your physician as to which arm to use for your measurements.

• Positioning the monitor with the [START/STOP] button towards the body 

makes it easier to use the device.

3.2 How to Sit Correctly

To take a measurement, you need to be relaxed and 
comfortably seated, at a comfortable room 
temperature. 

• Sit in a chair with your legs uncrossed and your feet flat on 

the floor.

• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at the same 

level as your heart.

3.3 Taking a Measurement

Notes:
• To stop the measurement, press the [START/STOP] button once to deflate 

the arm cuff.

• Remain still and do not talk while taking a measurement.
• This device will store 100 measurements in the device. Stored 

measurements cannot be viewed on the monitor's display. You will need to 
use the “OMRON connect” app to view all measurements prior to the most 
recent reading taken.

• Do not use the monitor with other electrical devices simultaneously. This 

may result in incorrect operation of the monitor.

1.

Press the [START/STOP] button.

The arm cuff will start to inflate.

 : The arm cuff was wrapped snugly enough.
 : The arm cuff was loose or not wrapped correctly.

Note: The cuff wrap guide result appears on the display during the 

measurement.

The following symbols indicate incorrect measurement

• If any of these symbols below are displayed, please take another 

measurement and try again.

2.

Remove the arm cuff.

3.

Transfer your readings.

Please make sure your monitor is within 5 m (16 ft.) of your smart device 
and your “OMRON connect” app is active.
As soon as the measurement is complete, your reading will be automatically 
transferred to the app.

Transferring

 your readings later

1. Open the app on your smart device and then follow the instructions.
2. Press connection button to transfer your readings.

• If your readings are successfully transferred to your smart device, 

“OK” will flash on the display.

If “Err” appears, refer to “Connection failure. / Data is not being 
transmitted.” in section 4.2 for more detail.

Notes:
• When the SYNC symbol “

” appears on the display, transfer the data 

immediately before the readings are deleted.

• When the Connection OFF symbol “

” appears on the display, turn on the 

Bluetooth

 feature of your monitor. (Refer to section 3.4)

• Once you send your readings to the app, the data cannot be resent to the 

app or smart device.

4.

After your readings have been transferred to the “OMRON 
connect” app, press the [START/STOP] button to turn off.

Notes:
• The monitor will automatically turn off after 2 minutes.
• Wait 2-3 minutes before taking another measurement. Waiting between 

measurements allows the arteries to return to their prior condition to taking 
the measurement.

• If you use any other app instead of “OMRON connect”, the data may not 

be displayed correctly after sending.

Data Transfer With Multiple Smart Devices

Once you transfer your data to a smart device, that data cannot be transferred 
again to the same smart device, nor to another smart device.

About Multiple Users 

This monitor is recommended for personal use. 
When this monitor is used by more than one person, the measurement data 
will be combined. 

Using your Monitor without Pairing

You can use this blood pressure monitor without pairing to a smart device.
1. Apply the arm cuff. (Refer to section 3.1)
2. Press [START/STOP] button. (Refer to step 1 to 2 in section 3.3)
3. Write down your SYS/DIA and Pulse readings, along with the date and time 

of your measurement on a piece of paper to keep a record of your history.

Note: When using this blood pressure monitor without a smart device, you 

will not be able to view your prior readings stored in the memory.

3.4 Turning “OFF” Bluetooth

®

Turn off the 

Bluetooth

®

 in your monitor in the following areas where use of 

wireless equipment is prohibited.
• On aircraft  • In hospitals  • While abroad
Your blood pressure monitor comes with the 

Bluetooth

®

 already turned on.

1.

Turn your monitor power off if it is turned on.

2.

Press and hold the connection button for more than 
10 seconds.

3.

Release the connection button when “

” appears on 

the display.

4.

Press the [START/STOP] button to turn the monitor off.

Note: The monitor will automatically turn off after 2 minutes.

How to turn “ON” Bluetooth

®

Operate the following step when the power has been off.
Press and hold the connection button for more than 2 seconds and 
then release the connection button when 

 appears on the 

display.

3.5 Restoring Your Monitor to the Default Settings

To delete all the information stored in your monitor, follow the instructions 
below. Make sure that your monitor is turned off. It will be off when the 
screen is not showing numbers or symbols. 

1.

While holding the connection button down, the Connection 
symbol “ ” and “P” appear on the display, and then press 
the [START/STOP] button for more than 5 seconds.

2.

When the “Clr” symbol appears on the display, release the 
connection button and [START/STOP] button.

The “Clr” will appear on the display when your monitor settings have 
reverted back to their original.

3.

Press the [START/STOP] button to turn the monitor off.

Notes:
• The monitor will automatically turn off after 2 minutes.
• Reverting to initial settings of this monitor will not delete the information in 

the app.

• When using this monitor again, please delete the registration of the 

monitor in the app first, and then pair your monitor.

Automatic Upper Arm 

Blood Pressure Monitor

EVOLV (HEM-7600T-E)

Instruction Manual

Connection symbol

Battery symbol (low)

Connection OFF symbol

Battery symbol (depleted)

Connection ON symbol

Deflation symbol

Movement error symbol

Heartbeat symbol
(Flashes during measurement.)

SYNC symbol

Cuff Wrap Guide Symbol

Irregular heartbeat symbol

Office

Home

Systolic Blood Pressure

140 mmHg

135 mmHg

Diastolic Blood Pressure

90 mmHg

85 mmHg

A

B

C

D

E

A. START/STOP button
B. Connection button
C. Display
D. Arm cuff
E. Battery compartment

F. Systolic blood pressure
G. Diastolic blood pressure
H. Pulse display

G

H

F

Display:

Irregular Heartbeat

Normal Heartbeat

Pulse

Blood pressure

Blood pressure

Short Long

Pulse

 : Make sure to press the connection button to pair with the smart 

device or transfer the data before taking a measurement when 
the SYNC symbol 

“ ”

 appears.

Review the list of compatible smart devices at: 
omronconnect.com/devices
For details, visit: omronconnect.com

1 - 2 cm

START

INFLATING

DEFLATING

COMPLETED

Example:

®

DO NOT adjust medication based on measurement results from this blood 
pressure monitor. Take medication as prescribed by your physician. Only a 
physician is qualified to diagnose and treat high blood pressure.
Always consult your physician. Self-diagnosis of measurement results and 
self-treatment are dangerous.

(2 sec.)

(8 sec.)

IM-HEM-7600T-E-EN-05-04/2021

9545878-0F

Issue Date: 2021-04-16

21D1080

9545878-0F_HEM-7600T-E_IM_M02_210419.pdf

9545878-0F_IM-HEM-7600T-E-EN-05-04-2021.fm  Page 1  Tuesday, July 6, 2021  12:13 PM

Summary of Contents for EVOLV

Page 1: ...esult in minor or moderate injury to the user or patient or damage to the equipment or other property Stop using this monitor and consult your physician if you experience skin irritation or discomfort...

Page 2: ...25 less or 25 more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure If the irregular heartbeat symbol displays with your measurement results w...

Page 3: ...ordance with the national local regulations for the disposal of batteries 2 2 Pairing Your Monitor with a Smart Device The date and time on your monitor will automatically be set when you pair it with...

Page 4: ...ff Remain still and do not talk while taking a measurement This device will store 100 measurements in the device Stored measurements cannot be viewed on the monitor s display You will need to use the...

Page 5: ...e use of wireless equipment is prohibited On aircraft In hospitals While abroad Your blood pressure monitor comes with the Bluetooth already turned on 1 Turn your monitor power off if it is turned on...

Page 6: ...ent Refer to section 3 3 Arm cuff is not applied correctly Apply the arm cuff correctly Refer to section 3 1 XX will change to display an error number 6 and higher Device error Contact your local OMRO...

Page 7: ...used for continuous operation This blood pressure monitor is designed according to the European Standard EN1060 Non invasive sphygmomanometers Part 1 General Requirements and Part 3 Supplementary req...

Page 8: ...e Technology by OMRON Healthcare in Japan To indicate generally elevated potentially hazardous levels of non ionizing radiation or to indicate equipment or systems e g in the medical electrical area t...

Page 9: ...sure is constantly changing during the course of the heart s cycle The highest pressure in the cycle is called the Systolic Blood Pressure the lowest is the Diastolic Blood Pressure Both pressures the...

Page 10: ...cility OMRON HEALTHCARE Co Ltd Matsusaka Factory 1855 370 Kubo cho Matsusaka shi Mie 515 8503 Japan Subsidiaries OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 0DG UK www omron health...

Reviews: