background image

3

Wash the parts.

Wash the inhalation top, medication tank and accessories (mouthpiece, 
mask or child accessory) in warm water with mild (neutral) detergent and 
rinse all pieces thoroughly with clean hot tap water. Wash the vaporizer 
head with running tap water.

 WARNING

•  DO NOT leave the cleaning solution on the parts. Rinse the parts with clean 

tap water after cleaning.

•  DO NOT immerse the main unit or the AC adapter in water or other 

liquid.

Inhalation 

top

Medication 

tank

Mouthpiece

Mask

Child 

accessory

Vaporizer head

Main unit

Air tube

Wash in warm water with mild (neutral) detergent

Wash with running tap water

DO NOT wash

AC adapter

4

Dry the parts.

Allow to air dry in a clean environment.

 WARNING

After cleaning, allow the parts to dry before reassembling. To avoid 
contamination, store the device and parts in a clean and safe location.

Cleaning the Device After Each Use

The following items should be disassembled and cleaned after each use in order to prevent any remaining medication in the tank from drying 
which will result in ineffective nebulization.
•  Nebulizer kit (inhalation top, medication tank and vaporizer head)
•  Mouthpiece
•  Mask

Refer to steps above to clean each item.

Cleaning the Main Unit and Air Tube

Clean the casing of the main unit and the air tube by using a soft cloth 
moistened with water or a mild detergent. Do not use abrasive cleaners. 
Dry the casing and air tube immediately using a soft clean cloth.

 WARNING

•  DO NOT immerse the main unit or the AC adapter in water or other liquid.

 CAUTION

•  Unplug the AC adapter before cleaning this device.
•  DO NOT insert any object into the main unit. 

Cleaning the Device Weekly

The following items should be cleaned by boiling once a week. If the parts are heavily stained, replace them with new ones.
•  Nebulizer kit (inhalation top, medication tank and vaporizer head)
•  Mouthpiece

Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess water and allow 
to air dry in a clean environment.

For additional detail, consult the COMP A-I-R Compressor Nebulizer Instruction Manual – Model RE-C800KD.