13
D6F-PH MEMS Differential pressure Sensor User’s Manual (A288)
7-3. Description of Registers
The internal memory and registers of the sensor module can be accessed via the interface
configuration registers.
7-3-1. Sensor Control (D040h)
Table10. SENS_CTRL
Address
MSB
D7
D6
D5
D4
D3
D2
D1
LSB
D0
D040h
MS
DV_PWR[
1]
DV_PWR[
0]
Write
Access
None
None
None
None
None
Host &
MCU
Host &
MCU
Host &
MCU
Default
0
0
0
0
0
0
0
0
•
DV_PWR[1
:
0] – Main Device power mode setting
0 0 = Standby – All blocks are powered down.
1 0 = MCU on – Used when only MCU is required. Basic analog and memories are
powered on and MCU clock is running.
Note
:
This register should not be changed during a measurement.
•
MS – MCU start – Begin execution of measurement or MCU mode based on the state of
DV_PWR.
•
0 = Stop
Sequences are stopped and MCU clock is turned off.
•
1 = Start
The MCU clock is started and the MCU mode is executed.